Flow injection measurement of lead using mercury-free disposable gold-sputtered screen-printed carbon electrodes (SPCE)

被引:50
作者
Masawat, P
Liawruangrath, S [1 ]
Slater, JM
机构
[1] Chiang Mai Univ, Fac Sci, Dept Chem, Chiang Mai 50200, Thailand
[2] Univ London Birkbeck Coll, Sch Biol & Chem Sci, London WC1H 0PP, England
关键词
gold-coated screen-printed carbon electrode; lead; flow injection analysis;
D O I
10.1016/S0925-4005(03)00066-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Screen-printed sensors are promising devices for disposable, cheap and reliable environmental monitoring. In previous studies, workers have used mercury-coated electrodes to allow stripping voltammetry to be performed. In this paper, we describe a disposable sputtered gold sensor which allows underpotential analyte preconcentration and avoids the environmental contamination associated with mercury-based sensors. It is used in combination with a specially fabricated thin-layer flow cell for stripping analysis. The sensor consists of a screen-printed strip with three electrodes; gold coated over carbon-silver ink as the working electrode, silver-silver chloride ink as the pseudo reference electrode, and a carbon-silver ink as the counter electrode. The optimized flow injection (FI) system allows the convenient monitoring of micrograms per litre lead concentrations following short deposition times (detection limit 0.8 mug l(-1) at 120 s deposition). The method was evaluated by determining lead in spiked drinking and tap water samples; the recoveries of Pb2+ were 103% (R.S.D. 2.8%) and 97.9% (R.S.D. 7.1%), n = 5, respectively. Measurements in the presence of typical interferences such as copper, cadmium, zinc, iron, chromium, and mercury are reported. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:52 / 59
页数:8
相关论文
共 29 条
[1]   Determination of lead and mercury in sea water by preconcentration in a flow injection system followed by atomic absorption spectrometry detection [J].
Bravo-Sánchez, LR ;
de la Riva, BS ;
Costa-Fernández, JM ;
Pereiro, R ;
Sanz-Medel, A .
TALANTA, 2001, 55 (06) :1071-1078
[2]  
BREEN J, 1995, LEAD POISONING
[3]   Simultaneous determination of Cd(II), Cu(II) and Pb(II) in surface waters by solid phase extraction and flow injection analysis with spectrophotometric detection [J].
Castillo, E ;
Cortina, JL ;
Beltrán, JL ;
Prat, MD ;
Granados, M .
ANALYST, 2001, 126 (07) :1149-1153
[4]   Determination of Cu, Pb, Cd, and Zn in river sediment extracts by sequential injection anodic stripping voltammetry with thin mercury film electrode [J].
da Silva, CL ;
Masini, JC .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 367 (03) :284-290
[5]  
FANG ZL, 1993, FLOW INJECTION SEPER
[6]   Nickel determination in saline matrices by ICP-AES after sorption on Amberlite XAD-2 loaded with PAN [J].
Ferreira, SLC ;
de Brito, CF ;
Dantas, AF ;
de Araújo, NML ;
Costa, ACS .
TALANTA, 1999, 48 (05) :1173-1177
[7]  
Honeychurch KC, 2000, ELECTROANAL, V12, P171, DOI 10.1002/(SICI)1521-4109(200002)12:3<171::AID-ELAN171>3.0.CO
[8]  
2-Q
[9]   Voltammetric studies of lead at calixarene modified screen-printed carbon electrodes and its trace determination in water by stripping voltammetry [J].
Honeychurch, KC ;
Hart, JP ;
Cowell, DC ;
Arrigan, DWM .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 77 (03) :642-652
[10]   Voltammetric studies of lead at a 1-(2-pyridylazo)-2-naphthol modified screen-printed carbon electrode and its trace determination in water by stripping voltammetry [J].
Honeychurch, KC ;
Hart, JP ;
Cowell, DC .
ANALYTICA CHIMICA ACTA, 2001, 431 (01) :89-99