Assessment of the gain transfer function of phosphors for application in medical imaging radiation detectors

被引:15
作者
Cavouras, D
Kandarakis, I
Maris, T
Panayiotakis, GS
Nomicos, CD
机构
[1] TEI Athens, Dept Med Instrumentat Technol, Athens 12210, Greece
[2] Univ Crete, Sch Med, Univ Hosp, Dept Radiol, Heraklion, Greece
[3] Univ Patras, Sch Med, Dept Phys Med, Patras 26500, Greece
[4] TEI Athens, Dept Elect, Athens 12210, Greece
关键词
phosphors; X-ray luminescence; MTF; spectral compatibility;
D O I
10.1016/S0720-048X(99)00106-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: to study various phosphors used in detectors of medical imaging systems by the gain transfer function (GTF), defined in terms of X-ray luminescence efficiency, light spectrum and modulation transfer function. Materials and methods: four phosphor materials, La2O2S:Tb, Y2O2S:Tb, Y2O2S:Eu and Y2O3:Eu were used in the form of fluorescent layers prepared in the laboratory. The GTF was determined at 30 kVp and 80 kVp X-ray tube voltages for various phosphor coating weights. Results: La2O2S:Tb, which was the highest density and effective atomic number phosphor used, was found to exhibit the best GTF performance at 80 kVp. At 30 kVp, the yttrium based phosphors were found of increased performance. This is mainly due to the proximity of the X-ray energy to the K-absorption edge of yttrium at 17 keV. Europium activated phosphors were found to perform very well when combined with the red sensitive film and the silicon photodiode. Conclusion: The GTF may be a useful method for comparing and selecting phosphor materials for use in various medical imaging applications. (C) 2000 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:70 / 77
页数:8
相关论文
共 16 条
[1]  
[Anonymous], PHYS MED IMAGING REC
[2]  
[Anonymous], 1986, 41 ICRU
[3]  
Barnes G. T., 1979, PHYSICS MED IMAGING, P138
[4]   An evaluation of the Y2O3:Eu3+ scintillator for application in medical x-ray detectors and image receptors [J].
Cavouras, D ;
Kandarakis, I ;
Panayiotakis, GS ;
Evangelou, EK ;
Nomicos, CD .
MEDICAL PHYSICS, 1996, 23 (12) :1965-1975
[5]   An experimental method to determine the effective luminescence efficiency of scintillator-photodetector combinations used in X-ray medical imaging systems [J].
Cavouras, D ;
Kandarakis, I ;
Bakas, A ;
Triantis, D ;
Nomicos, CD ;
Panayiotakis, GS .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (847) :766-772
[6]  
CURRY TS, 1990, CHRISTENSENS PHYSICS, P118
[7]  
DANTY JC, 1974, IMAGE SCI, P152
[8]  
GREENING JR, 1985, MED PHYSICS HDB, P56
[9]  
HENDEE WR, 1970, MED RAD PHYSICS, P145
[10]   An experimental method for the determination of spatial-frequency-dependent detective quantum efficiency (DQE) of scintillators used in X-ray imaging detectors [J].
Kandarakis, I ;
Cavouras, D ;
Panayiotakis, GS ;
Triantis, D ;
Nomicos, CD .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 399 (2-3) :335-342