Tandem Ion Mobility Spectrometry for the Detection of Traces of Explosives in Cargo at Concentrations of Parts Per Quadrillion

被引:35
作者
Amo-Gonzalez, Mario [1 ]
Perez, Sergio [1 ]
Delgado, Rafael [1 ]
Arranz, Gonzalo [1 ]
Carnicero, Irene [1 ]
机构
[1] SEADM, Parque Tecnol Boecillo 205, Valladolid, Spain
关键词
AMBIENT-PRESSURE; IMS-IMS; AIR; DECOMPOSITION; DISSOCIATION; KINETICS; FIELD;
D O I
10.1021/acs.analchem.9b03589
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A tandem ion mobility spectrometer (IMS2) built from two differential mobility analyzers (DMAs) is coupled at ambient pressure with a thermal fragmenter placed in between, such that the precursor ions selected in the first DMA are thermally decomposed at ambient pressure in the fragmenter and the product ions generated are filtered in the second DMA. A thermal desorber and a multicapillary gas chromatography (GC) column are coupled to a secondary electrospray (SESI) ion source, so the adsorption sampling filters are thermally desorbed and the liberated vapors are separated in the GC column, prior to their ionization and mobility/mobility classification. The new fragmenter allows the fragmentation of the five explosives studied: RDX, PETN, NG, EGDN, and TNT. The background of the analyzer is evaluated for the five explosives using air samples of 500 L volume. An atmospheric background of only 2.5 pg (5 ppq) is found for TNT, being somewhat higher for the rest of explosives studied. The architecture GC-IMS2 is compared with GC-IMS obtaining a 100-fold increase of sensitivity in the first configuration, confirming the high selectivity provided by the fragmentation cell and the second IMS stage for the product ion mobility analysis. The analyzer is tested also with real explosives hidden in cargo pallets achieving successful detection of four (EGDN, NG, TNT, and PETN) out of five explosives.
引用
收藏
页码:14009 / 14018
页数:10
相关论文
共 31 条
  • [1] Planar Differential Mobility Analyzer with a Resolving Power of 110
    Amo-Gonzalez, Mario
    Perez, Sergio
    [J]. ANALYTICAL CHEMISTRY, 2018, 90 (11) : 6735 - 6741
  • [2] Mobility Peak Tailing Reduction in a Differential Mobility Analyzer (DMA) Coupled with a Mass Spectrometer and Several Ionization Sources
    Amo-Gonzalez, Mario
    de la Mora, Juan Fernandez
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2017, 28 (08) : 1506 - 1517
  • [3] AMOGONZALEZ M, 2018, J ANAL CHEM+, V90, P6885, DOI DOI 10.1021/ACS.ANALCHEM.8B01086
  • [4] Gas phase fragmentation of protonated esters in air at ambient pressure through ion heating by electric field in differential mobility spectrometry
    An, X.
    Eiceman, G. A.
    Rodriguez, J. E.
    Stone, J. A.
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2011, 303 (2-3) : 181 - 190
  • [5] A determination of the effective temperatures for the dissociation of the proton bound dimer of dimethyl methylphosphonate in a planar differential mobility spectrometer
    An, Xinxia
    Eiceman, Gary A.
    Stone, John A.
    [J]. INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2010, 13 (01) : 25 - 36
  • [6] Flow driven transmission of charged particles against an axial field in antistatic tubes at the sample outlet of a Differential Mobility Analyzer
    Attoui, Michel
    de la Mora, Juan Fernandez
    [J]. JOURNAL OF AEROSOL SCIENCE, 2016, 100 : 91 - 96
  • [7] EVALUATION OF TENAX-GR ADSORBENT FOR THE PASSIVE SAMPLING OF VOLATILE ORGANIC-COMPOUNDS AT LOW CONCENTRATIONS
    CAO, XL
    HEWITT, CN
    [J]. ATMOSPHERIC ENVIRONMENT PART A-GENERAL TOPICS, 1993, 27 (12): : 1865 - 1872
  • [8] Tandem ion mobility spectrometry at ambient pressure and field decomposition of mobility selected ions of explosives and interferences Electronic supplementary information (ESI) available. See DOI: 10.1039/c8an02041h
    Chiluwal, Umesh
    Lee, Gyoungil
    Rajapakse, Maneeshin Y.
    Willy, Timothy
    Lukow, Stefan
    Schmidt, Hartwig
    Eiceman, Gary A.
    [J]. ANALYST, 2019, 144 (06) : 2052 - 2061
  • [9] Fast high-resolution nanoDMA measurements with a 25ms response time electrometer
    de la Mora, Juan Fernandez
    Perez-Lorenzo, Luis Javier
    Arranz, Gonzalo
    Amo-Gonzalez, Mario
    Burtscher, Heinz
    [J]. AEROSOL SCIENCE AND TECHNOLOGY, 2017, 51 (06) : 724 - 734
  • [10] FRAGMENTATION OF BUTYL ACETATE ISOMERS IN THE DRIFT REGION OF AN ION MOBILITY SPECTROMETER
    EICEMAN, GA
    SHOFF, DB
    HARDEN, CS
    SNYDER, AP