H*(10) and neutron spectra around linacs

被引:25
作者
Rene Vega-Carrillo, Hector [1 ]
Ortiz-Hernandez, Arturo [1 ]
Martin Hernandez-Davila, Victor [1 ]
Hernandez-Almaraz, Berenice [1 ]
Rivera Montalvo, Teodoro [2 ]
机构
[1] Univ Autonoma Zacatecas, Unidad Acad Estudios Nucl, Zacatecas 98000, Zac, Mexico
[2] CICATA IPN, Unidad Legaria, Mexico City, DF, Mexico
关键词
Photoneutrons; Neutron spectra; Dose; Bonner sphere spectrometer; MONTE-CARLO; DOSE-EQUIVALENT;
D O I
10.1007/s10967-009-0363-5
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Neutron spectrum and ambient dose equivalent has been measured around two 10 MV linear accelerators. Measurements were carried out in Mevatron and Primus model linacs, both made by Siemens. Main differences between those models are the beam collimator and the vault room. Here, Bonner sphere spectrometer with a passive thermal neutron detector has been utilized to measure the neutron spectrum inside the vault. Using an active detector the neutron spectrum was measured at the vault's door of both accelerators. With a neutron area monitor the dose equivalent was measured by the door at both vault doors. Neutron strength, total fluence rate and ambient dose equivalent were compared, from this was found that shielding conditions are better in the Primus model.
引用
收藏
页码:537 / 540
页数:4
相关论文
共 18 条
[1]   Monte Carlo simulation estimates of neutron doses to critical organs of a patient undergoing 18 MV x-ray LINAC-based radiotherapy [J].
Barquero, R ;
Edwards, TM ;
Iñiguez, MP ;
Vega-Carrillo, HR .
MEDICAL PHYSICS, 2005, 32 (12) :3579-3588
[2]   FRUIT:: An operational tool for multisphere neutron spectrometry in workplaces [J].
Bedogni, Roberto ;
Domingo, Carles ;
Esposito, Adolfo ;
Fernandez, Francisco .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (03) :1301-1309
[3]   Depth dose-equivalent and effective energies of photoneutrons generated by 6-18 MV X-ray beams for radiotherapy [J].
d'Errico, F ;
Luszik-Bhadra, M ;
Nath, R ;
Siebert, BRL ;
Wolf, U .
HEALTH PHYSICS, 2001, 80 (01) :4-11
[4]   Neutron measurements in a Varian 2100C LINAC facility using a Bonner sphere system based on passive gold activation detectors [J].
Fernandez, F. ;
Domingo, C. ;
Amgarou, K. ;
Castelo, J. ;
Bouassoule, T. ;
Garcia, M. J. ;
Luguera, E. .
RADIATION PROTECTION DOSIMETRY, 2007, 126 (1-4) :361-365
[5]   Monte Carlo calculations and validation of a gold foil-based Bonner sphere system [J].
Fernandez, F. ;
Bouassoule, T. ;
Amgarou, K. ;
Domingo, C. ;
Garcia, M. J. ;
Lacoste, V. ;
Gressier, V. ;
Muller, H. .
RADIATION PROTECTION DOSIMETRY, 2007, 126 (1-4) :366-370
[6]   Comparison of activation products and induced dose rates in different high-energy medical linear accelerators [J].
Fischer, Helmut W. ;
Tabot, Ben ;
Poppe, Bjoern .
HEALTH PHYSICS, 2008, 94 (03) :272-278
[7]  
FRY SF, 2009, MED PHYS, V36, P1244
[8]  
Hashemi SM, 2007, NUKLEONIKA, V52, P39
[9]   MULTILEAF COLLIMATION VERSUS CONVENTIONAL SHIELDING BLOCKS - A TIME AND MOTION STUDY OF BEAM SHAPING IN RADIOTHERAPY [J].
HELYER, SJ ;
HEISIG, S .
RADIOTHERAPY AND ONCOLOGY, 1995, 37 (01) :61-64
[10]   Investigation of secondary neutron dose for 18 MV dynamic MLC IMRT delivery [J].
Howell, RM ;
Ferenci, MS ;
Hertel, NE ;
Fullerton, GD .
MEDICAL PHYSICS, 2005, 32 (03) :786-793