Validity and Between-Day Reliability of the Cervical Range of Motion (CROM) Device

被引:276
作者
Audette, Isabel [2 ]
Dumas, Jean-Pierre [3 ,4 ]
Cote, Julie N. [1 ,4 ]
De Serres, Sophie J. [4 ]
机构
[1] McGill Univ, Dept Kinesiol & Phys Educ, Montreal, PQ H2W 1S4, Canada
[2] McGill Univ, Sch Phys & Occupat Therapy, Montreal, PQ H2W 1S4, Canada
[3] Univ Sherbrooke, Dept Rehabil, Sherbrooke, PQ J1K 2R1, Canada
[4] CRIR Jewish Rehabil Hosp Site, Laval, PQ, Canada
关键词
clinical tool; MDC; neck; psychometrics; spine; WHIPLASH-ASSOCIATED DISORDERS; FASTRAK MEASUREMENT SYSTEM; NECK PAIN; ACTIVE RANGE; SPINE; MANAGEMENT; HEALTH;
D O I
10.2519/jospt.2010.3180
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
STUDY DESIGN: Clinical measurement, validity and intrarater reliability study. OBJECTIVES: (1) To confirm the validity and assess between-day test-retest reliability of cervical spine motion measurements made with the cervical range of motion (CROM) device in flexion, extension, bilateral rotation, and bilateral side flexion; (2) to provide meaningful information to clinicians about the standard error of measurement and the minimal detectable change for the CROM device. BACKGROUND: Range of motion is a common outcome measure used in the assessment of the cervical spine. The CROM device is one of the tools used to measure cervical range of motion in the clinical setting. However, its psychometric properties are not well established, especially for measurements taken on separate days. METHODS: Quasi-experimental design with 1 group comparison. Twenty healthy adults (9 men and 11 women) participated in this study. Cervical range of motion was simultaneously recorded with the CROM device and the Fastrak motion analysis system for all 6 cervical movements mentioned above. The CROM device was placed on the participant's head consistent with standard clinical procedures. Two Fastrak sensors were positioned with 1 on the forehead and lover the spinous process of the T6 vertebra. Test-retest reliability of measurements made with the CROM device was assessed, as well as its standard error of measurement and minimal detectable change, with measures taken on 2 separate days spaced 48 hours apart. RESULTS: Values obtained by the 2 measuring devices yielded Pearson correlation coefficients ranging between 0.93 and 0.98. Test-retest reliability of measurements of cervical range of motion using the CROM was found to be good, with ICCs ranging between 0.89 and 0.98. The standard errors of measurement across the 6 movements ranged from 1.6 degrees to 2.8 degrees and the minimal detectable changes across the 6 movements ranged from 3.6 degrees to 6.5 degrees. CONCLUSIONS: The measurements made with the CROM were shown to be reliable in all movement directions. J Orthop Sports Phys Ther 2010;40(5):318-323. doi:10.2519/jospt.2010.3180
引用
收藏
页码:318 / 323
页数:6
相关论文
共 30 条
[1]   Measuring range of active cervical rotation in a position of full head flexion using the 3D Fastrak measurement system: an intra-tester reliability study [J].
Amiri, M ;
Jull, G ;
Bullock-Saxton, J .
MANUAL THERAPY, 2003, 8 (03) :176-179
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Meta-analysis of normative cervical motion [J].
Chen, J ;
Solinger, AB ;
Poncet, JF ;
Lantz, CA .
SPINE, 1999, 24 (15) :1571-1578
[4]   Neck pain: Clinical practice guidelines linked to the international classification of functioning, disability, and health from the orthopaedic section of the American physical therapy association [J].
Childs, John D. ;
Cleland, Joshua A. ;
Elliott, James M. ;
Teyhen, Deydre S. ;
Wainner, Robert S. ;
Whitman, Julie M. ;
Sopky, Bernard J. ;
Godges, Joseph J. ;
Flynn, Timothy W. .
JOURNAL OF ORTHOPAEDIC & SPORTS PHYSICAL THERAPY, 2008, 38 (09) :A1-A34
[5]   The annual incidence and course of neck pain in the general population:: a population-based cohort study [J].
Côté, P ;
Cassidy, JD ;
Carroll, LJ ;
Kristman, V .
PAIN, 2004, 112 (03) :267-273
[6]   Clinimetric evaluation of active range of motion measures in patients with non-specific neck pain: a systematic review [J].
de Koning, Chantal H. P. ;
van den Heuvel, Sylvia P. ;
Staal, J. Bart ;
Smits-Engelsman, Bouwien C. M. ;
Hendriks, Erik J. M. .
EUROPEAN SPINE JOURNAL, 2008, 17 (07) :905-921
[7]   Minimal changes in health status questionnaires: distinction between minimally detectable change and minimally important change [J].
de Vet, Henrica C. ;
Terwee, Caroline B. ;
Ostelo, Raymond W. ;
Beckerman, Heleen ;
Knol, Dirk L. ;
Bouter, Lex M. .
HEALTH AND QUALITY OF LIFE OUTCOMES, 2006, 4 (1)
[8]   Characterization of Acute and Chronic Whiplash-Associated Disorders [J].
Elliott, James M. ;
Noteboom, Jon Timothy ;
Flynn, Timothy W. ;
Sterling, Michele .
JOURNAL OF ORTHOPAEDIC & SPORTS PHYSICAL THERAPY, 2009, 39 (05) :312-323
[9]  
ENSINK FB, 1976, SPINE, V21, P1339
[10]   Intrarater Reliability of CROM Measurement of Cervical Spine Active Range of Motion in Persons With and Without Neck Pain [J].
Fletcher, James P. ;
Bandy, William D. .
JOURNAL OF ORTHOPAEDIC & SPORTS PHYSICAL THERAPY, 2008, 38 (10) :640-645