Dosimetric characteristics of a new 125I brachytherapy source

被引:32
作者
Gearheart, DM [1 ]
Drogin, A [1 ]
Sowards, K [1 ]
Meigooni, AS [1 ]
Ibbott, GS [1 ]
机构
[1] Univ Kentucky, Albert B Chandler Med Ctr, Dept Radiat Med, Lexington, KY 40536 USA
关键词
brachytherapy; dosimetry; anisotropy function; radial dose function; dose rate constant;
D O I
10.1118/1.1290486
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
I-125 brachytherapy sources are being used for interstitial implants in tumor sites such as the prostate. Recently, a new I-125 source has been introduced, which has a design different from that of other sources presently on the market. Dosimetric characteristics of this source, including dose rate constant, radial dose function, and anisotropy function, were determined experimentally following the AAPM Task Group 43 recommendations. The characteristics were related to the 1999 NIST calibration assigned to this source [S-K,S-99std]. Measurements were performed in a solid water phantom using LiF thermoluminescent dosimeters. For these measurements, slabs of solid water phantom material were machined to accommodate the source and LiF TLD chips of dimensions (3.1 x3.1x0.8 mm(3)) and (1.0x1.0x1.0 mm(3)). The TLD chips were surrounded by at least 10 cm of solid water phantom material to provide full scattering conditions. The results indicated a dose rate constant, Lambda, of 0.88+/-0.07 cGy h(-1) U-1 for the new I-125 source as compared to 0.98 and 1.04 cGy h(-1) U-1 for the Nycomed/Amersham model 6711 and 6702 seeds, respectively. Per TG-43, the values reported here represent the dose absorbed by water at 1 cm from the source in a water medium. The radial dose function, g(r), of the new I-125 source was measured at distances ranging from 0.5 to 10 cm. The anisotropy function, F(r, theta), of the new I-125 source was measured at distances of 2 and 5 cm from the source center. Calculations of anisotropy and radial dose function were also made using a Monte Carlo code. These calculations were made for both solid water and liquid water, the former to validate the Monte Carlo code and the latter to provide results in liquid water for clinical use. All data compared favorably with those from the Nycomed/Amersham models 6711 and 6702 sources. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)00310-2].
引用
收藏
页码:2278 / 2285
页数:8
相关论文
共 12 条
[1]   Biochemical outcome after radical prostatectomy, external beam radiation therapy, or interstitial radiation therapy for clinically localized prostate cancer [J].
D'Amico, AV ;
Whittington, R ;
Malkowicz, SB ;
Schultz, D ;
Blank, K ;
Broderick, GA ;
Tomaszewski, JE ;
Renshaw, AA ;
Kaplan, I ;
Beard, CJ ;
Wein, A .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1998, 280 (11) :969-974
[2]   Uncertainty of calibrations at the accredited dosimetry calibration laboratories [J].
Ibbott, GS ;
Attix, FH ;
Slowey, TW ;
Fontenla, DP ;
Rozenfeld, M .
MEDICAL PHYSICS, 1997, 24 (08) :1249-1254
[3]   INSTRUMENTATION AND DOSIMETER-SIZE ARTIFACTS IN QUANTITATIVE THERMOLUMINESCENCE DOSIMETRY OF LOW-DOSE FIELDS [J].
MEIGOONI, AS ;
MISHRA, V ;
PANTH, H ;
WILLIAMSON, J .
MEDICAL PHYSICS, 1995, 22 (05) :555-561
[4]   DOSIMETRY OF INTERSTITIAL BRACHYTHERAPY SOURCES - RECOMMENDATIONS OF THE AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO 43 [J].
NATH, R ;
ANDERSON, LL ;
LUXTON, G ;
WEAVER, KA ;
WILLIAMSON, JF ;
MEIGOONI, AS .
MEDICAL PHYSICS, 1995, 22 (02) :209-234
[5]  
PEREZ CA, 1997, PRINCIPLES PRACTICE, P487
[6]  
PEREZ CA, 1997, PRINCIPLES PRACTICE, P1583
[7]   Report on the dosimetry of a new design 125Iodine brachytherapy source [J].
Wallace, RE ;
Fan, JJ .
MEDICAL PHYSICS, 1999, 26 (09) :1925-1931
[8]   Dosimetric prerequisites for routine clinical use of new low energy photon interstitial brachytherapy sources [J].
Williamson, J ;
Coursey, BM ;
DeWerd, LA ;
Hanson, WF ;
Nath, R .
MEDICAL PHYSICS, 1998, 25 (12) :2269-2270
[9]   MONTE-CARLO EVALUATION OF KERMA AT A POINT FOR PHOTON TRANSPORT PROBLEMS [J].
WILLIAMSON, JF .
MEDICAL PHYSICS, 1987, 14 (04) :567-576
[10]   Guidance to users of Nycomed Amersham and North American Scientific, Inc., I-125 Interstitial Sources: Dosimetry and calibration changes: Recommendations of the American Association of Physicists in Medicine Radiation Therapy Committee Ad Hoc Subcommittee on Low-Energy Seed Dosimetry [J].
Williamson, JF ;
Coursey, BM ;
DeWerd, LA ;
Hanson, WF ;
Nath, R ;
Ibbott, G .
MEDICAL PHYSICS, 1999, 26 (04) :570-573