3D real time methodology monitoring cement failures in THA

被引:23
作者
Qi, G
Li, JH
Mann, KA
Mouchon, WP
Hamstad, MA
Salehi, A
Whitten, SA
机构
[1] Univ Memphis, Dept Mech Engn, Med Acoust Res Lab, Memphis, TN 38152 USA
[2] Univ Memphis, Dept Biomed Engn, Joint Program, Memphis, TN 38152 USA
[3] Univ Tennessee, Ctr Hlth Sci, Memphis, TN 38152 USA
[4] Upstate Med Univ, Dept Orthoped Surg, Syracuse, NY 13210 USA
[5] Univ Denver, Dept Engn, Denver, CO 80208 USA
[6] Smith Nephew Inc, Memphis, TN 38116 USA
关键词
acoustic emission; microcrack source location; cemented total hip arthroplasty; fatigue of cemented hip implant;
D O I
10.1002/jbm.a.30133
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The present work proposed a methodology to monitor cement microcrack formation in the cemented femoral stem construct using the acoustic emission technique. This technique provides a unique means to automatically tally the number of microcracks, to visualize microcrack distribution, and to animate the progress of crack formation in a given time window of a fatigue test. In this work, the formulation of microcrack source location was derived and a computer program was developed specifically for the proposed application. The program was validated using computer simulation and standard pencil lead break tests. It was found that the mathematical errors complied with the acceptable minimal error. Based on the pencil lead break tests, the average technical error used to estimate the resolution of this technique was 4.7 mm at the present stage. The program was then used to monitor the fatigue damage in precoated cemented femoral hip constructs loaded for a total of more than five million cycles. Two types of microcrack activities were observed in the experiments: Type I and Type II microcracks. A Type I microcrack was a crack that was captured by four or more sensors, and therefore its location was defined uniquely by a set of coordinates. A Type 11 microcrack was a crack that was captured by three or less sensors, therefore it was unlocatable. Both counts of Type I and Type 11 microcrack were tallied with respect to the day of fatigue tests. Acoustic emission microcrack graphs were used to visualize the distribution of Type I microcracks in the construct. It was found that the Type I microcracks distributed mainly over the proximal third of the stem. The amount of microcrack events decreased significantly as the number of loading cycles increased. (C) 2004 Wiley Periodicals, Inc.
引用
收藏
页码:391 / 402
页数:12
相关论文
共 36 条
[1]  
[Anonymous], 1979, CLIN ORTHOP RELAT RE
[2]   Characterization and control of selective corrosion of α,β′-brass by acoustic emission [J].
Assouli, B ;
Srhiri, A ;
Idrissi, H .
NDT & E INTERNATIONAL, 2003, 36 (02) :117-126
[3]  
*ASTM, 1994, E976 ASTM
[4]   Endurance testing of hip prostheses: a comparison between the load fixed in ISO 7206 standard and the physiological loads [J].
Baleani, M ;
Cristofolini, L ;
Viceconti, M .
CLINICAL BIOMECHANICS, 1999, 14 (05) :339-345
[5]   Detection of microfracture processes in composite laminates by thermo-acoustic emission [J].
Choi, NS ;
Kim, YB ;
Kim, TW ;
Rhee, KY .
JOURNAL OF MATERIALS SCIENCE, 2003, 38 (05) :1013-1019
[6]   Monitoring the integrity of the cement-metal interface of total joint components in vitro using acoustic emission and ultrasound [J].
Davies, JP ;
Tse, MK ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 1996, 11 (05) :594-601
[7]   IN-VITRO EVALUATION OF BONDING OF THE CEMENT-METAL INTERFACE OF A TOTAL HIP FEMORAL COMPONENT USING ULTRASOUND [J].
DAVIES, JP ;
TSE, MK ;
HARRIS, WH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1995, 13 (03) :335-338
[8]  
DOWNS KS, 1996, J ACOUST EMISS, V3, pS61
[9]  
Hamstad M A, 1995, Journal of Acoustic Emission, V13, P31
[10]   A 3-DIMENSIONAL NONLINEAR FINITE-ELEMENT STUDY OF THE EFFECT OF CEMENT-PROSTHESIS DEBONDING IN CEMENTED FEMORAL TOTAL HIP COMPONENTS [J].
HARRIGAN, TP ;
HARRIS, WH .
JOURNAL OF BIOMECHANICS, 1991, 24 (11) :1047-1058