Lupus anticoagulant assay cut-offs vary between reagents even when derived from a common set of normal donor plasmas

被引:14
作者
Moore, Gary W. [1 ]
Kumano, Osamu [2 ]
机构
[1] Guys & St Thomas Hosp, Dept Haemostasis & Thrombosis, Viapath Analyt, London, England
[2] Sysmex Corp, Prot Technol, Kobe, Hyogo, Japan
关键词
activated partial thromboplastin time; antiphospholipid syndrome; dilute Russell's viper venom time; lupus anticoagulant; phospholipid dependence; GUIDELINES; TESTS; SENSITIVITY; MANAGEMENT; UPDATE;
D O I
10.1111/jth.14669
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Multicenter studies reveal that diagnostic efficacy of lupus anticoagulant (LA) assays is enhanced if cut-offs are locally generated. However, a potential confounder is the inevitable use of separate normal donor populations. Objectives Generate cut-offs for multiple LA reagents with the same analyzer and normal donor plasmas. Methods Cut-offs for screen ratio, confirm ratio, percent correction of screen ratio by confirm ratio, and normalized screen/confirm ratio (NSCR) were derived from the same 50 normal donor plasmas for screen and confirm pairs for two dilute Russell's viper venom time reagents, LA1/LA2 and HEMOCLOT (TM) LA-S/LA-C, and two APTTs, Actin FSL/FS and Cephen LS/Cephen. The cut-offs were challenged with plasmas from 20 triple-positive APS patients and 25 plasmas from LA-negative, thrombotic patients. Results Cut-offs for screen ratio, confirm ratio, percent correction, and NSCR, respectively, were 1.12/1.08/8.3/1.09 for LA1/LA2; 1.17/1.10/13.6/1.13 for HEMOCLOT (TM) LA-S/LA-C; 1.12/1.13/9.7/1.10 for Actin FSL/FS; 1.09/1.13/11.0/1.11 for Cephen LS/Cephen. LA1 and LA-S screens were elevated in 19/20 and 16/20 triple-positive plasmas, respectively, while 20/20 were detected with both via integrated interpretation ie, percent correction or NSCR irrespective of screen elevation. Actin FSL and Cephen LS screens were elevated in 17/20 and 19/20 triple-positive plasmas, respectively, while one more LA was detected with Actin FSL via integrated interpretation, but not for Cephen LS. Integrated interpretation suggested 5/25 LA-negative plasmas contained weak LA (two with Actin FSL/FS, two with LA1/LA2, one with LA-S/LA-R). Conclusions Employing the same normal donor plasmas and analytical platform does not compensate for between-reagent differences when generating LA assay cut-offs.
引用
收藏
页码:439 / 444
页数:6
相关论文
共 25 条
[1]  
[Anonymous], 2008, DEFINING ESTABLISHIN
[2]  
Clinical and Laboratory Standards Institute (CLSI), 2014, CLSI document H60-A
[3]   Clinical and laboratory practice for lupus anticoagulant testing: An International Society of Thrombosis and Haemostasis Scientific and Standardization Committee survey [J].
Cohen, Hannah ;
Mackie, Ian J. ;
Devreese, Katrien M. J. .
JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2019, 17 (10) :1715-1732
[4]   Laboratory criteria for antiphospholipid syndrome: communication from the SSC of the ISTH [J].
Devreese, K. M. J. ;
Ortel, T. L. ;
Pengo, V. ;
de Laat, B. .
JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2018, 16 (04) :809-813
[5]   The importance of locally derived reference ranges and standardized calculation of dilute Russell's viper venom time results in screening for lupus anticoagulant [J].
Gardiner, C ;
Mackie, IJ ;
Malia, RG ;
Jones, DW ;
Winter, M ;
Leeming, D ;
Taberner, DA ;
Machin, SJ ;
Greaves, M .
BRITISH JOURNAL OF HAEMATOLOGY, 2000, 111 (04) :1230-1235
[6]   Reference intervals of the dilute tissue thromboplastin inhibition and dilute Russell's viper venom tests revisited [J].
Gerbutavicius, R ;
Fareed, J ;
Messmore, HL ;
Iqbal, O ;
Hoppensteadt, DA ;
Wehrmacher, WH ;
Demir, M ;
Piccolo, P ;
Ahmad, S ;
Ma, Q ;
Griniute, R .
CLINICAL AND APPLIED THROMBOSIS-HEMOSTASIS, 2002, 8 (02) :115-124
[7]   Guidelines on the investigation and management of the antiphospholipid syndrome [J].
Greaves, M ;
Cohen, H ;
Machin, SJ ;
Mackie, I .
BRITISH JOURNAL OF HAEMATOLOGY, 2000, 109 (04) :704-715
[8]   Achieving and maintaining quality in the laboratory [J].
Hertzberg, MS ;
Mammen, J ;
McCraw, A ;
Nair, SC ;
Srivastava, A .
HAEMOPHILIA, 2006, 12 :61-67
[9]   False negative screening tests for lupus anticoagulants - An unrecognized problem? [J].
Jacobsen, EM ;
Wisloff, F .
THROMBOSIS RESEARCH, 1996, 82 (05) :445-451
[10]   Guidelines on the investigation and management of antiphospholipid syndrome [J].
Keeling, David ;
Mackie, Ian ;
Moore, Gary W. ;
Greer, Ian A. ;
Greaves, Michael .
BRITISH JOURNAL OF HAEMATOLOGY, 2012, 157 (01) :47-58