Validation of a method to measure three-dimensional hip joint kinematics in subjects with femoroacetabular impingement

被引:2
作者
Ward, Thomas R. [1 ]
Hussain, Mafruha M. [2 ]
Pickering, Mark [2 ]
Perriman, Diana [1 ]
Burns, A., I [1 ]
Scarvell, Jennie [1 ,3 ]
Smith, Paul N. [1 ,4 ]
机构
[1] Canberra Hosp, Trauma & Orthopaed Res Unit, Bldg 6 Level 1,POB 2, Woden, ACT 2606, Australia
[2] Univ New South Wales, Australian Def Force Acad, Canberra, ACT, Australia
[3] Univ Canberra, Canberra, ACT, Australia
[4] Australian Natl Univ, Canberra, ACT, Australia
关键词
Femoroacetabular impingement; fluoroscopy; hip; image registration; kinematics; OSTEOARTHRITIS; ACCURACY; MOTION;
D O I
10.1177/1120700019883548
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: A kinematic measurement method combining dynamic motion and imaging, which captures the behaviour of the hip at terminal motion, may offer improved diagnostic accuracy and enhance our understanding of the mechanics of femoroacetabular impingement (FAI). Methods: 3 embalmed cadaveric hip/pelvis specimens with implanted Roentgen Stereophotogrammetric Analysis (RSA) beads were mounted on a custom rig and imaged with a fluoroscope in four poses to simulate a clinical impingement examination: in hip extension and in three positions: near impingement, early impingement and late impingement while simulating a flexion/adduction/internal rotation manoeuvre. Hip joint kinematics were measured using 2 methods and compared: RSA (gold standard) and a custom 3-dimensional to 2-dimensional (3D-2D) image registration method which matches 3D models developed from CT to 2D fluoroscopic images. Results: Using RSA as the gold standard, bias and precision of hip joint rotations measured using 3D-2D registration demonstrated maximums of 1.64 degrees and 3.96 degrees, respectively. However, if the single outlier was removed, bias and precision were 0.55 degrees and 1.38 degrees. Bias and precision of translations had maximums of 0.51 mm and 0.77 mm, respectively. Conclusions: This 3D to 2D registration method may offer a clinically useful solution for dynamic assessment of hip impingement. If 5-mm translation and 10 degrees of rotation represent a clinically significant difference in hip kinematics, the method's accuracy of approximately 1 mm displacement and 1 degrees rotation should enable detection of significant clinical differences.
引用
收藏
页码:133 / 139
页数:7
相关论文
共 19 条
[1]   Hip morphology influences the pattern of damage to the acetabular cartilage - Femoroacetabular impingement as a cause of early osteoarthritis of the hip [J].
Beck, M ;
Kalhor, M ;
Leunig, M ;
Ganz, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2005, 87B (07) :1012-1018
[2]   Femoroacetabular impingement -: A cause for osteoarthritis of the hip [J].
Ganz, R ;
Parvizi, J ;
Beck, M ;
Leunig, M ;
Nötzli, H ;
Siebenrock, KA .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) :112-120
[3]   Kinematic Analysis of Healthy Hips during Weight-Bearing Activities by 3D-to-2D Model-to-Image Registration Technique [J].
Hara, Daisuke ;
Nakashima, Yasuharu ;
Hamai, Satoshi ;
Higaki, Hidehiko ;
Ikebe, Satoru ;
Shimoto, Takeshi ;
Hirata, Masanobu ;
Kanazawa, Masayuki ;
Kohno, Yusuke ;
Iwamoto, Yukihide .
BIOMED RESEARCH INTERNATIONAL, 2014, 2014
[4]   A trabecular metal tibial component in total knee replacement in patients younger than 60 years A TWO-YEAR RADIOSTEREOPHOTOGRAMMETRIC ANALYSIS [J].
Henricson, A. ;
Linder, L. ;
Nilsson, K. G. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (12) :1585-1593
[5]  
Hossain MM, 2014, IEEE ENG MED BIO, P6044, DOI 10.1109/EMBC.2014.6945007
[6]   Accuracy and Feasibility of Dual Fluoroscopy and Model-Based Tracking to Quantify in Vivo Hip Kinematics During Clinical Exams [J].
Kapron, Ashley L. ;
Aoki, Stephen K. ;
Peters, Christopher L. ;
Maas, Steve A. ;
Bey, Michael J. ;
Zauel, Roger ;
Anderson, Andrew E. .
JOURNAL OF APPLIED BIOMECHANICS, 2014, 30 (03) :461-470
[7]   Fluoroscopic Demonstration of Femoroacetabular Impingement During Hip Arthroscopy [J].
Lee, Cara Beth ;
Clark, John .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2011, 27 (07) :994-1004
[8]   In-vitro validation of a non-invasive dual fluoroscopic imaging technique for measurement of the hip kinematics [J].
Lin, Hao ;
Wang, Shaobai ;
Tsai, Tsung-Yuan ;
Li, Guoan ;
Kwon, Young-Min .
MEDICAL ENGINEERING & PHYSICS, 2013, 35 (03) :411-416
[9]   Model-Based Tracking of the Hip: Implications for Novel Analyses of Hip Pathology [J].
Martin, Daniel E. ;
Greco, Nicholas J. ;
Klatt, Brian A. ;
Wright, Vonda J. ;
Anderst, William J. ;
Tashman, Scott .
JOURNAL OF ARTHROPLASTY, 2011, 26 (01) :88-97
[10]   Image-assisted non-invasive and dynamic biomechanical analysis of human joints [J].
Muhit, Abdullah A. ;
Pickering, Mark R. ;
Scarvell, Jennifer M. ;
Ward, Tom ;
Smith, Paul N. .