4D flow MRI: impact of region of interest size, angulation and spatial resolution on aortic flow assessment

被引:15
作者
Casciaro, M. E. [1 ]
Pascaner, A. F. [1 ]
Guilenea, F. N. [1 ]
Alcibar, J. [1 ]
Gencer, U. [2 ]
Soulat, G. [2 ]
Mousseaux, E. [2 ]
Craiem, D. [1 ]
机构
[1] Univ Favaloro, CONICET, Inst Med Traslac Trasplante & Bioingn IMeTTyB, Solis 453, RA-1078 Buenos Aires, DF, Argentina
[2] Hop Europeen Georges Pompidou, INSERM U970, Cardiovasc Imaging Unit, Paris, France
关键词
Blood flow measurement; Aorta; Phase-contrast MRI; Image processing; WALL SHEAR-STRESS;
D O I
10.1088/1361-6579/abe525
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Objectives. In cardiovascular magnetic resonance, the 3D time-resolved phase-contrast technique, also known as 4D flow, is gaining increasing attention due to applications that exploit three-directional velocity encoding throughout the cardiac cycle. Blood flow volume assessment usually requires an expert to draw regions of interest (ROI) around the vessel cross section, whereas the errors involved in this estimation have not been thoroughly investigated. Our objective is to quantify the influence of ROI sizing, angulation and spatial resolution of the reconstructed plane employed in blood flow measurements using 4D flow. Approach. Three circular ROIs were drawn around the ascending, arch and descending aorta of healthy volunteers (n = 27) and patients with a dilated ascending aorta or bicuspid valve (n = 37). We applied systematic changes of ROI diameter (up to +/- 10%), tilt angle (up to +/- 25 degrees) and spatial resolution (from 0.25 to 2 mm) of the reconstructed oblique planes, calculating the effects on net, forward and backward blood flow volumes. Main results. Patients had a larger ascending aorta than healthy volunteers with similar ages and male sex proportion (60 +/- 15 y.o. vs 58 +/- 16 y.o. and 84% vs 70%, respectively). Higher forward and backward flow volumes were observed in the ascending aorta and the aortic arch of the patients with respect to controls (p < 0.001), whereas net volumes were similar: 74.0 +/- 20.8 ml versus 75.7 +/- 21.8 ml (p = 0.37), respectively. The ascending aorta was the most sensitive to ROI modifications. Changes of +/- 10% in the ROI diameter and +/- 25 degrees in tilt angles produced flow volume differences of up to 9 ml (10%) and 18 ml (15%) in controls and patients, respectively. Modifying the reconstructed planes spatial resolution produced flow volume changes below 2 ml. Significance. Since the setting of the ROI size and plane angle could produce errors that represent up to 20% of the forward and/or backward aortic flow volume, a good standardization for vessel segmentation and plane positioning is desirable.
引用
收藏
页数:11
相关论文
共 18 条
[1]   Four-dimensional Flow MRI: Principles and Cardiovascular Applications [J].
Azarine, Arshid ;
Garcon, Philippe ;
Stansal, Audrey ;
Canepa, Nadia ;
Angelopoulos, Giorgios ;
Silvera, Stephane ;
Sidi, Daniel ;
Marteau, Veronique ;
Zins, Marc .
RADIOGRAPHICS, 2019, 39 (03) :632-+
[2]   Aortic Dilation in Bicuspid Aortic Valve Disease Flow Pattern Is a Major Contributor and Differs With Valve Fusion Type [J].
Bissell, Malenka M. ;
Hess, Aaron T. ;
Biasiolli, Luca ;
Glaze, Steffan J. ;
Loudon, Margaret ;
Pitcher, Alex ;
Davis, Anne ;
Prendergast, Bernard ;
Markl, Michael ;
Barker, Alex J. ;
Neubauer, Stefan ;
Myerson, Saul G. .
CIRCULATION-CARDIOVASCULAR IMAGING, 2013, 6 (04) :499-507
[3]   Validation and reproducibility of cardiovascular 4D-flow MRI from two vendors using 2 x 2 parallel imaging acceleration in pulsatile flow phantom and in vivo with and without respiratory gating [J].
Bock, Jelena ;
Toger, Johannes ;
Bidhult, Sebastian ;
Bloch, Karin Markenroth ;
Arvidsson, Per ;
Kanski, Mikael ;
Arheden, Hakan ;
Testud, Frederik ;
Greiser, Andreas ;
Heiberg, Einar ;
Carlsson, Marcus .
