Respiration-induced attenuation artifact at PET/CT: Technical considerations

被引:150
作者
Goerres, GW
Burger, C
Kamel, E
Seifert, B
Kaim, AH
Buck, A
Buehler, TC
von Schulthess, GK
机构
[1] Univ Zurich, Div Nucl Med, CH-8091 Zurich, Switzerland
[2] Univ Zurich, Dept Biostat, CH-8091 Zurich, Switzerland
关键词
heart; PET; images; processing; lung; positron emission tomography (PET); artifact; technology;
D O I
10.1148/radiol.2263011732
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Combined positron emission tomographic (PET)/computed tomographic (CT) scanners allow the use of CT data for attenuation correction of PET images. Eight patients with cancer underwent PET/CT scanning. Transmission scanning was performed with conventional attenuation correction and with CT scanning during maximum inspiration and normal expiration. Image quality was visually compared and fluorine 18 activities were measured in volumes of interest in the lung and myocardium. Analysis of variance for repeated measures revealed a significant decrease (P = .0001) in measured activities be tween PET images corrected with CT data acquired during maximum inspiration and those corrected with the conventional attenuation correction method or with CT data acquired during normal expiration. Deep inspiration during CT can result in severe deterioration in the final PET image. (C) RSNA, 2003.
引用
收藏
页码:906 / 910
页数:5
相关论文
共 7 条
  • [1] FDG imaging of lung nodules: A phantom study comparing SPECT, camera-based PET, and dedicated PET
    Coleman, RE
    Laymon, CM
    Turkington, TG
    [J]. RADIOLOGY, 1999, 210 (03) : 823 - 828
  • [2] PET-CT image co-registration in the thorax: influence of respiration
    Goerres, GW
    Kamel, E
    Heidelberg, TNH
    Schwitter, MR
    Burger, C
    von Schulthess, GK
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (03) : 351 - 360
  • [3] CT vs 68Ge attenuation correction in a combined PET/CT system:: evaluation of the effect of lowering the CT tube current
    Kamel, E
    Hany, TF
    Burger, C
    Treyer, V
    Lonn, AHR
    von Schulthess, GK
    Buck, A
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (03) : 346 - 350
  • [4] Attenuation correction for a combined 3D PET/CT scanner
    Kinahan, PE
    Townsend, DW
    Beyer, T
    Sashin, D
    [J]. MEDICAL PHYSICS, 1998, 25 (10) : 2046 - 2053
  • [5] Kluetz P G., 2000, Clin Positron Imaging, V3, P223, DOI 10.1016/S1095-0397(01)00055-3
  • [6] A JAVA']JAVA environment for medical image data analysis: initial application for brain PET quantitation
    Mikolajczyk, K
    Szabatin, M
    Rudnicki, P
    Grodzki, M
    Burger, C
    [J]. MEDICAL INFORMATICS, 1998, 23 (03): : 207 - 214
  • [7] MOVEMENTS OF THE THORACIC CAGE AND DIAPHRAGM IN RESPIRATION
    WADE, OL