The effectiveness of pulse oximetry sonification enhanced with tremolo and brightness for distinguishing clinically important oxygen saturation ranges: a laboratory study

被引:15
作者
Paterson, E. [1 ]
Sanderson, P. M. [1 ,2 ,3 ]
Paterson, N. A. B. [4 ]
Liu, D. [2 ]
Loeb, R. G. [5 ]
机构
[1] Univ Queensland, Sch Psychol, Brisbane, Qld, Australia
[2] Univ Queensland, Sch ITEE, Brisbane, Qld, Australia
[3] Univ Queensland, Sch Med, Brisbane, Qld, Australia
[4] Lady Cilento Childrens Hosp, Brisbane, Qld, Australia
[5] Univ Arizona, Coll Med, Dept Anesthesiol, Tucson, AZ USA
基金
澳大利亚研究理事会;
关键词
auditory alarms; auditory displays; patient monitoring; pulse oximetry; sonification; AUDITORY DISPLAYS; PITCH; PERCEPTION; ATTENTION; ALARMS; MUSIC; SOUND; NOISE;
D O I
10.1111/anae.13424
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Our study examined the effectiveness of pulse oximetry sonification enhanced with acoustic tremolo and brightness to help listeners differentiate clinically relevant oxygen saturation ranges. In a series of trials lasting 30s each, 76 undergraduate participants identified final oxygen saturation range (Target: 100% to 97%; Low: 96% to 90%; Critical: 89% and below), and detected threshold transitions into and out of the target range using conventional sonification (n=38) or enhanced sonification (n=38). Median (IQR [range]) accuracy for range identification with the conventional sonification was 80 (70-85 [45-95])%, whereas with the enhanced sonification it was 100 (99-100 [80-100])%; p<0.001. Accuracy for detecting threshold transitions with the conventional sonification was 60 (50-75 [30-95])%, but with the enhanced sonification it was 100 (95-100 [75-100]%; p<0.001. Participants can identify clinically meaningful oxygen saturation ranges and detect threshold transitions more accurately with enhanced sonification than with conventional sonification.
引用
收藏
页码:565 / 572
页数:8
相关论文
共 35 条
[1]  
[Anonymous], 2010, Sonification Report: Status of the Field and Research Agenda
[2]   Comparison of the identification and ease of use of two alarm sound sets by critical and acute care nurses with little or no music training: a laboratory study [J].
Atyeo, J. ;
Sanderson, P. M. .
ANAESTHESIA, 2015, 70 (07) :818-827
[3]   A comparison of linear and logarithmic auditory tones in pulse oximeters [J].
Brown, Zoe ;
Edworthy, Judy ;
Sneyd, J. Robert ;
Schlesinger, Joseph .
APPLIED ERGONOMICS, 2015, 51 :350-357
[4]   Continuous auditory monitoring - how much information do we register? [J].
Craven, RM ;
McIndoe, AK .
BRITISH JOURNAL OF ANAESTHESIA, 1999, 83 (05) :747-749
[5]   Intra-operative monitoring - many alarms with minor impact [J].
de Man, F. R. ;
Greuters, S. ;
Boer, C. ;
Veerman, D. P. ;
Loer, S. A. .
ANAESTHESIA, 2013, 68 (08) :804-810
[6]  
Deschamps M-L, 2015, THESIS U QUEENSLAND
[7]   Alarms are still a problem! [J].
Edworthy, J. .
ANAESTHESIA, 2013, 68 (08) :791-794
[8]   Medical audible alarms: a review [J].
Edworthy, Judy .
JOURNAL OF THE AMERICAN MEDICAL INFORMATICS ASSOCIATION, 2013, 20 (03) :584-589
[9]   Heterogeneity in auditory alarm sets makes them easier to learn [J].
Edworthy, Judy ;
Hellier, Elizabeth ;
Titchener, Kirsteen ;
Naweed, Anjum ;
Roels, Richard .
INTERNATIONAL JOURNAL OF INDUSTRIAL ERGONOMICS, 2011, 41 (02) :136-146
[10]   G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences [J].
Faul, Franz ;
Erdfelder, Edgar ;
Lang, Albert-Georg ;
Buchner, Axel .
BEHAVIOR RESEARCH METHODS, 2007, 39 (02) :175-191