Brachial plexus impingement secondary to implantable cardioverter defibrillator: A case report

被引:1
作者
Jumper, Natalie [1 ]
Radotra, Ishan [1 ]
Witt, Paulina [1 ]
Campbell, Niall G. [2 ]
Mishra, Anuj [1 ]
机构
[1] Univ Hosp South Manchester NHS Fdn Trust, Dept Plast & Reconstruct Surg, Southmoor Rd, Manchester M23 9LT, Lancs, England
[2] Univ Hosp South Manchester NHS Fdn Trust, Dept Cardiol, Manchester, Lancs, England
来源
ARCHIVES OF PLASTIC SURGERY-APS | 2019年 / 46卷 / 06期
关键词
Brachial plexus; Defibrillators; implantable; Brachial plexus neuropathies; SUBCLAVIAN VEIN CATHETERIZATION; INJURY;
D O I
10.5999/aps.2018.01158
中图分类号
R61 [外科手术学];
学科分类号
摘要
Overall complication rates of 9.1% have been reported following implantable cardioverter defibrillator (ICD) placement Brachial plexus injury is infrequently reported in the literature. We describe a 26-year-old female experiencing left arm nerve pain, a positive Tinel's sign, numbness in the median nerve distribution of the hand and biceps muscle weakness following revision ICD via subclavian vein approach. Nerve conduction studies identified severe partial left brachial plexopathy, which remained incompletely resolved with conservative management. Surgical exploration revealed lateral cord impingement by the ICD generator and a loop of the ICD lead, along with fibrosis, necessitating surgical neurolysis and ICD generator repositioning. As increasing numbers of patients undergo cardiac device implantation, it is incumbent on practitioners to be aware of potential increases in the prevalence of this complication.
引用
收藏
页码:594 / 598
页数:5
相关论文
共 15 条
[1]   A systematic review of ICD complications in randomised controlled trials versus registries: is our 'real-world' data an underestimation? [J].
Ezzat, Vivienne A. ;
Lee, Victor ;
Ahsan, Syed ;
Chow, Anthony W. ;
Segal, Oliver ;
Rowland, Edward ;
Lowe, Martin D. ;
Lambiase, Pier D. .
OPEN HEART, 2015, 2 (01)
[2]   Brachial plexopathies: Classification, causes, and consequences [J].
Ferrante, MA .
MUSCLE & NERVE, 2004, 30 (05) :547-568
[3]  
FIORISTA F, 1986, Archivos del Instituto de Cardiologia de Mexico, V56, P309
[4]  
Higgins SL, 2001, CARD ELECTROPHYSIOL, V5, P58
[5]   Brachial plexus injury during subclavian vein catheterization for hemodialysis [J].
Karakaya, D ;
Baris, S ;
Güldogus, F ;
Incesu, L ;
Sarihasan, B ;
Tür, A .
JOURNAL OF CLINICAL ANESTHESIA, 2000, 12 (03) :220-223
[6]   Comparison of the Axillary Venous Approach and Subclavian Venous Approach for Efficacy of Permanent Pacemaker Implantation [J].
Kim, Ki-Hun ;
Park, Kyoung-Min ;
Nam, Gi-Byoung ;
Kim, Dae-Kyeong ;
Oh, Minkyung ;
Choi, HyungOh ;
Hong, Taek Jong ;
Park, Bo-Min ;
Seo, Guang-Won ;
Song, Pil-Sang ;
Kim, Dong-Kie ;
Seol, Sang-Hoon ;
Kim, Doo-Il ;
Kim, You-Ho ;
Choi, Kee-Joon .
CIRCULATION JOURNAL, 2014, 78 (04) :865-871
[7]   Brachial Plexus Injury Caused by Indwelling Axillary Venous Pacing Leads [J].
Kim, So Yeon ;
Park, Jong Sung ;
Bang, Jung Hee ;
Kang, Eun Ju .
KOREAN CIRCULATION JOURNAL, 2015, 45 (05) :428-431
[8]  
Love CJ, 1999, CARD ELECTROPHYSIOL, V3, P36
[9]  
Moradzadeh A, 2010, ARCH FACIAL PLAST S, V12, P40, DOI 10.1001/archfacial.2009.104
[10]   Systematic review and meta-analysis of percutaneous subclavian vein puncture versus surgical venous cutdown for the insertion of a totally implantable venous access device [J].
Orci, L. A. ;
Meier, R. P. H. ;
Morel, P. ;
Staszewicz, W. ;
Toso, C. .
BRITISH JOURNAL OF SURGERY, 2014, 101 (02) :8-16