Molecular depth profiling of histamine in ice using a buckminsterfullerene probe

被引:77
作者
Wucher, A [1 ]
Sun, SX
Szakal, C
Winograd, N
机构
[1] Univ Duisburg, Dept Phys, D-45117 Essen, Germany
[2] Penn State Univ, Dept Chem, University Pk, PA 16802 USA
关键词
D O I
10.1021/ac049641t
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We employ a buckminsterfullerene ion source to probe the distribution of histamine molecules at the water-ice/vacuum interface. The experiments utilize secondary ion mass spectrometry to detect molecular ions that are desorbed from a frozen aqueous histamine solution. The results show that this cluster ion probe induces an extraordinarily high sputter yield of 2400 ice molecules per impact event as determined by a quartz crystal microbalance. As a consequence of this high yield, we show that it is possible to produce molecular depth profiles of the top several hundred nanometers below the ice surface without destruction of the molecular ion signal by accumulation of beam-induced chemical damage. Similar profiles are reported for desorbed neutral molecular fragments by utilizing a high-power femtosecond-pulsed laser for photoionization. While this type of information could not be achieved using atomic projectiles, it is possible to remove the damage induced by such projectiles by subsequent cluster bombardment. These experiments are particularly important for organic surface analysis since they suggest that cluster ion probes may successfully be employed to remove overlayers that may mask the desired molecular information in static secondary ion mass spectral analysis.
引用
收藏
页码:7234 / 7242
页数:9
相关论文
共 30 条
[1]   Sputtering of water ice [J].
Baragiola, RA ;
Vidal, RA ;
Svendsen, W ;
Schou, J ;
Shi, M ;
Bahr, DA ;
Atteberrry, CL .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2003, 209 :294-303
[2]  
Braun RM, 1998, RAPID COMMUN MASS SP, V12, P1246, DOI 10.1002/(SICI)1097-0231(19980930)12:18<1246::AID-RCM316>3.0.CO
[3]  
2-C
[4]   SPUTTERING OF ICES BY KEV IONS [J].
CHRISTIANSEN, JW ;
CARPINI, DD ;
TSONG, IST .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1986, 15 (1-6) :218-221
[5]   Development of instrumentation for routine ToF-SIMS imaging analysis of biological material [J].
Cliff, B ;
Lockyer, NP ;
Corlett, C ;
Vickerman, JC .
APPLIED SURFACE SCIENCE, 2003, 203 :730-733
[6]  
CORNETT DS, 1994, RAPID COMMUN MASS SP, V8, P996
[7]  
DIEHNELT CW, 1998, SECONDARY ION MASS S, P593
[8]   Surface analysis studies of yield enhancements in secondary ion mass spectrometry by polyatomic projectiles [J].
Fuoco, ER ;
Gillen, G ;
Wijesundara, MBJ ;
Wallace, WE ;
Hanley, L .
JOURNAL OF PHYSICAL CHEMISTRY B, 2001, 105 (18) :3950-3956
[9]   Formation of sputtered silver clusters under bombardment with SF5+ ions [J].
Ghalab, S ;
Staudt, C ;
Maksimov, SE ;
Marsarov, P ;
Tugushev, VI ;
Dzhemilev, NK ;
Wucher, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2002, 197 (1-2) :43-48
[10]   Secondary ion mass spectrometry using cluster primary ion beams [J].
Gillen, G ;
Fahey, A .
APPLIED SURFACE SCIENCE, 2003, 203 :209-213