Comparison of Visual Assessment Techniques for Wear and Corrosion in Modular Hip Replacement Systems

被引:5
作者
Di Prima, Matthew A. [1 ]
Vesnovsky, Oleg [1 ]
Kovacs, Paul
Hopper, Robert H., Jr. [2 ]
Ho, Henry [2 ]
Engh, Charles A., Jr. [2 ]
Topoleski, L. D. Timmie [1 ,3 ]
机构
[1] US FDA, Ctr Device & Radiol Hlth, 10903 New Hampshire Ave, Silver Spring, MD 20993 USA
[2] Anderson Orthopaed Res Inst, Alexandria, VA 22306 USA
[3] Univ Maryland Baltimore Cty, Dept Mech Engn, Baltimore, MD 21250 USA
来源
MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES | 2015年 / 1591卷
关键词
metal-on-metal; corrosion; orthopaedic; wear; modular; hip arthroplasty; FRETTING CORROSION; TAPER JUNCTION; INTERFACE; FAILURE; HEAD; ARTHROPLASTY; RETRIEVAL; CREVICE;
D O I
10.1520/STP159120140152
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Wear and corrosion of implant metals have been associated with adverse tissue reactions and failures of total hips. However, the factors leading to component wear and corrosion, as well as the clinical effects of wear and corrosion, are not well understood. A visual inspection standard for taper corrosion and fretting, the Goldberg method, currently is used to quantify the amount of apparent corrosion damage present on an implant using a subjective score. We developed an approach to quantify the damage by creating digital microscopic images of the entire taper and trunnion surfaces, which enables an objective quantification of detailed wear and corrosion damage of the contacting surfaces, which we call the digital mosaic method (DMM). We compared the results of this new methodology to the Goldberg method and Anderson method (an expanded version of the Goldberg method). We examined the modular surfaces of ten retrieved metal-on-metal hip replacement systems. Each set of taper and trunnion surfaces were imaged and stitched into a composite image of the entire taper-trunnion surface. Three independent observers measured the surface area fraction of damage (wear or corrosion) for each surface, which in turn was compared with visual scores performed by independent observers on the same retrieved joints. To determine a measure of the intraobserver repeatability of the new method, a single reviewer measured the surface fraction of damage for the same set of explanted surfaces across different days, blinded to any previous specimen identification, with the image order randomized. The results of this study show that the DMM damage quantification correlates to the current visual scoring approaches (Goldberg and Anderson). The potential advantages to using this approach include better damage quantification, detailed analysis of specific locations, ability to identify matching damage locations on the two surfaces, and higher resolution of the contact area.
引用
收藏
页码:147 / 163
页数:17
相关论文
共 23 条
[1]   Fretting corrosion fatigue study of modular joints in total hip replacements by accelerated in vitro testing [J].
Bhambri, SK ;
Gilbertson, LN .
MODULARITY OF ORTHOPEDIC IMPLANTS, 1997, 1301 :146-156
[2]   High failure rates with a large-diameter hybrid metal-on-metal total hip replacement CLINICAL, RADIOLOGICAL AND RETRIEVAL ANALYSIS [J].
Bolland, B. J. R. F. ;
Culliford, D. J. ;
Langton, D. J. ;
Millington, J. P. S. ;
Arden, N. K. ;
Latham, J. M. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05) :608-615
[3]   Effects of neck extension, coverage, and frequency on the fretting corrosion of modular THR bore and cone interface [J].
Brown, SA ;
Abera, A ;
DOnofrio, M ;
Flemming, C .
MODULARITY OF ORTHOPEDIC IMPLANTS, 1997, 1301 :189-198
[4]   Metal-on-Metal Local Tissue Reaction Is Associated With Corrosion of the Head Taper Junction [J].
Fricka, Kevin B. ;
Ho, Henry ;
Peace, William J. ;
Engh, Charles A., Jr. .
JOURNAL OF ARTHROPLASTY, 2012, 27 (08) :26-31
[5]   IN-VIVO CORROSION OF MODULAR HIP-PROSTHESIS COMPONENTS IN MIXED AND SIMILAR METAL COMBINATIONS - THE EFFECT OF CREVICE, STRESS, MOTION, AND ALLOY COUPLING [J].
GILBERT, JL ;
BUCKLEY, CA ;
JACOBS, JJ .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (12) :1533-1544
[6]   The mechanical and electrochemical processes associated with taper fretting crevice corrosion: A review [J].
Gilbert, JL ;
Jacobs, JJ .
MODULARITY OF ORTHOPEDIC IMPLANTS, 1997, 1301 :45-59
[7]   Corrosion testing of modular hip implants [J].
Goldberg, JR ;
Buckley, CA ;
Jacobs, JJ ;
Gilbert, JL .
MODULARITY OF ORTHOPEDIC IMPLANTS, 1997, 1301 :157-176
[8]  
Goldberg JR, 2002, CLIN ORTHOP RELAT R, P149
[9]   Fretting corrosion mechanisms at modular implant interfaces [J].
Jani, SC ;
Sauer, WL ;
McLean, TW ;
Lambert, RD ;
Kovacs, P .
MODULARITY OF ORTHOPEDIC IMPLANTS, 1997, 1301 :211-225
[10]  
Kovacs P., 1996, MED APPL TITANIUM IT, P163