Optical method of sub-cooled water recognition in dew point hygrometer

被引:6
作者
Weremczuk, J [1 ]
机构
[1] Warsaw Univ Technol, Inst Elect Syst, PL-00665 Warsaw, Poland
来源
IMTC/2000: PROCEEDINGS OF THE 17TH IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE: SMART CONNECTIVITY: INTEGRATING MEASUREMENT AND CONTROL | 2000年
关键词
D O I
10.1109/IMTC.2000.846802
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A new optical method of sub-cooled water recognition in dew point hygrometer is presented in this paper. The physical phenomenon of sub-cooled water and its influence on the hygrometer accuracy is described. A hypothesis relating scattering of light by water and ice layers is proposed. A new type of dew point hygrometer based on the silicone mirror with the embedded capacitive detector and multi-section optical detector is presented. Measurement data obtained from the experimental hygrometer are provided.
引用
收藏
页码:2 / 7
页数:6
相关论文
共 4 条
  • [1] BANNISTER M, 1998, 3 INT S HUM MOIST LO
  • [2] JACHOWICZ R, 1992, TC4 IMEKO
  • [3] JACHOWICZ R, 2000, IN PRESS SENSORS A A
  • [4] JACHOWICZ R, 1999, 13 EUR C SOL STAT TR