Functional apparent moduli as predictors of oral implant osseointegration dynamics

被引:9
作者
Chang, Po-Chun [1 ,2 ,3 ]
Seol, Yang-Jo [1 ,3 ]
Kikuchi, Noboru [4 ]
Goldstein, Steven A. [2 ,5 ]
Giannobile, William V. [1 ,2 ,6 ]
机构
[1] Univ Michigan, Sch Dent, Dept Periodont & Oral Med, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Coll Engn, Dept Biomed Engn, Ann Arbor, MI 48109 USA
[3] Natl Univ Singapore, Fac Dent, Dept Prevent Dent, Singapore 119083, Singapore
[4] Univ Michigan, Coll Engn, Dept Mech Engn, Ann Arbor, MI 48109 USA
[5] Univ Michigan, Sch Med, Dept Orthopaed Surg, Ann Arbor, MI 48109 USA
[6] Univ Michigan, Sch Dent, Michigan Ctr Oral Hlth Res, Ann Arbor, MI 48109 USA
关键词
dental implant; osseointegration; micro-computed tomography; finite element analysis; FINITE-ELEMENT-ANALYSIS; MICROFOCUS COMPUTED-TOMOGRAPHY; TRABECULAR BONE; TITANIUM IMPLANTS; MECHANICAL-PROPERTIES; ALVEOLAR BONE; STABILITY; BEHAVIOR; HISTOMORPHOMETRY; NANOINDENTATION;
D O I
10.1002/jbm.b.31631
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
At present, limited functional data exists regarding the application and use of biomechanical and imaging technologies for oral implant osseointegration assessment. The objective of this investigation was to determine the functional apparent moduli (FAMs) that could predict the dynamics of oral implant osseointegration. Using an in vivo dental implant osseous healing model, two FAMs, functional bone apparent modulus (FBAM), and composite tissue apparent modulus (FCAM), of the selected pen-implant structures were calculated via microcomputed tomography (micro-CT) and finite element (FE) simulations in order to support this concept. Results showed significant sensitivity between FAMs and micro-CT parameters, especially between bone mineral density and FBAM, while at extraction defect sites the strongest correlations existed between bone-implant contact and FCAM. Significant enhancement of FCAM indicated progressive functional repair during early osseointegration. Further, the resultant interfacial resistance was predicted by bone mineral content (BMC) and FBAM within a similar to 200 mu m pen-implant thickness, while the extraction defects gave zones of similar to 575 mu m and 200 mu m for BMC and FCAM, respectively. These results suggest that the function of dental implant support can be predicted from a peri-implant structural zone. We conclude that FAMs can be used to predict the dynamics of dental implant osseointegration in vivo. (C) 2010 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 94B:118-126, 2010.
引用
收藏
页码:118 / 126
页数:9
相关论文
共 37 条
[1]   Validity and clinical significance of biomechanical testing of implant/bone interface [J].
Aparicio, Carlos ;
Lang, Niklaus R. ;
Rangert, Bo .
CLINICAL ORAL IMPLANTS RESEARCH, 2006, 17 :2-7
[2]  
Atsumi M, 2007, INT J ORAL MAX IMPL, V22, P743
[3]   Artifacts in CT: Recognition and avoidance [J].
Barrett, JF ;
Keat, N .
RADIOGRAPHICS, 2004, 24 (06) :1679-1691
[4]   Scanning acoustic microscopy study of human cortical and trabecular bone [J].
Bumrerraj, S ;
Katz, JL .
ANNALS OF BIOMEDICAL ENGINEERING, 2001, 29 (12) :1034-1042
[5]   Harder and stiffer bone Osseointegrated to roughened titanium [J].
Butz, F. ;
Aita, H. ;
Wang, C. J. ;
Ogawa, T. .
JOURNAL OF DENTAL RESEARCH, 2006, 85 (06) :560-565
[6]   Validation of a voxel-based FE method for prediction of the uniaxial apparent modulus of human trabecular bone using macroscopic mechanical tests and nanoindentation [J].
Chevalier, Yan ;
Pahr, Dieter ;
Allmer, Helga ;
Charlebois, Mathieu ;
Zysset, Philippe .
JOURNAL OF BIOMECHANICS, 2007, 40 (15) :3333-3340
[7]   A FINITE-ELEMENT ANALYSIS OF THE PUSH-OUT TEST - INFLUENCE OF TEST CONDITIONS [J].
DHERT, WJA ;
VERHEYEN, CCPM ;
BRAAK, LH ;
DEWIJN, JR ;
KLEIN, CPAT ;
DEGROOT, K ;
ROZING, PM .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (01) :119-130
[8]   BMP gene delivery for alveolar bone engineering at dental implant defects [J].
Dunn, CA ;
Jin, QM ;
Taba, M ;
Franceschi, RT ;
Rutherford, RB ;
Giannobile, WV .
MOLECULAR THERAPY, 2005, 11 (02) :294-299
[9]   Mechanical properties of OI type III bone tissue measured by nanoindentation [J].
Fan, Zaifeng ;
Smith, Peter A. ;
Eckstein, Eugene C. ;
Harris, Gerald F. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2006, 79A (01) :71-77
[10]   Biological fixation of endosseous implants [J].
Franchi, M ;
Fini, M ;
Martini, D ;
Orsini, E ;
Leonardi, L ;
Ruggeri, A ;
Giavaresi, G ;
Ottani, V .
MICRON, 2005, 36 (7-8) :665-671