The effects of pulse rate, artefact and pulse strength on oscillometric non-invasive blood pressure measurements.

被引:0
|
作者
Amoore, JN [1 ]
Geake, WB [1 ]
Scott, DHT [1 ]
机构
[1] Royal Infirm, Dept Med Phys & Med Engn, Edinburgh EH3 9YW, Midlothian, Scotland
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We examined the effects of pulse rate, pulse and artefact on 20 adult oscillometric non-invasive blood pressure measurements using NIBP test simulators. Over a range of pulse rates from 40 to 200 bpm the 20 NIBP monitors studied recorded little change in registered pressures. The monitors achieved consistent readings at low pulse rates by decreasing the deflation rate and hence increasing the determination time (from under 20 sec at 200 bpm to over 30 sec at 40 bpm). High levels of artefact produced artificially high systolic pressures in 12 of the 16 monitors that recorded pressures, 6 of them being over 30 mmHg above the target 120 mmHg; 4 of the monitors signalled failure to record due to excessive artefact. Pulse strengths of 25% of the nominal were coped with well by most monitors with a slight tendency for the monitors to record low systolic pressures.
引用
收藏
页码:1977 / 1978
页数:2
相关论文
共 50 条
  • [21] Occlusion spectroscopy as a new paradigm for non-invasive blood measurements.
    Fine, I
    Fikhte, B
    Shvartsman, LD
    OPTICAL DIAGNOSTICS AND SENSING OF BIOLOGICAL FLUIDS AND GLUCOSE AND CHOLESTEROL MONITORING, 2001, 4263 : 122 - 130
  • [22] NON-INVASIVE, CONTINUOUS, PULSE PRESSURE MONITORING METHOD
    Kuwahara, Makiko
    Yavari, Ehsan
    Boric-Lubecke, Olga
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 6574 - 6577
  • [23] Comparing invasive and oscillometric blood pressure measurements
    Fahd, G.
    Mokhtar, O. Ait
    Boiron, O.
    Morgado, I.
    Paganelli, F.
    Deplano, V.
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2011, 14 : 149 - 151
  • [24] Non-invasive measurements of pulse pressure variation and stroke volume variation in anesthetized patients using the Nexfin blood pressure monitor
    Stens, Jurre
    Oeben, Jeroen
    Van Dusseldorp, Ab A.
    Boer, Christa
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2016, 30 (05) : 587 - 594
  • [25] A comparison of invasive arterial blood pressure measurement with oscillometric non-invasive blood pressure measurement in patients with sepsis
    Jiang, Ziqing
    Li, Shaoying
    Wang, Lin
    Yu, Feng
    Zeng, Yanping
    Li, Hongbo
    Li, Jun
    Zhang, Zhanfeng
    Zuo, Junling
    JOURNAL OF ANESTHESIA, 2024, 38 (02) : 222 - 231
  • [26] A comparison of invasive arterial blood pressure measurement with oscillometric non-invasive blood pressure measurement in patients with sepsis
    Ziqing Jiang
    Shaoying Li
    Lin Wang
    Feng Yu
    Yanping Zeng
    Hongbo Li
    Jun Li
    Zhanfeng Zhang
    Junling Zuo
    Journal of Anesthesia, 2024, 38 : 222 - 231
  • [27] Non-invasive measurements of pulse pressure variation and stroke volume variation in anesthetized patients using the Nexfin blood pressure monitor
    Jurre Stens
    Jeroen Oeben
    Ab A. Van Dusseldorp
    Christa Boer
    Journal of Clinical Monitoring and Computing, 2016, 30 : 587 - 594
  • [28] High Fidelity Pressure Wires Provide Accurate Validation of Non-Invasive Central Blood Pressure and Pulse Wave Velocity Measurements
    Scalia, Alessandro
    Ghafari, Chadi
    Navarre, Wivine
    Delmotte, Philippe
    Phillips, Rob
    Carlier, Stephane
    BIOMEDICINES, 2023, 11 (04)
  • [29] Non-invasive blood pressure monitoring in dialysis patients using pulse wave velocity
    Koball, Sebastian
    Hermann, Hans
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2024, 39 : I2428 - I2428
  • [30] A comparison of blood pressure and pulse pressure values obtained by oscillometric and central measurements in hypertensive patients
    Kayrak, Mehmet
    Ulgen, Mehmet S.
    Yazici, Mehmet
    Yilmaz, Remzi
    Demir, Kenan
    Dogan, Yildiz
    Ozhan, Hakan
    Alihanoglu, Yusuf
    Koc, Fatih
    Bodur, Sait
    BLOOD PRESSURE, 2010, 19 (02) : 98 - 103