EIS measurements for characterization of muscular tissue by means of equivalent electrical parameters

被引:29
作者
Clemente, Fabrizio [1 ]
Romano, Maria [2 ]
Bifulco, Paolo [2 ]
Cesarelli, Mario [2 ]
机构
[1] CNR, Ist Ingn Biomed, Ist Biostrutture & Bioimmagini, Rome, Italy
[2] Univ Naples Federico II, DIETI, Naples, Italy
关键词
Biological measurement; Electrical impedance spectroscopy; Muscle; Electrical models of biological tissue; Isometric contraction; BIOELECTRICAL-IMPEDANCE TECHNIQUES; PROSTHESIS OSSEOINTEGRATION; BIOIMPEDANCE MEASUREMENT; STRIATED-MUSCLE; DRUG-DELIVERY; HUMAN SKIN; SPECTROSCOPY; MYOGRAPHY; MEDICINE;
D O I
10.1016/j.measurement.2014.09.013
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The objective of this work was to study the use of electrical impedance spectroscopy (EIS) measurement to characterize muscle electrical properties in different conditions. In vivo EIS measurements were carried out in the range 1-60 kHz on 32 forearm flexor muscles of healthy volunteers. Each subject underwent to a protocol consisting of three trials: rest condition, isometric contraction, 4 min relaxation. Measured data were fitted with Cole-Cole equivalent electrical circuit using the complex nonlinear least square method. Finally, the relative variations from the steady state of the electrical parameters of the equivalent electrical circuit were estimated. Results obtained are in agreement with the hypothesis that, in the frequency range considered, the current preferentially flows through the extracellular fluid, being the component of intracellular resistance negligible. Among circuital parameters, in fact, the relative variations from steady state of R-0 changed highly significant (p < 0.001) both during contraction and after 4 min relaxation. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:476 / 482
页数:7
相关论文
共 31 条
[1]   Effects of age on muscle as measured by electrical impedance myography [J].
Aaron, Ronald ;
Esper, Gregory J. ;
Shiffman, Carl A. ;
Bradonjic, Kaca ;
Lee, Kyungmouk S. ;
Rutkove, Seward B. .
PHYSIOLOGICAL MEASUREMENT, 2006, 27 (10) :953-959
[2]  
Arpaia P., 2007, 2007 IEEE INSTR MEAS, P1
[3]   Low-invasive diagnosis of metallic prosthesis osseointegration by electrical impedance spectroscopy [J].
Arpaia, Pasquale ;
Clemente, Fabrizio ;
Zanesco, Antonio .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2007, 56 (03) :784-789
[4]   An instrument for prosthesis osseointegration assessment by electrochemical impedance spectrum measurement [J].
Arpaia, Pasquale ;
Clemente, Fabrizio ;
Romanucci, Carmine .
MEASUREMENT, 2008, 41 (09) :1040-1044
[5]  
Barsoukov E, 2005, IMPEDANCE SPECTROSCOPY: THEORY, EXPERIMENT, AND APPLICATIONS, 2ND EDITION, P1, DOI 10.1002/0471716243
[6]  
Basmajian J.V., MUSCLE FATIGUE TIME
[7]   A FLEXIBLE FFT ALGORITHM FOR PROCESSING BIOMEDICAL SIGNALS USING A PERSONAL-COMPUTER [J].
CESARELLI, M ;
CLEMENTE, F ;
BRACALE, M .
JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (06) :527-530
[8]   Characterization of the Muscle Electrical Properties in Low Back Pain Patients by Electrical Impedance Myography [J].
Ching, Congo Tak-Shing ;
Chen, Yueh-Chi ;
Lu, Li-Hua ;
Hsieh, Peiyuan F. ;
Hsiao, Chin-Sung ;
Sun, Tai-Ping ;
Shieh, Hsiu-Li ;
Chang, Kang-Ming .
PLOS ONE, 2013, 8 (04)
[9]  
Clemente F., 2012, IEEE INT S MED MEAS, P1
[10]   Study of muscular tissue in different physiological conditions using electrical impedance spectroscopy measurements [J].
Clemente, Fabrizio ;
Romano, Maria ;
Bifulco, Paolo ;
Cesarelli, Mario .
BIOCYBERNETICS AND BIOMEDICAL ENGINEERING, 2014, 34 (01) :4-9