Primary Cementless Total Hip Arthroplasty with an Alumina Ceramic-on-Ceramic Bearing Results After a Minimum of Twenty Years of Follow-up

被引:65
作者
Petsatodis, George E. [1 ]
Papadopoulos, Pericles P. [1 ]
Papavasiliou, Kyriakos A. [1 ]
Hatzokos, Ippokratis G. [1 ]
Agathangelidis, Filon G. [1 ]
Christodoulou, Anastasios G. [1 ]
机构
[1] Aristotle Univ Thessaloniki, Orthopaed Dept 1, G Papanikolaou Gen Hosp, Sch Med, Thessaloniki 57010, Greece
关键词
YOUNG-PATIENTS; WEAR; DISLOCATION; STEM;
D O I
10.2106/JBJS.H.01829
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The biological problems related to wear debris after total hip arthroplasty have stimulated renewed interest in alternatives to meta I-on-polyethylene bearing surfaces. Methods: We retrospectively evaluated the clinical and radiographic results of 100 patients who had undergone a total of 109 primary total hip arthroplasties with a cementless alumina ceramic-on-ceramic prosthesis between January 1985 and December 1989. The mean age of the patients at the time of the index arthroplasty was forty-six years. Clinical evaluation was performed with use of the Charnley modification of the Merle d'Aubigne-Postel scale. Seventy-eight patients who had had a total of eighty-five arthroplasties were available for follow-up evaluation at an average of 20.8 years. The patients' average age at the time of the latest follow-up was 66.8 years. Results: Six hips (six acetabular cups and one femoral stem) in six patients underwent revision. Aseptic loosening of the cup combined with focal osteolysis was the cause of all six revisions. In one patient, the stem was also revised because of aseptic loosening. At the time of final follow-up, the result was excellent (according to the Merle d'Aubigne-Postel scale) in 68% of the hips, good in 19%, fair in 9%, and poor in 4%. The mean Merle d'Aubigne-Postel score improved from 7.9 points preoperatively to 16.9 points postoperatively (p < 0.001). The cumulative rate of survival of the prostheses was 84.4% at 20.8 years. Conclusions: The results of these cementless ceramic-on-ceramic total hip arthroplasties continued to be satisfactory at a minimum of twenty years postoperatively. The improved design of contemporary prostheses and the new generation of ceramic-on-ceramic bearing surfaces may lead to even better long-term results.
引用
收藏
页码:639 / 644
页数:6
相关论文
共 27 条
[21]   Serum Ion Levels after Ceramic-on-Ceramic and Metal-on-Metal Total Hip Arthroplasty: 8-Year Minimum Follow-up [J].
Savarino, Lucia ;
Padovanni, Giovanni ;
Ferretti, Massimo ;
Greco, Michelina ;
Cenni, Elisabetta ;
Perrone, Giorgio ;
Greco, Francesco ;
Baldini, Nicola ;
Giunti, Armando .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2008, 26 (12) :1569-1576
[22]   PERIPROSTHETIC BONE LOSS IN TOTAL HIP-ARTHROPLASTY - POLYETHYLENE WEAR DEBRIS AND THE CONCEPT OF THE EFFECTIVE JOINT SPACE [J].
SCHMALZRIED, TP ;
JASTY, M ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (06) :849-863
[23]  
Sherk HH, 2005, CLIN ORTHOP RELAT R, P3, DOI 10.1097/01.blo.0000150447.40645.50
[24]   A study for a retrieved implant of ceramic-on-ceramic total hip arthroplasty [J].
Shishido, T ;
Yamamoto, K ;
Tanaka, S ;
Masaoka, T ;
Clarke, IC ;
Williams, P .
JOURNAL OF ARTHROPLASTY, 2006, 21 (02) :294-298
[25]   The role of stripe wear in causing acoustic emissions from alumina ceramic-on-ceramic bearings [J].
Taylor, Scott ;
Manley, Michael T. ;
Sutton, Kate .
JOURNAL OF ARTHROPLASTY, 2007, 22 (07) :47-51
[26]   Squeaking in ceramic-on-ceramic hips - The importance of acetabular component orientation [J].
Walter, William L. ;
O'Toole, Gary C. ;
Walter, William K. ;
Ellis, Andrew ;
Zicat, Bernard A. .
JOURNAL OF ARTHROPLASTY, 2007, 22 (04) :496-503
[27]   Alumina-on-alumina total hip arthroplasty - A five-year minimum follow-up study [J].
Yoo, JJ ;
Kim, YM ;
Yoon, KS ;
Koo, KH ;
Song, WS ;
Kim, HJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (03) :530-535