Percentage depth dose evaluation in heterogeneous media using thermoluminescent dosimetry

被引:19
作者
da Rosa, L. A. R. [2 ]
Cardoso, S. C. [1 ]
Campos, L. T. [2 ]
Alves, V. G. L. [1 ]
Batista, D. V. S. [3 ]
Facure, A. [4 ]
机构
[1] Univ Fed Rio de Janeiro, Inst Fis, IRD CNEN, BR-21941972 Rio De Janeiro, Brazil
[2] Univ Fed Rio de Janeiro, Inst Radioprotecao & Dosimetria, BR-21941972 Rio De Janeiro, Brazil
[3] INCA, Inst Nacl Canc, Rio De Janeiro, Brazil
[4] Comissao Nacl Energia Nucl, Rio De Janeiro, Brazil
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2010年 / 11卷 / 01期
关键词
thermoluminescent detector; lung heterogeneity; small field size; 15; MV; treatment planning system; MONTE-CARLO; STEREOTACTIC RADIOSURGERY; ENERGY; BEAMS; DISTRIBUTIONS; RADIOTHERAPY; ALGORITHMS; TRANSPORT; TLD-100; IMPACT;
D O I
10.1120/jacmp.v11i1.2947
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study is to investigate the influence of lung heterogeneity inside a soft tissue phantom on percentage depth dose (PDD). PDD curves were obtained experimentally using LiF: Mg,Ti (TLD-100) thermoluminescent detectors and applying Eclipse treatment planning system algorithms Batho, modified Batho (M-Batho or BMod), equivalent TAR (E-TAR or EQTAR), and anisotropic analytical algorithm (AAA) for a 15 MV photon beam and field sizes of 1 x 1, 2 x 2, 5 x 5, and 10 x 10 cm(2). Monte Carlo simulations were performed using the DOSRZnrc user code of EGSnrc. The experimental results agree with Monte Carlo simulations for all irradiation field sizes. Comparisons with Monte Carlo calculations show that the AAA algorithm provides the best simulations of PDD curves for all field sizes investigated. However, even this algorithm cannot accurately predict PDD values in the lung for field sizes of 1 x 1 and 2 x 2 cm(2). An overdos-age in the lung of about 40% and 20% is calculated by the AAA algorithm close to the interface soft tissue/lung for 1 x 1 and 2 x 2 cm(2) field sizes, respectively. It was demonstrated that differences of 100% between Monte Carlo results and the algorithms Batho, modified Batho, and equivalent TAR responses may exist inside the lung region for the 1 x 1 cm(2) field.
引用
收藏
页码:117 / 127
页数:11
相关论文
共 33 条
[1]   The impact of electron transport on the accuracy of computed dose [J].
Arnfield, MR ;
Siantar, CH ;
Siebers, J ;
Garmon, P ;
Cox, L ;
Mohan, R .
MEDICAL PHYSICS, 2000, 27 (06) :1266-1274
[2]  
BATHO H, 1964, J CAN ASSOC RADIOL, V15, P70
[3]  
BJAMGARD BE, 1990, MED PHYS, V17, P794
[4]   Energy dependence of TLD 100 and MCP-N detectors [J].
Carinoua, E. ;
Boziari, A. ;
Askounisa, P. ;
Mikulis, A. ;
Kamenopoulou, V. .
RADIATION MEASUREMENTS, 2008, 43 (2-6) :599-602
[5]   Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium [J].
Carrasco, P ;
Jornet, N ;
Duch, MA ;
Weber, L ;
Ginjaume, M ;
Eudaldo, T ;
Jurado, D ;
Ruiz, A ;
Ribas, M .
MEDICAL PHYSICS, 2004, 31 (10) :2899-2911
[6]   Precision for low dose assessment using TLD-100 chips and computerised glow curve analysis [J].
da Rosa, LAR ;
Regulla, DF ;
Fill, UA .
RADIATION PROTECTION DOSIMETRY, 1999, 85 (1-4) :175-178
[7]   Dose evaluation in lung-equivalent media in high-energy photon external radiotherapy [J].
Duch, M. A. ;
Carrasco, P. ;
Ginjaume, M. ;
Jornet, N. ;
Ortega, X. ;
Ribas, M. .
RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) :43-47
[8]   Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee [J].
Ezzell, GA ;
Galvin, JM ;
Low, D ;
Palta, JR ;
Rosen, I ;
Sharpe, MB ;
Xia, P ;
Xiao, Y ;
Xing, L ;
Yu, CX .
MEDICAL PHYSICS, 2003, 30 (08) :2089-2115
[9]   Comparison of TLD-100 and MCP-Ns for use as an extremity dosemeter for PET nuclear medicine staff [J].
Ginjaume, M. ;
Perez, S. ;
Duch, M. A. ;
Ortega, X. .
RADIATION MEASUREMENTS, 2008, 43 (2-6) :607-610
[10]  
HOROWITZ YS, 1990, RADIAT PROT DOSIM, V30, P219