Development, safety and efficacy of a novel circular-irrigated deca-channel mapping and ablation catheter for pulmonary vein isolation

被引:0
作者
Feng, Xiang-Fei [1 ]
Chen, Mu [1 ]
Sun, Jian [1 ]
Sung, Jie [2 ]
Liang, Bo [2 ]
Sun, Yi-Yong [2 ]
Li, Yi-Gang [1 ]
机构
[1] Shanghai Jiao Tong Univ, Xinhua Hosp, Sch Med, Dept Cardiol, 1665 KongJiang Rd, Shanghai 200092, Peoples R China
[2] Shanghai MicroPort EP MedTech Co Ltd, Bldg 28,Lane 588,Tianxiong Rd, Shanghai, Peoples R China
来源
BMC CARDIOVASCULAR DISORDERS | 2018年 / 18卷
基金
中国国家自然科学基金;
关键词
Atrial fibrillation; Catheter ablation; Circular-irrigated deca-channel ablation catheter; Pulmonary vein isolation; Design requirement; ATRIAL-FIBRILLATION; NMARQ ABLATION; FOLLOW-UP; LESION; SIZE;
D O I
10.1186/s12872-018-0886-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Pulmonary vein isolation (PVI), a cornerstone for catheter ablation of atrial fibrillation (AF), remains a complex and time-consuming procedure. Present study introduces a novel, circular-irrigated, deca-channel mapping and ablation catheter (CIDMA), describes the in vitro test results on feasibility, safety, and acute efficacy of the CIDMA catheter. Methods: An assembled CIDMA catheter was subjected to a number of in vitro tests. With this catheter, ablation procedures were first performed in a pig's myocardial strips in vitro to determine the effects in unipolar or bipolar configuration. Results: Three catheters were assembled. The adjustable circular diameter was changed from initial state of 32.41 +/- 0.61 mm into controlled state of 28.61 +/- 0.47 mm (P = 0.013). In the plastic model, the push-ability, torque-ability, and kink resistance of CIDMA catheter were shown to be satisfactory. In vitro, our findings showed that ablation could produce obvious ablation lesions, and unipolar ablation (at length, width and depth of 5.0 +/- 1.3, 4.6 +/- 0.7, and 4.2 +/- 0.6 mm, respectively) was more effective than bipolar (at length, width and depth of 2.8 +/- 0.2, 4.2 +/- 0.5, and 2.3 +/- 0.4 mm, respectively) (P < 0.01). Conclusions: In vitro, our preliminary data suggest that the CIDMA catheter produced optimal ablation lesions, especially in the unipolar ablation mode. Future in vivo animal and clinical studies are warranted to test the efficacy of this catheter in real-world scenario.
引用
收藏
页数:7
相关论文
共 20 条
  • [1] Phrenic nerve lesion: A potential complication of the nMARQ ablation technique
    Arroja, Jose David
    Zimmermann, Marc
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 2015, 180 : 91 - 92
  • [2] Arunachalam SP, 2017, NUCL MED BIOMED IMAG, V2, P1
  • [3] Pulmonary vein isolation by duty-cycled bipolar and unipolar radiofrequency energy with a multielectrode ablation catheter
    Boersma, Lucas V. A.
    Wijffets, Maurits C. E. F.
    Oral, Hakan
    Wever, Eric F. D.
    Morady, Fred
    [J]. HEART RHYTHM, 2008, 5 (12) : 1635 - 1642
  • [4] Calkins H, 2012, HEART RHYTHM, V9, P632, DOI 10.1016/j.hrthm.2011.12.016
  • [5] Updated Worldwide Survey on the Methods, Efficacy, and Safety of Catheter Ablation for Human Atrial Fibrillation
    Cappato, Riccardo
    Calkins, Hugh
    Chen, Shih-Ann
    Davies, Wyn
    Iesaka, Yoshito
    Kalman, Jonathan
    Kim, You-Ho
    Klein, George
    Natale, Andrea
    Packer, Douglas
    Skanes, Allan
    Ambrogi, Federico
    Biganzoli, Elia
    [J]. CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2010, 3 (01) : 32 - 38
  • [6] EVALUATION OF THE ELASTIC BEHAVIOR OF CENTRAL VENOUS PVC, POLYURETHANE AND SILICONE CATHETERS
    CERVERA, M
    DOLZ, M
    HERRAEZ, JV
    BELDA, R
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (02) : 177 - 183
  • [7] Multi-selective catheter for MR-guided endovascular interventions
    Clogenson, Helene C. M.
    van Lith, Joris Y.
    Dankelman, Jenny
    Melzer, Andreas
    van den Dobbelsteen, John J.
    [J]. MEDICAL ENGINEERING & PHYSICS, 2015, 37 (07) : 623 - 630
  • [8] iphotoacoustic visualization of myocardial ablation lesions
    Dana, Nicholas
    Di Biase, Luigi
    Natale, Andrea
    Emelianov, Stanislav
    Bouchard, Richard
    [J]. HEART RHYTHM, 2014, 11 (01) : 150 - 157
  • [9] Deneke Thomas, 2009, Expert Rev Cardiovasc Ther, V7, P1341, DOI 10.1586/erc.09.116
  • [10] Visual Balloon-Guided Point-by-Point Ablation Reliable, Reproducible, and Persistent Pulmonary Vein Isolation
    Dukkipati, Srinivas R.
    Neuzil, Petr
    Skoda, Jan
    Petru, Jan
    d'Avila, Andre
    Doshi, Shephal K.
    Reddy, Vivek Y.
    [J]. CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2010, 3 (03) : 266 - U78