Open-cell design stents in congenital heart disease: A comparison of IntraStent vs. Palmaz stents

被引:28
作者
Kreutzer, J [1 ]
Rome, JJ [1 ]
机构
[1] Univ Penn, Childrens Hosp Philadelphia, Dept Pediat, Div Cardiol, Philadelphia, PA 19104 USA
关键词
cardiac catheterization; congenital heart defects; stent;
D O I
10.1002/ccd.10180
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Slotted stainless Palmaz stents (PS) remain the most commonly used in congenital heart disease (CND). Limitations of PS include rigidity, foreshortening, poor conformability on expansion, balloon rupture, and jailing of side branches. Recently, stents with open-cell design (IntraStent; IS) in appropriate sizes for CHD have been introduced. We reviewed our experience with the IS, comparing performance with the PS in a retrospective nonrandomized, uncontrolled observational study. Between May 1999 and March 2001, 34 IS (10-36 mm) and 34 PS (10-40 mm) were implanted in 57 patients (3 months to 25 years old; median, 3.24 years) in congenital or postoperative lesions. There was no statistically significant difference between the two groups in efficacy, % diameter increase (162% +/- 203% for IS vs. 153% +/- 177% for PS), and % gradient reduction (69% +/- 23% for IS vs. 80% +/- 27% for PS). Other aspects of stent performance differed significantly (P < 0.05): PS forshortened more (mean 18% for PS vs. 0% for IS) and were associated with balloon rupture (9/34 for PS vs. 0/34 for IS; P = 0.002), but had no evidence of intimal protrusion in between struts (0/34 for PS vs. 14/34 for IS; P < 0.001) and no recoil > 15% (0/34 for PS vs. 7/34 for IS; P = 0.006); IS conformed almost twice as well to vessel curvature (P = 0.003). Although these factors did not affect immediate procedural success, balloon rupture in two PS resulted in stent malpositions. Among IS, the origins of three covered side branches were accessed and balloon dilated up to 6 mm through the side of the stent. In conclusion, stents with open-cell design have some characteristics that seem beneficial for their use in CHD: they foreshorten less, are conformable, are less likely to cause balloon rupture, and can allow access to jailed branches. However, they have a higher incidence of significant percent recoil and protrusion of intimal tissue in between struts, which may impact long-term stent performance.
引用
收藏
页码:400 / 409
页数:10
相关论文
共 21 条
  • [1] BALLOON EXPANDABLE STENT IMPLANTATION IN STENOTIC RIGHT HEART VALVED CONDUITS
    ALMAGOR, Y
    PREVOSTI, LG
    BARTORELLI, AL
    KEREN, G
    FERRANS, VJ
    JONES, M
    LEON, MB
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (05) : 1310 - 1314
  • [2] Use of self expanding stents in stenotic aortopulmonary shunts in adults with complex cyanotic heart disease
    Bader, R
    Somerville, J
    Redington, A
    [J]. HEART, 1999, 82 (01) : 27 - 29
  • [3] Early and intermediate-term complications of self-expanding stents limit its potential application in children with congenital heart disease
    Cheung, YF
    Sanatani, S
    Leung, MP
    Human, DG
    Chau, AKT
    Culham, JAG
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (04) : 1007 - 1015
  • [4] BALLOON-EXPANDABLE STENT REPAIR OF SEVERE COARCTATION OF AORTA
    DELEZO, JS
    PAN, M
    ROMERO, M
    MEDINA, A
    SEGURA, J
    PAVLOVIC, D
    MARTINEZ, C
    TEJERO, I
    NAVERO, JP
    TORRES, F
    LAFUENTE, M
    HERNANDEZ, E
    MELIAN, F
    CONCHA, M
    [J]. AMERICAN HEART JOURNAL, 1995, 129 (05) : 1002 - 1008
  • [5] Use of balloon-expandable stents for coarctation of the aorta: Initial results and intermediate-term follow-up
    Ebeid, MR
    Prieto, LR
    Latson, LA
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 30 (07) : 1847 - 1852
  • [6] ENDOVASCULAR STENTS IN THE PULMONARY CIRCULATION - CLINICAL IMPACT ON MANAGEMENT AND MEDIUM-TERM FOLLOW-UP
    FOGELMAN, R
    NYKANEN, D
    SMALLHORN, JF
    MCCRINDLE, BW
    FREEDOM, RM
    BENSON, LN
    [J]. CIRCULATION, 1995, 92 (04) : 881 - 885
  • [7] INTRAVASCULAR STENT PROSTHESIS FOR RIGHT VENTRICULAR OUTFLOW OBSTRUCTION
    HOSKING, MCK
    BENSON, LN
    NAKANISHI, T
    BURROWS, PE
    WILLIAMS, WG
    FREEDOM, RM
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (02) : 373 - 380
  • [8] REPEAT DILATION OF INTRAVASCULAR STENTS IN CONGENITAL HEART-DEFECTS
    ING, FF
    GRIFKA, RG
    NIHILL, MR
    MULLINS, CE
    [J]. CIRCULATION, 1995, 92 (04) : 893 - 897
  • [9] Kreutzer J, 2000, DIAGNOSTIC INTERVENT, V2nd, P221
  • [10] Early results and medium-term follow-up of stent implantation for mild residual or recurrent aortic coarctation
    Marshall, AC
    Perry, SB
    Keane, JF
    Lock, JE
    [J]. AMERICAN HEART JOURNAL, 2000, 139 (06) : 1054 - 1060