Validation study of PulseCO system for continuous cardiac output measurement

被引:17
作者
Berberian, G
Quinn, TA
Vigilance, DW
Park, DY
Cabreriza, SE
Curtis, LJ
Spotnitz, HM
机构
[1] Columbia Univ Coll Phys & Surg, Dept Surg, New York, NY 10032 USA
[2] Columbia Univ Coll Phys & Surg, Dept Biomed Engn, New York, NY 10032 USA
关键词
D O I
10.1097/01.mat.0000150329.11072.C2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ultrasonic flow probes have been used to optimize biventricular pacing immediately after cardiopulmonary bypass, improving cardiac output (CO) by 10%; however, flow probes must be removed with chest closure. The PulseCO system (LiDCO Limited, Cambridge, UK) may extend optimization into the postoperative period, but controlled validations have not been reported. Six anesthetized pigs were instrumented for right heart bypass. Flow was varied from 3 to 1 L/min and then back to 3 in 0.5 L/min increments for 60 second intervals. CO was measured by ultrasonic flow probe on the aorta and by PulseCO using a femoral arterial line. PulseCO and flow probe accurately measured CO (PulseCO R-2: 0.79-0.95; flow probe R-2: 0.96-0.99). At flow of 2 L/min, when the heart was paced 30 bpm over the sinus rate, PulseCO falsely indicated an increase in CO (2.13 vs. 2.30 L/min, p = 0.014). When mean arterial pressure was increased by 20% using a phenylephrine infusion, PulseCO falsely indicated an increase in CO (2.13 vs. 2.47 L/min, p = 0.014). When mean arterial pressure was decreased by 20% using a nitroprusside infusion, PulseCO falsely indicated a decrease in CO (2.13 vs. 1.79 L/min, p = 0.003). PulseCO appears to be useful for assessing acute changes in CO if its limitations are recognized.
引用
收藏
页码:37 / 40
页数:4
相关论文
共 9 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Validation study of a new transit time ultrasonic flow probe for continuous great vessel measurements [J].
Dean, DA ;
Jia, CX ;
Cabreriza, SE ;
DAlessandro, DA ;
Dickstein, ML ;
Sardo, MJ ;
Chalik, N ;
Spotnitz, HM .
ASAIO JOURNAL, 1996, 42 (05) :M671-M676
[3]  
Hamilton TT, 2002, ANN THORAC SURG, V74, pS1408
[4]  
Jonas Max M, 2002, Curr Opin Crit Care, V8, P257, DOI 10.1097/00075198-200206000-00010
[5]   ESTIMATION OF STROKE VOLUME IN DOG BY A PULSE CONTOUR METHOD [J].
KOUCHOUKOS, NT ;
SHEPPARD, LC ;
MCDONALD, DA .
CIRCULATION RESEARCH, 1970, 26 (05) :611-+
[6]   Estimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb [J].
Linton, NWF ;
Linton, RAF .
BRITISH JOURNAL OF ANAESTHESIA, 2001, 86 (04) :486-496
[7]   VOLUME ELASTICITY CHARACTERISTICS OF THE HUMAN AORTA AND PREDICTION OF THE STROKE VOLUME FROM THE PRESSURE PULSE [J].
REMINGTON, JW ;
NOBACK, CR ;
HAMILTON, WF ;
GOLD, JJ .
AMERICAN JOURNAL OF PHYSIOLOGY, 1948, 153 (02) :298-308
[8]  
SPOTNITZ HM, IN PRESS J THORAC CA
[9]   COMPUTATION OF AORTIC FLOW FROM PRESSURE IN HUMANS USING A NONLINEAR, 3-ELEMENT MODEL [J].
WESSELING, KH ;
JANSEN, JRC ;
SETTELS, JJ ;
SCHREUDER, JJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1993, 74 (05) :2566-2573