Cost-effective immobilization for whole brain radiation therapy

被引:5
作者
Rubinstein, Ashley E. [1 ,2 ]
Ingram, W. Scott [1 ,2 ]
Anderson, Brian M. [1 ,2 ]
Gay, Skylar S. [1 ,2 ]
Fave, Xenia J. [1 ,2 ]
Ger, Rachel B. [1 ,2 ]
McCarroll, Rachel E. [1 ,2 ]
Owens, Constance A. [1 ,2 ]
Netherton, Tucker J. [1 ,2 ]
Kisling, Kelly D. [1 ,2 ]
Court, Laurence E. [1 ,2 ]
Yang, Jinzhong [1 ,2 ]
Li, Yuting [1 ,2 ]
Lee, Joonsang [1 ]
Mackin, Dennis S. [1 ]
Cardenas, Carlos E. [1 ,2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Grad Sch Biomed Sci, Houston, TX 77030 USA
关键词
accuracy; immobilization; low- and middle-income countries; palliation; setup; whole-brain treatment; CONE-BEAM CT; NECK-CANCER RADIOTHERAPY; STEREOTACTIC RADIOTHERAPY; IMAGE GUIDANCE; ACCURACY; MASK; POSITION; SETUP; RADIOSURGERY; FRACTION;
D O I
10.1002/acm2.12101
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To investigate the inter- and intra-fraction motion associated with the use of a low-cost tape immobilization technique as an alternative to thermoplastic immobilization masks for whole-brain treatments. The results of this study may be of interest to clinical staff with severely limited resources (e.g., in low-income countries) and also when treating patients who cannot tolerate standard immobilization masks. Setup reproducibility of eight healthy volunteers was assessed for two different immobilization techniques. (a) One strip of tape was placed across the volunteer's forehead and attached to the sides of the treatment table. (b) A second strip was added to the first, under the chin, and secured to the table above the volunteer's head. After initial positioning, anterior and lateral photographs were acquired. Volunteers were positioned five times with each technique to allow calculation of inter-fraction reproducibility measurements. To estimate intra-fraction reproducibility, 5-minute anterior and lateral videos were taken for each technique per volunteer. An in-house software was used to analyze the photos and videos to assess setup reproducibility. The maximum intra-fraction displacement for all volunteers was 2.8mm. Intra-fraction motion increased with time on table. The maximum inter-fraction range of positions for all volunteers was 5.4mm. The magnitude of inter-fraction and intra-fraction motion found using the 1-strip and 2-strip tape immobilization techniques was comparable to motion restrictions provided by a thermoplastic mask for whole-brain radiotherapy. The results suggest that tape-based immobilization techniques represent an economical and useful alternative to the thermoplastic mask.
引用
收藏
页码:116 / 122
页数:7
相关论文
共 28 条
[1]   Comparison of two dead and neck immobilization systems [J].
Bentel, GC ;
Marks, LB ;
Hendren, K ;
Brizel, DM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :867-873
[2]  
Beyzadeoglu Murat., 2015, RAD THERAPY HEAD NEC
[3]   Trend analysis of radiation therapy incidents over seven years [J].
Bissonnette, Jean-Pierre ;
Medlam, Gaylene .
RADIOTHERAPY AND ONCOLOGY, 2010, 96 (01) :139-144
[4]   Repositioning accuracy of two different mask systems - 3D revisited: Comparison using true 3D/3D matching with cone-beam CT [J].
Boda-Heggemann, Judit ;
Walter, Cornelia ;
Rahn, Angelika ;
Wertz, Hansjoerg ;
Loeb, Iris ;
Lohr, Frank ;
Wenz, Frederik .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (05) :1568-1575
[5]   Effect of Radiosurgery Alone vs Radiosurgery With Whole Brain Radiation Therapy on Cognitive Function in Patients With 1 to 3 Brain Metastases A Randomized Clinical Trial [J].
Brown, Paul D. ;
Jaeckle, Kurt ;
Ballman, Karla V. ;
Farace, Elana ;
Cerhan, Jane H. ;
Anderson, S. Keith ;
Carrero, Xiomara W. ;
Barker, Fred G., II ;
Deming, Richard ;
Burri, Stuart H. ;
Menard, Cynthia ;
Chung, Caroline ;
Stieber, Volker W. ;
Pollock, Bruce E. ;
Galanis, Evanthia ;
Buckner, Jan C. ;
Asher, Anthony L. .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2016, 316 (04) :401-409
[6]  
Chelvarajah R., 2004, Radiographer, V51, P29
[7]   The management of radiation treatment error through incident learning [J].
Clark, Brenda G. ;
Brown, Robert J. ;
Ploquin, Jodi L. ;
Kind, Anneke L. ;
Grimard, Laval .
RADIOTHERAPY AND ONCOLOGY, 2010, 95 (03) :344-349
[8]   Time dependence of intrafraction patient motion assessed by repeat stereoscopic imaging [J].
Hoogeman, Mischa S. ;
Nuyttens, Joost J. ;
Levendag, Peter C. ;
Heumen, Ben J. M. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (02) :609-618
[9]  
Kang Hyejoo, 2010, J Appl Clin Med Phys, V12, P3270
[10]   Estimation of adequate setup margins and threshold for position errors requiring immediate attention in head and neck cancer radiotherapy based on 2D image guidance [J].
Kapanen, Mika ;
Laaksomaa, Marko ;
Tulijoki, Tapio ;
Peltola, Seppo ;
Wigren, Tuija ;
Hyodynmaa, Simo ;
Kellokumpu-Lehtinen, Pirkko-Liisa .
RADIATION ONCOLOGY, 2013, 8