Reversed-phase liquid chromatography column testing: robustness study of the test

被引:30
作者
Le Mapihan, K [1 ]
Vial, J [1 ]
Jardy, A [1 ]
机构
[1] CNRS, UMR 7121, Ecole Super Phys & Chim Ind, Lab Environm & Chim Analyt, F-75005 Paris, France
关键词
robustness; experimental design; stationary phases; column characterization; solvent effect; basic compounds;
D O I
10.1016/j.chroma.2004.10.093
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Choosing the right RPLC column for an actual separation among the more than 600 commercially available ones still represents a real challenge for the analyst particularly when basic solutes are involved. Many tests dedicated to the characterization and the classification of stationary phases have been proposed in the literature and some of them highlighted the need of a better understanding of retention properties to lead to a rational choice of columns. However, unlike classical chromatographic methods. the problem of their robustness evaluation has often been left unaddressed. In the present study, we present a robustness study that was applied to the chromatographic testing procedure we had developed and optimized previously. A design of experiment (DoE) approach was implemented. Four factors. previously identified as potentially influent, were selected and subjected to small controlled variations: solvent fraction, temperature, pH and buffer concentration. As our model comprised quadratic terms instead of a simple linear model, we chose a D-optimal design in order to minimize the experiment number. As a previous batch-to-batch study [K. Le Mapihan, Caracterisation et classification des phases stationnaires utilisees pour l'analyse CPL de produits pharmaceutiques, Ph.D. Thesis, Pierre and Marie Curie University, 2004] had shown a low variability on the selected stationary phase, it was then possible to split the design into two pans, according to the solvent nature, each using one column. Actually, our testing procedure involving assays both with methanol and with acetonitrile as organic modifier, such an approach enabled to avoid a possible bias due to the column ageing considering the number of experiments required (16 + 6 center points). Experimental results were computed thanks to a Partial Least Squares regression procedure, more adapted than the classical regression to handle factors and responses not completely independent. The results showed the behavior of the solutes in relation to their physico-chemical properties and the relevance of the second term degree of our model. Finally, the robust domain of the test has been fairly identified, so that any potential user precisely knows to which extend each experimental parameter must be controlled when our testing procedure is to be implemented. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:149 / 158
页数:10
相关论文
共 44 条
[1]   Comparative study of test methods for reversed-phase columns for high-performance liquid chromatography [J].
Claessens, HA ;
van Straten, MA ;
Cramers, CA ;
Jezierska, M ;
Buszewski, B .
JOURNAL OF CHROMATOGRAPHY A, 1998, 826 (02) :135-156
[2]   Trends and progress in the characterization of stationary phases for reversed-phase liquid chromatography [J].
Claessens, HA .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2001, 20 (10) :563-583
[3]   CHROMATOGRAPHIC CHARACTERIZATION OF SILICA-BASED REVERSED PHASES [J].
ENGELHARDT, H ;
LOW, H ;
GOTZINGER, W .
JOURNAL OF CHROMATOGRAPHY, 1991, 544 (1-2) :371-379
[4]   Chromatographic determination of metallic impurities in reversed-phase HPLC columns [J].
Engelhardt, H ;
Lobert, T .
ANALYTICAL CHEMISTRY, 1999, 71 (09) :1885-1892
[5]  
Engelhardt H, 1997, LC GC-MAG SEP SCI, V15, P856
[6]   Mixture design design generation, PLS analysis, and model usage [J].
Eriksson, L ;
Johansson, E ;
Wikström, C .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1998, 43 (1-2) :1-24
[7]   Chromatographic classification and comparison of commercially available reversed-phase liquid chromatographic columns using principal component analysis [J].
Euerby, M ;
Petersson, P .
JOURNAL OF CHROMATOGRAPHY A, 2003, 994 (1-2) :13-36
[8]  
Euerby MR, 2000, LC GC EUR, V13, P665
[9]   Identification of the factors that influence the reproducibility of chromatographic retention data [J].
Felinger, A ;
Kele, M ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 2001, 913 (1-2) :23-48
[10]  
HEYDEN YV, 1996, ROBUSTNESS ANAL CHEM, P79