Quantitative evaluation of increase in surface dose by immobilization thermoplastic masks and superficial dosimetry using Gafchromic EBT film and Monte Carlo calculations

被引:5
作者
Ali, Imad [1 ]
Matthiesen, Chance [1 ]
Algan, Ozer [1 ]
Thompson, Spencer [1 ]
Bogardus, Carl [1 ]
Herman, Terence [1 ]
Ahmad, Salahuddin [1 ]
机构
[1] Univ Oklahoma, Hlth Sci Ctr, Dept Radiat Oncol, Oklahoma City, OK 73104 USA
关键词
Skin dose; toxicity; thermoplastic immobilization mask; Gafchromic EBT film; IMRT; Monte Carlo simulation; MODULATED RADIATION-THERAPY; BUILDUP REGION; RADIOCHROMIC FILM; QUALITY-ASSURANCE; PHOTON BEAMS; KODAK EDR2; RADIOTHERAPY; SKIN; VERIFICATION; FIELDS;
D O I
10.3233/XST-2010-0263
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Purpose: To investigate the increase in surface dose under immobilization thermoplastic masks by measurements and calculation in the build-up region using Gafchromic films and Monte Carlo simulation. Materials and methods: Surface doses were measured underneath three thermoplastic masks in open fields using 6 and 18 MV photon beams. These masks are used to immobilize patients for head and neck (H&N), pelvis and thoracic treatment. Gafchromic EBT films were placed on the top of the flat surface of a phantom partially underneath the mask and exposed in open 10 x 10 cm(2) photon fields. The depth doses were calculated using BEAMnrc Monte Carlo code for water-equivalent film detectors with different layers of thickness ranging from 50 mu m to 2.5 mm and compared with film measurements. Results: Surface dose increased by a factor of 3 to 4 underneath the mask relative to the open areas and 6MV beam delivers more skin dose than 18 MV. H&N mask increased surface dose by a factor of 3 using 18 MV and a factor of 4 using 6MV. In addition, increase in surface dose depended on the type of the mask, the size of openings, and the amount of stretching performed during the mask preparation. The measured depth doses were compared with BEAMnrc Monte Carlo calculation for water-equivalent detectors using different sizes. The calculated depth dose depended significantly on the thickness of film detector and varies by more than 15% using layer thickness of 2.5 mm compared to 50 mu m. Surface doses measured by Gafchromic EBT films agreed within 3% with the Monte Carlo calculations using a small detector layer of 50 mu m. Conclusion: Thermoplastic masks used for patient immobilization can significantly increase skin doses by up to a factor of 4 more than that without the mask using 6 MV beams. The skin reactions resulting from thermoplastic masks should be monitored and corrective measures should be taken during treatment such as partially removing the mask over skin areas with complications and optimizing the skin dose in IMRT planning. Gafchromic EBT films provide accurate skin dosimetry which agrees within 3% with Monte Carlo calculations. Gafchromic EBT film makes an excellent tool for measuring depth doses in the buildup region and these data can be applied for treatment planning calculations and IMRT optimization.
引用
收藏
页码:319 / 326
页数:8
相关论文
共 22 条
[1]   Validation of Monte Carlo calculated surface doses for megavoltage photon beams [J].
Abdel-Rahman, W ;
Seuntjens, JP ;
Verhaegen, F ;
Deblois, F ;
Podgorsak, EB .
MEDICAL PHYSICS, 2005, 32 (01) :286-298
[2]   Dependence of radiochromic film optical density post-exposure kinetics on dose and dose fractionation [J].
Ali, I ;
Costescu, C ;
Vicic, M ;
Dempsey, JF ;
Williamson, JF .
MEDICAL PHYSICS, 2003, 30 (08) :1958-1967
[3]   Evaluation of surface and build-up region dose for intensity-modulated radiation therapy in head and neck cancer [J].
Chung, HT ;
Jin, HS ;
Dempsey, JF ;
Liu, CH ;
Palta, J ;
Suh, TS ;
Kim, SY .
MEDICAL PHYSICS, 2005, 32 (08) :2682-2689
[4]   Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system [J].
Court, Laurence E. ;
Tishler, Roy B. ;
Allen, Aaron M. ;
Xiang, Hong ;
Makrigiorgos, Mike ;
Chin, Lee .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2008, 9 (01) :29-35
[5]   Experimental evaluation of the impact of different headand-neck intensity-modulated radiation therapy planning techniques on doses to the skin and shallow targets [J].
Court, Laurence E. ;
Tishler, Roy B. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (02) :607-613
[6]   Accurate skin dose measurements using radiochromic film in clinical applications [J].
Devic, S ;
Seuntjens, J ;
Abdel-Rahman, W ;
Evans, M ;
Olivares, M ;
Podgorsak, EB ;
Vuong, T ;
Soares, CG .
MEDICAL PHYSICS, 2006, 33 (04) :1116-1124
[7]   Surface and build-up region dosimetry for obliquely incident intensity modulated radiotherapy 6 MV x rays [J].
Dogan, N ;
Glasgow, GP .
MEDICAL PHYSICS, 2003, 30 (12) :3091-3096
[8]   Comparative evaluation of Kodak EDR2 and XV2 films for verification of intensity modulated radiation therapy [J].
Dogan, N ;
Leybovich, LB ;
Sethi, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (22) :4121-4130
[9]   Clinical use of EBT model Gafchromic™ film in radiotherapy [J].
Fiandra, Christian ;
Ricardi, Umberto ;
Ragona, Riccardo ;
Anglesio, Silvia ;
Giglioli, Francesca Romana ;
Calamia, Elisa ;
Lucio, Francesco .
MEDICAL PHYSICS, 2006, 33 (11) :4314-4319
[10]   American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning [J].
Fraass, B ;
Doppke, K ;
Hunt, M ;
Kutcher, G ;
Starkschall, G ;
Stern, R ;
Van Dyke, J .
MEDICAL PHYSICS, 1998, 25 (10) :1773-1829