Comparison of pure tone audiometry and auditory steady-state responses in subjects with normal hearing and hearing loss

被引:26
作者
Ozdek, Ali [1 ,2 ]
Karacay, Mahmut [2 ]
Saylam, Guleser [2 ]
Tatar, Emel [2 ]
Aygener, Nurdan [3 ]
Korkmaz, Mehmet Hakan [2 ]
机构
[1] SB Ankara Diskapi Egitim & Arastirma Hastanesi, KBB Klin, TR-06510 Ankara, Turkey
[2] Ankara Diskapi Yildirim Beyazit Res & Training Ho, ENT Dept, Minist Hlth, Ankara, Turkey
[3] Ankara Diskapi Yildirim Beyazit Res & Training Ho, Dept Audiol, Minist Hlth, Ankara, Turkey
关键词
Pure tone audiometry; Auditory steady-state response; Behavioral threshold; ASSR threshold; BRAIN-STEM RESPONSE; THRESHOLDS; CHILDREN;
D O I
10.1007/s00405-009-1014-8
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
The objective of this study is to compare pure tone audiometry and auditory steady-state response (ASSR) thresholds in normal hearing (NH) subjects and subjects with hearing loss. This study involved 23 NH adults and 38 adults with hearing loss (HI). After detection of behavioral thresholds (BHT) with pure tone audiometry, each subject was tested for ASSR responses in the same day. Only one ear was tested for each subject. The mean pure tone average was 9 +/- A 4 dB for NH group and 57 +/- A 14 for HI group. There was a very strong correlation between BHT and ASSR measurements in HI group. However, the correlation was weaker in the NH group. The mean differences of pure tone average of four frequencies (0.5, 1, 2, and 4 kHz) and ASSR threshold average of same frequencies were 13 +/- A 6 dB in NH group and 7 +/- A 5 dB in HI group and the difference was significant (P = 0.01). It was found that 86% of threshold difference values were less than 20 dB in NH group and 92% of threshold difference values were less than 20 dB in HI group. In conclusion, ASSR thresholds can be used to predict the configuration of pure tone audiometry. Results are more accurate in HI group than NH group. Although ASSR can be used in cochlear implant decision-making process, findings do not permit the utilization of the test for medico-legal reasons.
引用
收藏
页码:43 / 49
页数:7
相关论文
共 23 条
[1]   Comparing pure-tone audiometry and auditory steady state response for the measurement of hearing loss [J].
Ahn, Joong Ho ;
Lee, Hyo-Sook ;
Kim, Young-Jin ;
Yoon, Tae Hyun ;
Chung, Jong Woo .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2007, 136 (06) :966-971
[2]  
Arlinger S, 2000, SEMIN HEAR, V21, P370, DOI DOI 10.1055/S-2000-13469
[3]   Steady-state response audiometry in a group of patients with steeply sloping sensorineural hearing loss [J].
Ballay, C ;
Tonini, R ;
Waninger, T ;
Yoon, C ;
Manolidis, S .
LARYNGOSCOPE, 2005, 115 (07) :1243-1246
[4]  
BROOKHOUSER PE, 1990, LARYNGOSCOPE, V100, P803
[5]   Auditory steady-state responses and clinical applications [J].
Canale, Andrea ;
Lacilla, Michelangelo ;
Cavalot, Andrea Luigi ;
Albera, Roberto .
EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2006, 263 (06) :499-503
[7]   Auditory steady-state response thresholds of adults with sensorineural hearing impairments [J].
Herdman, AT ;
Stapells, DR .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2003, 42 (05) :237-248
[8]   Threshold prediction using the auditory steady-state response and the tone burst auditory brain stem response: A within-subject comparison [J].
Johnson, TA ;
Brown, CJ .
EAR AND HEARING, 2005, 26 (06) :559-576
[9]   Frequency-specific audiometry using steady-state responses [J].
Lins, OG ;
Picton, TW ;
Boucher, BL ;
DurieuxSmith, A ;
Champagne, SC ;
Moran, LM ;
PerezAbalo, MC ;
Martin, V ;
Savio, G .
EAR AND HEARING, 1996, 17 (02) :81-96
[10]   Comparison of MASTER and AUDERA for measurement of auditory steady-state responses [J].
Luts, H ;
Wouters, J .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2005, 44 (04) :244-253