Determination of manganese in brain samples by slurry sampling graphite furnace atomic absorption spectrometry

被引:11
作者
Marcó, LM
Caraballo, EAH
Pascusso, C
Alvarado, J
机构
[1] Univ Simon Bolivar, Dept Quim, Lab Espectroscopia Atom, Caracas 1080A, Venezuela
[2] Univ Ctr Occidental Lisandro Alvarado, Decanato Med, Dept Ciencias Func, Barquisimeto, Venezuela
[3] Univ Los Andes, Fac Ciencias, IVAIQUIM, Merida 5101, Venezuela
[4] Univ Ctr Occidental Lisandro Alvarado, Dept Quim & Suelos, Unidad Anal Instrumental, Decanato Agron, Barquisimeto, Venezuela
关键词
cadmium; slurry; brain; electrothermal atomization atomic absorption spectrometry;
D O I
10.1016/S0039-9140(02)00652-5
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple procedure for the determination of manganese in different sections of human brain samples by graphite furnace atomic absorption spectrometry has been developed. Brain sections included cerebellum, hypothalamus, frontal cortex, vermix and encephalic trunk. Two sample preparation procedures were evaluated, namely, slurry sampling and microwave-assisted acid digestion. Brain slurries (2% w/v) could be prepared in distilled, de-ionized water, with good stability for up to 30 min. Brain samples were also digested in a domestic microwave oven using 5 ml of concentrated HNO3. A mixed palladium+magnesium nitrate chemical modifier was used for thermal stabilization of the analyte in the electrothermal atomizer up to pyrolysis temperatures of 1300 degreesC, irrespective of the matrix. Quantitation of manganese was conducted in both cases by means of aqueous standards calibration. The detection limits were 0.3 and 0.4 ng ml(-1) for the slurry and the digested samples, respectively. The accuracy of the procedure was checked by comparing the results obtained in the analysis of slurries and digested brain samples, and by analysis of the NIST Bovine Liver standard reference material (SRM 1577a). The ease of slurry preparation, together with the conventional set of analytical and instrumental conditions selected for the determination of manganese make such methodology suitable for routine clinical applications. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:897 / 904
页数:8
相关论文
共 25 条
[1]  
ALVARADO J, 1991, J TRACE ELEM ELECT H, V5, P173
[2]   CHARACTERIZATION OF BIOLOGICAL RMS FOR POTENTIAL USE IN HUMAN BRAIN ANALYSIS [J].
ANDRASI, E ;
DOZSA, A ;
BEZUR, L ;
ERNYEI, L ;
MOLNAR, Z .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 345 (2-4) :340-342
[3]   CONCENTRATIONS OF ELEMENTS IN BRAIN-TUMORS [J].
ANDRASI, E ;
OROSZ, L ;
SCHEIBLER, H ;
BEZUR, L ;
ERNYEI, L .
MIKROCHIMICA ACTA, 1995, 118 (1-2) :113-121
[4]   Several methods to determine heavy metals in the human brain [J].
Andrási, E ;
Igaz, S ;
Szoboszlai, N ;
Farkas, E ;
Ajtony, Z .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1999, 54 (05) :819-825
[5]   SOLID SAMPLING IN ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY USING COMMERCIAL ATOMIZERS - A REVIEW [J].
BENDICHO, C ;
DELOOSVOLLEBREGT, MTC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (05) :353-374
[6]   PLATFORM IN FURNACE ZEEMAN-EFFECT ATOMIC-ABSORPTION SPECTROMETRIC DETERMINATION OF ARSENIC IN BEER BY ATOMIZATION OF SLURRIES OF SAMPLE ASH [J].
CERVERA, ML ;
NAVARRO, A ;
MONTORO, R ;
DELAGUARDIA, M ;
SALVADOR, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (06) :477-481
[7]   AUTOMATED SLURRY SAMPLE INTRODUCTION FOR ANALYSIS OF A RIVER SEDIMENT BY GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY [J].
EPSTEIN, MS ;
CARNRICK, GR ;
SLAVIN, W ;
MILLERIHLI, NJ .
ANALYTICAL CHEMISTRY, 1989, 61 (13) :1414-1419
[8]   ANALYSIS OF SILICON-NITRIDE POWDERS FOR AL, CR, CU, PA, K, MG, MN, NA, AND ZN BY SLURRY-SAMPLING ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
FRIESE, KC ;
KRIVAN, V .
ANALYTICAL CHEMISTRY, 1995, 67 (02) :354-359
[9]   RAPID FURNACE PROGRAMS FOR THE SLURRY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRIC DETERMINATION OF CHROMIUM, LEAD AND COPPER IN DIATOMACEOUS-EARTH [J].
GARCIA, IL ;
CORTEZ, JA ;
CORDOBA, MH .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (01) :103-108
[10]  
*HGA, 1990, HGA 700 GRAPH FURN O