Favourable alignment outcomes with MRI-based patient-specific instruments in total knee arthroplasty

被引:28
作者
Schotanus, Martijn G. M. [1 ]
Thijs, Elke [1 ]
Heijmans, Marion [2 ]
Vos, Rein [3 ]
Kort, Nanne P. [1 ]
机构
[1] Zuyderland Med Ctr, Dr H vd Hoffpl 1, NL-6162 AG Sittard Geleen, Netherlands
[2] Zuyderland Med Ctr, Zuyderland Acad, Sittard Geleen, Netherlands
[3] Maastricht Univ, Med Ctr, Dept Methodol & Stat, Maastricht, Netherlands
关键词
Patient-specific instruments; PSI; Conventional instruments; Total knee arthroplasty; TKA; MRI; CT; Long-leg standing radiograph; Alignment; Outliers; Meta-analysis; Systematic literature review; CUTTING BLOCKS; NO DIFFERENCE; COMPUTED-TOMOGRAPHY; COMPONENT ALIGNMENT; CT; GUIDES; ACCURACY; TKA; MULTICENTER; ROTATION;
D O I
10.1007/s00167-017-4637-0
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose Patient-specific instruments (PSIs) are already in relatively common use, and their post-operative radiographic results are equal to those for total knee arthroplasty (TKA) with conventional instrumentation. PSI use requires a preoperative MRI scan, CT scan, or a combination of MRI and a long-leg standing radiograph. However, there is no consensus as to which of these modalities, MRI or CT, is the preferred imaging modality when performing TKA with PSIs. Methods This systematic literature review and meta-analysis studied the differences in alignment outliers between CT- and MRI-based PSI for TKA. A search of the Cochrane Database of Systematic Reviews, MED-LINE/PubMed and Embase was conducted, without restriction on date of publication. Only level I evidence studies written in English that included TKA with the use of MRI- and CT-based PSI were selected. A meta-analysis was then performed of the rate of outliers in the bio-mechanical axis and individual femoral and tibial component alignment. Where considerable heterogeneity among studies was present or the data did not provide sufficient information for performing the meta-analysis, a qualitative synthesis was undertaken. Results Twelve randomized controlled trials, studying 841 knees, were eligible for data extraction and meta-analysis. MRI-based PSI resulted in a significantly lower proportion of coronal plane outliers with regard to the lateral femoral component (OR 0.52, 95% CI 0.30-0.89, P = 0.02), without significant heterogeneity (n.s.). There were no significant differences regarding the biomechanical axis or frontal femoral and individual tibial component alignment. Conclusion This systematic review and meta-analysis demonstrate that alignment with MRI-based PSI is at least as good as, if not better than, that with CT-based PSI. To prevent for malalignment, MRI should be the imaging modality of choice when performing TKA surgery with PSI.
引用
收藏
页码:2659 / 2668
页数:10
相关论文
共 43 条
  • [1] Accuracy of MRI-basedvs. CT-based patient-specific instrumentation in total knee arthroplasty: A meta-analysis
    An, Vincent V. G.
    Sivakumar, Brahman S.
    Phan, Kevin
    Levy, Yadin David
    Bruce, Warwick J. M.
    [J]. JOURNAL OF ORTHOPAEDIC SCIENCE, 2017, 22 (01) : 116 - 120
  • [2] [Anonymous], 2009, Introd. Meta-Anal., DOI DOI 10.1002/9780470743386.CH7
  • [3] [Anonymous], 2010, SURGERY, DOI DOI 10.1016/J.SURG.2009.06.030
  • [4] Patient specific cutting blocks are currently of no proven value
    Barrack, R. L.
    Ruh, E. L.
    Williams, B. M.
    Ford, A. D.
    Foreman, K.
    Nunley, R. M.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2012, 94B (11): : 95 - 99
  • [5] No difference in clinical outcome between patient-matched positioning guides and conventional instrumented total knee arthroplasty two years post-operatively A MULTICENTRE, DOUBLE-BLIND, RANDOMISED CONTROLLED TRIAL
    Boonen, B.
    Schotanus, M. G. M.
    Kerens, B.
    van der Weegen, W.
    Hoekstra, H. J.
    Kort, N. P.
    [J]. BONE & JOINT JOURNAL, 2016, 98B (07) : 939 - 944
  • [6] Inter-observer reliability of measurements performed on digital long-leg standing radiographs and assessment of validity compared to 3D CT-scan
    Boonen, B.
    Kerens, B.
    Schotanus, M. G. M.
    Emans, P.
    Jong, B.
    Kort, N. P.
    [J]. KNEE, 2016, 23 (01) : 20 - 24
  • [7] Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial
    Boonen, B.
    Schotanus, M. G. M.
    Kerens, B.
    van der Weegen, W.
    van Drumpt, R. A. M.
    Kort, N. P.
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2013, 21 (10) : 2206 - 2212
  • [8] A prospective randomised controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement
    Chareancholvanich, K.
    Narkbunnam, R.
    Pornrattanamaneewong, C.
    [J]. BONE & JOINT JOURNAL, 2013, 95B (03) : 354 - 359
  • [9] Analysis of errors in medical rapid prototyping models
    Choi, JY
    Choi, JH
    Kim, NK
    Kim, Y
    Lee, JK
    Kim, MK
    Lee, JH
    Kim, MJ
    [J]. INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2002, 31 (01) : 23 - 32
  • [10] The accuracy of component alignment in custom cutting blocks compared with, conventional total knee arthroplasty instrumentation: Prospective control trial
    Chotanaphuti, Thanainit
    Wangwittayakul, Visit
    Khuangsirikul, Saradej
    Foojareonyos, Trakul
    [J]. KNEE, 2014, 21 (01) : 185 - 188