A sheath flow gating interface for the on-line coupling of solid- phase extraction with capillary electrophoresis

被引:12
作者
Weng, Qianfeng [1 ]
Fu, Lianjuan [1 ]
Li, Xintong [1 ]
An, Ran [1 ]
Li, Jinxiang [1 ]
机构
[1] Liaoning Normal Univ, Dept Chem, Dalian 116029, Peoples R China
关键词
Solid-phase extraction; Capillary electrophoresis; Interface; Phenols; COMPREHENSIVE 2-DIMENSIONAL SYSTEM; PERFORMANCE LIQUID-CHROMATOGRAPHY; C18 TRAPPING COLUMN; MASS-SPECTROMETRY; ZONE-ELECTROPHORESIS; MICROCOLUMN-LC; SIZE-EXCLUSION; HUMAN PLASMA; PRECONCENTRATION; INJECTION;
D O I
10.1016/j.aca.2014.10.050
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sheath flow gating interface (SFGI) is presented for the on-line coupling of solid-phase extraction (SPE) with capillary electrophoresis (CE). The design, construction and operation of the SFGI are described in detail. After operating conditions were investigated and selected, the SFGI was evaluated on a SPE-CE-UV setup using hydroxylated poly(glycidyl methacrylate-co-ethylene dimethacrylate) monolith as the absorbent and using three phenols as the test analytes. The preconcentration factors obtained with the SPE-CE-UV system and the SPE-UV part are 530 and 550, respectively. The plate numbers obtained using the SPE-CE-UV system are slightly better than or comparable to those with the CE-UV part. The precisions (RSDs) of 100 consecutive injections are 2.43%, 3.86%, and 4.25% for peak height, peak area and migration time, respectively. The measured recoveries for the river water samples spiked at three different levels are in the range of 93.6-102.8% with the interday RSD values ranging from 2.0 to 4.5% (n = 3). These data collectively demonstrate that the SFGI has the ability to exactly and reproducibly transfer nanoliters of fractions from SPE onto CE with no degradation of the efficiencies of SPE and CE, suggesting a great potential to be routinely used for the coupling of SPE, microcolumn LC or FIA with CE. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:46 / 52
页数:7
相关论文
共 29 条
[1]  
Baker D.R., 1995, CAPILLARY ELECTROPHO
[2]   On-line system for peptide mapping by capillary electrophoresis at sub-micromolar concentrations [J].
Bonneil, E ;
Waldron, KC .
TALANTA, 2000, 53 (03) :687-699
[3]   AUTOMATED INSTRUMENTATION FOR COMPREHENSIVE 2-DIMENSIONAL HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY CAPILLARY ZONE ELECTROPHORESIS [J].
BUSHEY, MM ;
JORGENSON, JW .
ANALYTICAL CHEMISTRY, 1990, 62 (10) :978-984
[4]   Combination of flow injection with capillary electrophoresis. Part 5. Automated preconcentration and determination of pseudoephedrine in human plasma [J].
Chen, HW ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1999, 394 (01) :13-22
[5]   Combination of flow injection with capillary electrophoresis. Part 3. On-line sorption column preconcentration capillary electrophoresis system [J].
Chen, HW ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1997, 355 (2-3) :135-143
[6]   Poly(dimethylsiloxane)-based microchip for two-dimensional solid-phase extraction-capillary electrophoresis with an integrated electrospray emitter tip [J].
Dahlin, AP ;
Bergström, SK ;
Andrén, PE ;
Markides, KE ;
Bergquist, J .
ANALYTICAL CHEMISTRY, 2005, 77 (16) :5356-5363
[7]   Reversed-phase capillary liquid chromatography coupled on-line to capillary electrophoresis immunoassays [J].
German, I ;
Kennedy, RT .
ANALYTICAL CHEMISTRY, 2000, 72 (21) :5365-5372
[8]   Large-bore particle-entrapped monolithic precolumns prepared by a sol-gel method for on-line peptides trapping and preconcentration in multidimensional liquid chromatography system for proteome analysis [J].
Gu, X ;
Wang, Y ;
Zhang, XM .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1072 (02) :223-232
[9]   Improved solid-phase microextraction device for use in on-line immunoaffinity capillary electrophoresis [J].
Guzman, NA .
ELECTROPHORESIS, 2003, 24 (21) :3718-3727
[10]   Evaluation of on-line solid phase extraction-capillary electrophoresis-electrospray-mass spectrometry for the analysis of neuropeptides in human plasma [J].
Hernandez, Elena ;
Benavente, Fernando ;
Sanz-Nebot, Victoria ;
Barbosa, Jose .
ELECTROPHORESIS, 2008, 29 (16) :3366-3376