ACTA RADIOLOGICA, 2019, 60 (03) :327-337
[4]   Two-Minute k-Space and Time-accelerated Aortic Fourdimensional Flow MRI: Dual-Center Study of Feasibility and Impact on Velocity and Wall Shear Stress Quantification [J].
Bollache, Emilie ;
Knott, Kristopher D. ;
Jarvis, Kelly ;
Boubertakh, Redha ;
Dolan, Ryan Scott ;
Camaioni, Claudia ;
Collins, Louise ;
Scully, Paul ;
Rabin, Sydney ;
Treibel, Thomas ;
Carr, James C. ;
van Ooij, Pim ;
Collins, Jeremy D. ;
Geiger, Julia ;
Moon, James C. ;
Barker, Alex J. ;
Petersen, Steffen E. ;
Markl, Michael .
RADIOLOGY-CARDIOTHORACIC IMAGING, 2019, 1 (02)
[5]   Automatic correction of background phase offset in 4D-flow of great vessels and of the heart in MRI using a third-order surface model [J].
Craiem, Damian ;
Pascaner, Ariel F. ;
Casciaro, Mariano E. ;
Gencer, Umit ;
Alcibar, Joaquin ;
Soulat, Gilles ;
Mousseaux, Elie .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2019, 32 (06) :629-642
[6]   Low and Oscillatory Wall Shear Stress Is Not Related to Aortic Dilation in Patients With Bicuspid Aortic Valve A Time-Resolved 3-Dimensional Phase-Contrast Magnetic Resonance Imaging Study [J].
Dux-Santoy, Lydia ;
Guala, Andrea ;
Sotelo, Julio ;
Uribe, Sergio ;
Teixido-Tura, Gisela ;
Ruiz-Munoz, Aroa ;
Hurtado, Daniel E. ;
Valente, Filipa ;
Galian-Gay, Laura ;
Gutierrez, Laura ;
Gonzalez-Alujas, Teresa ;
Johnson, Kevin M. ;
Wieben, Oliver ;
Ferreira, Ignacio ;
Evangelista, Arturo ;
Rodriguez-Palomares, Jose F. .
ARTERIOSCLEROSIS THROMBOSIS AND VASCULAR BIOLOGY, 2020, 40 (01) :E10-E20
[7]   4D flow cardiovascular magnetic resonance consensus statement [J].
Dyverfeldt, Petter ;
Bissell, Malenka ;
Barker, Alex J. ;
Bolger, Ann F. ;
Carlhall, Carl-Johan ;
Ebbers, Tino ;
Francios, Christopher J. ;
Frydrychowicz, Alex ;
Geiger, Julia ;
Giese, Daniel ;
Hope, Michael D. ;
Kilner, Philip J. ;
Kozerke, Sebastian ;
Myerson, Saul ;
Neubauer, Stefan ;
Wieben, Oliver ;
Markl, Michael .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2015, 17
[8]   Influence of Aortic Dilation on the Regional Aortic Stiffness of Bicuspid Aortic Valve Assessed by 4-Dimensional Flow Cardiac Magnetic Resonance Comparison With Marfan Syndrome and Degenerative Aortic Aneurysm [J].
Guala, Andrea ;
Rodriguez-Palomares, Jose ;
Dux-Santoy, Lydia ;
Teixido-Tura, Gisela ;
Maldonado, Giuliana ;
Galian, Laura ;
Huguet, Marina ;
Valente, Filipa ;
Gutierrez, Laura ;
Gonzalez-Alujas, Teresa ;
Johnson, Kevin M. ;
Wieben, Oliver ;
Aviles, Augusto Sao ;
Garcia-Dorado, David ;
Evangelista, Arturo .
JACC-CARDIOVASCULAR IMAGING, 2019, 12 (06) :1020-1029
[9]   IN-VIVO VALIDATION OF MAGNETIC-RESONANCE BLOOD-VOLUME FLOW MEASUREMENTS WITH LIMITED SPATIAL-RESOLUTION IN SMALL VESSELS [J].
HOFMAN, MBM ;
VISSER, FC ;
VANROSSUM, AC ;
VINK, GQM ;
SPRENGER, M ;
WESTERHOF, N .
MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (06) :778-784
[10]   Quantifying errors in flow measurement using phase contrast magnetic resonance imaging: comparison of several boundary detection methods [J].
Jiang, Jing ;
Kokeny, Paul ;
Ying, Wang ;
Magnano, Chris ;
Zivadinov, Robert ;
Haacke, E. Mark .
MAGNETIC RESONANCE IMAGING, 2015, 33 (02) :185-193