Perimembranous ventricular septal defect closure in infants with KONAR-MF occluder

被引:5
作者
Singhi, Anil Kumar [1 ,4 ]
Mukherji, Aritra [2 ]
De, Arnab [3 ]
机构
[1] Med Super Specialty Hosp, Dept Pediat Cardiol, Kolkata 700099, India
[2] Hlth City Cayman Isl, Dept Pediat Cardiol, 1283 Sea View Rd, Grand Cayman KY1 1005, Cayman Islands
[3] Med Super Specialty Hosp, Kolkata, India
[4] Med Super Specialty Hosp, Dept Pediat Cardiol, Kolkata 700099, West Bengal, India
关键词
Ventricular septal defect; Perimembranous; Device; Infant; KONAR-MF; CHILDREN;
D O I
10.1016/j.ppedcard.2022.101489
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Ventricular septal defect (VSD) with significant left to right shunt requires closure. Device closure in Infants with perimembranous VSD and moderate left to right shunt is emerging as an indication in selective cases. Objective: Analysis of infants with moderate perimembranous VSD closed with new KONAR MF occluder. Methods: Infants with perimembranous VSD and moderate left to right shunt who underwent device closure with KONAR MF device (Lifetech Inc., Shenzhen, China) in a tertiary care setup were analyzed retrospectively. Results: Five infants had perimembranous VSD with moderate left to right shunt. They underwent device closure with a KONAF MF device. The mean age was 7.8 months and the mean defect size was 4.3 mm. The VSD was crossed antegradely in two patients and retrogradely in three patients. All but one patient had an antegrade deployment. The KONAR MF device size ranged from 6-4 mm to 10-8 mm. The device was deployed with either Launcher guiding catheter (Medtronic AVE, Santa Rosa, CA) 5F and 6F or 5F Lifetech delivery system. Successful deployment was possible in all patients. One patient had tilting of the device 3 h after the procedure with significant residual flow. The patient was sent for surgery after combined cardiology and surgical discussion. All other infants had stable device positions. There was no residual flow and no significant aortic regurgitation or any hemolysis. In the mean follow up of 15 months, they were all doing well, had good device performance with no evidence of heart block. Conclusions: Device closure of moderate sized perimembranous VSD in infants with KONAR MF occluder appeared to be safe and effective in short term follow up. Larger studies with long term data will help with future recommendations.
引用
收藏
页数:7
相关论文
共 16 条
[1]   Transcarotid Approach to Ventricular Septal Defect Closure in Small Infants [J].
Breatnach, Colm R. ;
Kenny, Damien ;
Linnane, Niall ;
Al Nasef, Mohamed ;
Ng, Li Yen ;
McGuinness, Jonathan ;
McCrossan, Brian ;
Nolke, Lars ;
Oslizlok, Paul ;
Redmond, Mark ;
Walsh, Kevin .
PEDIATRIC CARDIOLOGY, 2021, 42 (07) :1539-1545
[2]   Transcatheter closure of a membranous ventricular septal defect in a 1.8-kg infant using Amplatzer Duct Occluder II additional size device [J].
Champaneri, Bhavik ;
Kappanayil, Mahesh ;
Kumar, Raman K. .
CARDIOLOGY IN THE YOUNG, 2017, 27 (07) :1437-1440
[3]   Percutaneous closure of restrictive-type perimembranous ventricular septal defect using the new KONAR multifunctional occluder: Midterm outcomes of the first middle-eastern experience [J].
Haddad, Raymond N. ;
Daou, Linda S. ;
Saliba, Zakhia S. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2020, 96 (03) :E295-E302
[4]   The incidence of congenital heart disease [J].
Hoffman, JIE ;
Kaplan, S .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (12) :1890-1900
[5]   Transcatheter Retrograde Closure of Perimembranous Ventricular Septal Defects in Children With the Amplatzer Duct Occluder II Device [J].
Koneti, Nageswara Rao ;
Sreeram, Narayanswami ;
Penumatsa, Raghava Raju ;
Arramraj, Srinivas K. ;
Karunakar, Vadlamudi ;
Trieschmann, Uwe .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (23) :2421-2422
[6]   Safety and efficacy of transcatheter closure of outlet-type ventricular septal defects in children and adults with Amplatzer Duct Occluder II [J].
Lin, Hsin-Chia ;
Lin, Ming-Tai ;
Chen, Chun-An ;
Hsu, Jui-Yu ;
Lin, Su-Man ;
Wu, Mei-Hwan ;
Wang, Jou-Kou .
JOURNAL OF THE FORMOSAN MEDICAL ASSOCIATION, 2021, 120 (01) :180-188
[7]   Novel use of Konar-multifunctional occluder [J].
Mukherji, Aritra ;
Chattopadhyay, Amitabha ;
Ghosh, Sanjiban ;
Mohapatra, Soumya Kanti ;
De, Arnab ;
Singhi, Anil Kumar .
CARDIOLOGY IN THE YOUNG, 2020, 30 (11) :1722-1727
[8]   Percutaneous Ventricular Septal Defect Closure in Patients Under 1 Year of Age [J].
Narin, Nazmi ;
Pamukcu, Ozge ;
Tuncay, Aydin ;
Baykan, Ali ;
Sunkak, Suleyman ;
Tasci, Onur ;
Uzum, Kazim ;
Saltik, Levent .
PEDIATRIC CARDIOLOGY, 2018, 39 (05) :1009-1015
[9]   Transcatheter Closure of Moderate to Large Perimembranous Ventricular Septal Defects in Children Weighing 10 kilograms or less [J].
Pillai, Ajith Ananthakrishna ;
Rangasamy, Sasinthar ;
Balasubramonian, Vidhyakar Rangaswamy .
WORLD JOURNAL FOR PEDIATRIC AND CONGENITAL HEART SURGERY, 2019, 10 (03) :278-285
[10]   Transcatheter closure of ventricular septal defects in children less than 10 kg: experience from a tertiary care referral hospital in Eastern India [J].
Roy, Mahua ;
Gangopadhyay, Debasree ;
Goyal, Noopur ;
Murthy, Savita ;
Nandi, Debabrata ;
Bandyopadhyay, Biswajit ;
Dutta, Joyeeta .
CARDIOLOGY IN THE YOUNG, 2022, 32 (01) :48-54