A trucut biopsy needle for bipolar radiofrequency ablation of needle tract: a proof-of-concept experiment

被引:5
作者
Bruners, Philipp [1 ,2 ]
Penzkofer, Tobias [1 ,2 ]
Isfort, Peter [1 ,2 ]
Pfeffer, Jochen [2 ]
Schmitz-Rode, Thomas [2 ]
Guenther, Rolf W. [1 ]
Mahnken, Andreas H. [1 ]
机构
[1] Rhein Westfal TH Aachen, Univ Hosp, Dept Diagnost Radiol, D-52074 Aachen, Germany
[2] Rhein Westfal TH Aachen, Helmholtz Inst Biomed Engn, D-52074 Aachen, Germany
关键词
Biopsy; Radiofrequency ablation; Trucut; Bleeding; Tumour cell seeding; HEPATOCELLULAR-CARCINOMA; LIVER-BIOPSY; IMPLANTATION; COMPLICATIONS; COAGULATION; EXPERIENCE; MODEL; RISK;
D O I
10.1007/s00330-010-1739-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To develop a trucut biopsy needle featuring two electrodes that allow for bipolar radiofrequency (RF) coagulation of the puncture tract. We modified a 14-G trucut biopsy needle to contain two insulated electrodes and connected the device to an RF generator. Biopsies in ex vivo porcine liver and kidney were performed. The puncture tract was coagulated by using different RF energy settings (5 W, 10 W, 20 W). Tissue specimens were dissected along the puncture tract and the coagulation area was macroscopically evaluated. CT-guided in vivo liver and kidney biopsies were performed in two domestic pigs. Lengths of specimens were measured. Post-biopsy contrast-enhanced CT examinations were performed to rule out biopsy-related bleeding. Animals were euthanised and coagulation areas macroscopically explored. The mean diameters of the coagulated area around the ex vivo biopsy tract were 4.2 +/- 1.1 mm (5 W), 6.0 +/- 2.0 mm (10 W) and 5.2 +/- 0.51 mm (20 W) in liver and 5.0 +/- 0.7 mm (5 W), 6.6 +/- 0.9 (10 W) and 6.0 +/- 2.0 mm (20 W) in kidney. After biopsies CT revealed no bleeding. Mean maximum coagulation diameters were 10.1 +/- 4.6 mm (10 W) in liver and 6.0 +/- 2.5 mm (10 W) in kidney. Mean length of the specimens was 12.2 +/- 4.4 mm in kidney and 11.1 +/- 3.6 mm in liver tissue. Bipolar RF biopsy is a promising tool for tract coagulation after percutaneous biopsy.
引用
收藏
页码:2000 / 2004
页数:5
相关论文
共 13 条
[1]   Bipolar Radiofrequency ablation: Is the shape of the coagulation volume different in comparison to monopolar RF-ablation using variable active tip lengths? [J].
Bruners, Philipp ;
Lipka, Joerg ;
Guenther, Rolf W. ;
Schmitz-Rode, Thomas ;
Mahnken, Andreas H. .
MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2008, 17 (05) :267-274
[2]   Needle tract implantation after percutaneous interventional procedures in hepatocellular carcinomas: Lessons learned from a 10-year experience [J].
Chang, Samuel ;
Kim, Seong Hyun ;
Lim, Hyo K. ;
Kim, Seung Hoon ;
Lee, Won Jae ;
Choi, Dongil ;
Kim, Young-Sun ;
Rhim, Hyunchul .
KOREAN JOURNAL OF RADIOLOGY, 2008, 9 (03) :268-274
[3]   ELECTROCAUTERY OF THE TRACT AFTER NEEDLE-BIOPSY OF THE LIVER TO REDUCE BLOOD-LOSS - EXPERIENCE IN THE CANINE MODEL [J].
KIM, EH ;
KOPECKY, KK ;
CUMMINGS, OW ;
DREESEN, RG ;
POUND, DC .
INVESTIGATIVE RADIOLOGY, 1993, 28 (03) :228-230
[4]   Needle-tract implantation in hepatocellular carcinoma: frequency and CT findings after biopsy with a 19.5-gauge automated biopsy gun [J].
Kim, SH ;
Lim, HK ;
Lee, WJ ;
Cho, JM ;
Jang, HJ .
ABDOMINAL IMAGING, 2000, 25 (03) :246-250
[5]   Postbiopsy bleeding in a porcine model: Reduction with radio-frequency ablation preliminary results [J].
Laeseke, PF ;
Winter, TC ;
Davis, CL ;
Stevens, KR ;
Johnson, CD ;
Fronczak, FJ ;
Webster, JG ;
Lee, FT .
RADIOLOGY, 2003, 227 (02) :493-499
[6]   Risk of tumour seeding after percutaneous radiofrequency ablation for hepatocellular carcinoma [J].
Livraghi, T ;
Lazzaroni, S ;
Meloni, F ;
Solbiati, L .
BRITISH JOURNAL OF SURGERY, 2005, 92 (07) :856-858
[7]   Increased risk of tumor seeding after percutaneous radiofrequency ablation for single hepatocellular carcinoma [J].
Llovet, JM ;
Vilana, R ;
Brú, C ;
Bianchi, L ;
Salmeron, JM ;
Boix, L ;
Ganau, S ;
Sala, M ;
Pagès, M ;
Ayuso, C ;
Solé, M ;
Rodés, J ;
Bruix, R .
HEPATOLOGY, 2001, 33 (05) :1124-1129
[8]   Complications of radiofrequency coagulation of liver tumours [J].
Mulier, S ;
Mulier, P ;
Ni, Y ;
Miao, Y ;
Dupas, B ;
Marchal, G ;
De Wever, I ;
Michel, L .
BRITISH JOURNAL OF SURGERY, 2002, 89 (10) :1206-1222
[9]   COMPLICATIONS FOLLOWING PERCUTANEOUS LIVER-BIOPSY - A MULTICENTER RETROSPECTIVE STUDY ON 68 276 BIOPSIES [J].
PICCININO, F ;
SAGNELLI, E ;
PASQUALE, G ;
GIUSTI, G .
JOURNAL OF HEPATOLOGY, 1986, 2 (02) :165-173
[10]   Radiofrequency cauterization with biopsy introducer needle [J].
Pritchard, WF ;
Wray-Cahen, D ;
Karanian, JW ;
Hilbert, S ;
Wood, BJ .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 15 (02) :183-187