The dosage of small volumes for chromatographic quantifications using a drop-on-demand dispenser system

被引:19
作者
Englmann, Matthias [1 ]
Fekete, Agnes [1 ]
Gebefugi, Istvan [1 ]
Schmitt-Kopplin, Philippe [1 ]
机构
[1] GSF, Natl Res Ctr Environm & Hlth, Inst Ecol Chem, D-85764 Neuherberg, Germany
关键词
piezo-driven drop-on-demand dispenser; ultra performance liquid chromatography; droplets; dispenser; calibration standards; N-acyl homoserines; MALDI-TOF-MS; LESS SAMPLE PREPARATION; FLOW-THROUGH DISPENSER; MASS-SPECTROMETRY; ACOUSTIC LEVITATION; ELECTROPHORESIS; IONIZATION; PICOLITER; PROTEINS; BACTERIA;
D O I
10.1007/s00216-007-1335-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A commercially available piezo-driven drop-on-demand dispenser was tested for its suitability for the preparation of analytical calibration standards and in a standard addition approach prior to quantitative ultra performance liquid chromatography (UPLC) analysis of homoserines. The reproducibility of the drop-on-demand dosing system was tested and the verification of the droplet volume was performed by preparing a series of 1.0 mg/L caffeine standard solutions from a 1,000.0 mg/L stock solution and analysis of the concentrations obtained by UPLC. The reproducibility was better than 1% relative standard deviation from measurement to measurement and the highest was 1.6% from day to day. The results were compared with the conventional way of generating standard solutions (pipetting). A gravimetric method and a photography-based method for the determination of the average single droplet volume were compared and found to be in very good agreement. The system was employed for the quantification of N-decanoyl homoserine by standard addition in bacterial culture supernatants containing this analyte. The agreement with conventional quantification techniques was high. The paper shows the feasibility of the approach with advantages in low sample and solvent volume consumption and very good reproducibility and reliability combined with easy usage.
引用
收藏
页码:1109 / 1116
页数:8
相关论文
共 20 条
[1]  
Allmaier G, 1997, RAPID COMMUN MASS SP, V11, P1567, DOI 10.1002/(SICI)1097-0231(199709)11:14<1567::AID-RCM32>3.0.CO
[2]  
2-0
[3]   Single-pulse nanoelectrospray ionization [J].
Berggren, WT ;
Westphall, MS ;
Smith, LM .
ANALYTICAL CHEMISTRY, 2002, 74 (14) :3443-3448
[4]   Wall-less sample preparation of μm-sized sample spots for femtomole detection limits of proteins from liquid based UV-MALDI matrices [J].
Bogan, MJ ;
Agnes, GR .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2004, 15 (04) :486-495
[5]   MALDI-TOF-MS analysis of droplets prepared in an electrodynamic balance: "Wall-less" sample preparation [J].
Bogan, MJ ;
Agnes, GR .
ANALYTICAL CHEMISTRY, 2002, 74 (03) :489-496
[6]   A novel approach for high throughput cultivation assays and the isolation of planktonic bacteria [J].
Bruns, A ;
Hoffelner, H ;
Overmann, J .
FEMS MICROBIOLOGY ECOLOGY, 2003, 45 (02) :161-171
[7]   Acoustic levitation device for sample pretreatment in microanalysis and trace analysis [J].
Eberhardt, R ;
Neidhart, B .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 365 (06) :475-479
[8]   Signal amplification using "spot on-a-chip" technology for the identification of proteins via MALDI-TOF MS [J].
Ekström, S ;
Ericsson, D ;
Önnerfjord, P ;
Bengtsson, M ;
Nilsson, J ;
Marko-Varga, G ;
Laurell, T .
ANALYTICAL CHEMISTRY, 2001, 73 (02) :214-219
[9]  
ENGLMANN M, 2007, IN PRESS J CHROMAT A
[10]   Analysis of N-acylhomoserine lactones after alkaline hydrolysis and anion-exchange solid-phase extraction by capillary zone electrophoresis-mass spectrometry [J].
Frommberger, M ;
Hertkorn, N ;
Englmann, M ;
Jakoby, S ;
Hartmann, A ;
Kettrup, A ;
Schmitt-Kopplin, P .
ELECTROPHORESIS, 2005, 26 (7-8) :1523-1532