In vitro biomechanical investigation of the stability and stress-shielding effect of lumbar interbody fusion devices

被引:113
作者
Kanayama, M
Cunningham, BW
Haggerty, CJ
Abumi, K
Kaneda, K
McAfee, PC
机构
[1] Union Mem Hosp, Orthopaed Biomech Lab, Baltimore, MD USA
[2] Scoliosis & Spine Ctr, Towson, MD USA
[3] Hokkaido Univ, Sch Med, Dept Orthopaed Surg, Sapporo, Hokkaido 060, Japan
关键词
lumbar interbody fusion; threaded fusion cage; biomechanics; stress shielding; spine model;
D O I
10.3171/spi.2000.93.2.0259
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. Interbody fusion devices are rapidly gaining acceptance as a method of ensuring lumbar interbody arthrodesis. Although different types of devices have been developed, the comparative reconstruction stability remains controversial. It also remains unclear how different stress-shielded environments are created within the devices. Using a calf spine model, this study was designed to compare the construct stiffness afforded by II differently designed lumbar interbody fusion devices and to quantify their stress-shielding effects by measuring pressure within the devices. Methods. Sixty-six lumbar specimens obtained from calves were subjected to anterior interbody reconstruction at L4-5 by using one of the following interbody fusion devices: four different threaded fusion cages (BAK device, BAK Proximity, Ray TFC, and Danek TIBFD), five different nonthreaded fusion devices (oval and circular Harms cages, Brantigan PLIF and ALIF cages, and InFix device); two different types of allograft (femoral ring and bone dowel) were used. Construct stiffness was evaluated in axial compression, torsion, flexion, and lateral bending. Prior to testing, a silicon elastomer was injected into the cages and intracage pressures were measured using pressure needle transducers. Conclusions. No statistical differences were observed in construct stiffness among the threaded cages and nonthreaded devices in most of the testing modalities. Threaded fusion cages demonstrated significantly lower intracage pressures compared with nonthreaded cages and structural allografts. Compared with nonthreaded cages and structural allografts, threaded fusion cages afforded equivalent reconstruction stiffness but provided more stress-shielded environment within the devices.
引用
收藏
页码:259 / 265
页数:7
相关论文
共 24 条
[1]   INTERBODY LUMBAR FUSION USING A CARBON-FIBER CAGE IMPLANT VERSUS ALLOGRAFT BONE - AN INVESTIGATIONAL STUDY IN THE SPANISH GOAT [J].
BRANTIGAN, JW ;
MCAFEE, PC ;
CUNNINGHAM, BW ;
WANG, H ;
ORBEGOSO, CM .
SPINE, 1994, 19 (13) :1436-1444
[2]   A CARBON-FIBER IMPLANT TO AID INTERBODY LUMBAR FUSION - 2-YEAR CLINICAL-RESULTS IN THE 1ST 26 PATIENTS [J].
BRANTIGAN, JW ;
STEFFEE, AD .
SPINE, 1993, 18 (14) :2106-2117
[3]   Posterior lumbar interbody fusion - A biomechanical comparison, including a new threaded cage [J].
Brodke, DS ;
Dick, JC ;
Kunz, DN ;
McCabe, R ;
Zdeblick, TA .
SPINE, 1997, 22 (01) :26-31
[4]  
Cockin J., 1971, J BONE JOINT SURG BR, V53, P153
[5]   The effect of spinal destabilization and instrumentation on lumbar intradiscal pressure an in vitro biomechanical analysis [J].
Cunningham, BW ;
Kotani, Y ;
McNulty, PS ;
Cappuccino, A ;
McAfee, PC .
SPINE, 1997, 22 (22) :2655-2663
[6]   Osteogenic protein versus autologous interbody arthrodesis in the sheep thoracic spine -: A comparative endoscopic study using the Bagby and Kuslich interbody fusion device [J].
Cunningham, BW ;
Kanayama, M ;
Parker, LM ;
Weis, JC ;
Sefter, JC ;
Fedder, IL ;
McAfee, PC .
SPINE, 1999, 24 (06) :509-518
[7]  
CUNNINGHAM BW, 1998, T ORTHOP RES SOC, V23, P250
[8]   COMPARISON OF DISK SPACE HEIGHTS AFTER ANTERIOR LUMBAR INTERBODY FUSION [J].
DENNIS, S ;
WATKINS, R ;
LANDAKER, S ;
DILLIN, W ;
SPRINGER, D .
SPINE, 1989, 14 (08) :876-878
[9]  
EVANS JH, 1985, CLIN ORTHOP RELAT R, P38
[10]  
GLANTZ SA, 1990, PRIMER APPL REGRESSI, P710