Analysis of quantitative [I-123] mIBG SPECT/CT in a phantom and in patients with neuroblastoma

被引:13
作者
Brady, Samuel L. [1 ]
Shulkin, Barry L. [2 ]
机构
[1] Univ Cincinnati, Cincinnati Childrens Hosp Med Ctr, Dept Radiol, Cincinnati, OH 45229 USA
[2] St Jude Childrens Res Hosp, Dept Diagnost Imaging, MS 220,262 Danny Thomas Pl, Memphis, TN 38105 USA
关键词
mIBG; Nuclear medicine; Neuroblastoma; Pediatrics; SCATTER CORRECTION; MONTE-CARLO; ATTENUATION; ACCURACY; COMPENSATION; VALIDATION; RECOVERY;
D O I
10.1186/s40658-019-0267-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To determine the accuracy of quantitative SPECT, intersystem and interpatient standardized uptake value (SUV) calculation consistency for a manufacturer-independent quantitative SPECT/CT reconstruction algorithm, and the range of SUVs of normal and neoplastic tissue. Methods A NEMA body phantom with 6 spheres (ranging 10-37 mm) was filled with a known activity-to-volume ratio and used to determine the contrast recovery coefficient (CRC) for each visible sphere, and the measured SUV accuracy of those spheres and background water solution. One hundred eleven 123I-metaiodobenzylguanidine ([I-123] mIBG) SPECT/CT examinations from 43 patients were reconstructed using SUV SPECT (R) (HERMES Medical Solutions Inc.); 42 examinations were acquired using a GE Infinia Hawkeye 4 SPECT/CT, and 69 were acquired on a Siemens Symbia Intevo SPECT/CT. Inter scanner SUV analysis of 9 regions of normal [I-123] mIBG tissue uptake was conducted. Intrapatient mean SUV variability was calculated by measuring normal liver uptake within patients scanned on both cameras. The intensity of uptake by neoplastic tissue in the images was quantified using maximum SUV and, if present, compared over time. Results The phantom results of the visible spheres and background resulted in accuracy calculations better than 5-10% with CRC correction. Interscanner SUV variability showed no statistical difference (average p value 0.559; range 0.066-1.0) among the 9 normal tissues analyzed. Intrapatient liver mean SUV varied <= 16% as calculated for 28 patients (87 examinations) studied on both scanners. In one patient, a thoracic tumor evaluated over 10 time points (18 months) underwent a 74% (3.1/12.0) reduction in maximum SUV with treatment. Conclusion The results demonstrate quantitative accuracy to better than 10%, and both consistent SUV calculation between 2 different SPECT/CT scanners for 9 tissues, and low intrapatient measurement variability for quantitative SPECT/CT analysis in a pediatric population with neuroblastoma. Quantitative SPECT/CT offers the opportunity for objective analysis of tumor response using [I-123] mIBG by normalizing the uptake to injected dose and patient weight, as is done for PET.
引用
收藏
页数:14
相关论文
共 29 条
[1]   Factors affecting the repeatability of gamma camera calibration for quantitative imaging applications using a sealed source [J].
Anizan, N. ;
Wang, H. ;
Zhou, X. C. ;
Wahl, R. L. ;
Frey, E. C. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (03) :1325-1337
[2]   Quantitative SPECT/CT: SPECT joins PET as a quantitative imaging modality [J].
Bailey, Dale L. ;
Willowson, Kathy P. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 :S17-S25
[3]   An Evidence-Based Review of Quantitative SPECT Imaging and Potential Clinical Applications [J].
Bailey, Dale L. ;
Willowson, Kathy P. .
JOURNAL OF NUCLEAR MEDICINE, 2013, 54 (01) :83-89
[4]   Implementation of GPU accelerated SPECT reconstruction with Monte Carlo-based scatter correction [J].
Bexelius, Tobias ;
Sohlberg, Antti .
ANNALS OF NUCLEAR MEDICINE, 2018, 32 (05) :337-347
[5]   Measuring total liver function on sulfur colloid SPECT/CT for improved risk stratification and outcome prediction of hepatocellular carcinoma patients [J].
Bowen, Stephen R. ;
Chapman, Tobias R. ;
Borgman, Joshua ;
Miyaoka, Robert S. ;
Kinahan, Paul E. ;
Liou, Iris W. ;
Sandison, George A. ;
Vesselle, Hubert J. ;
Nyflot, Matthew J. ;
Apisarnthanarax, Smith .
EJNMMI RESEARCH, 2016, 6
[6]   Quantification of 99mTc-DPD concentration in the lumbar spine with SPECT/CT [J].
Cachovan, Michal ;
Vija, Alexander Hans ;
Hornegger, Joachim ;
Kuwert, Torsten .
EJNMMI RESEARCH, 2013, 3 :1-8
[7]   Hybrid scatter correction applied to quantitative holmium-166 SPECT [J].
de Wit, Tim C. ;
Xiao, Jianbin ;
Nijsen, J. Frank W. ;
Schip, Fred D. van het ;
Staelens, Steven G. ;
van Rijk, Peter P. ;
Beekman, Freek J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (19) :4773-4787
[8]   Model-based compensation for quantitative 123I brain SPECT imaging [J].
Du, Y ;
Tsui, BMW ;
Frey, EC .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (05) :1269-1282
[9]   A Monte Carlo and physical phantom evaluation of quantitative In-111SPECT [J].
He, B ;
Du, Y ;
Song, XY ;
Segars, WP ;
Frey, EC .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (17) :4169-4185
[10]   Quantitative accuracy of 177Lu SPECT reconstruction using different compensation methods: phantom and patient studies [J].
Hippelainen, Eero ;
Tenhunen, Mikko ;
Maenpaa, Hanna ;
Sohlberg, Antti .
EJNMMI RESEARCH, 2016, 6 :1-8