Semi-automated 96-well solid-phase extraction and gas chromatography-negative chemical ionization tandem mass spectrometry for the trace analysis of fluprostenol in rat plasma

被引:5
|
作者
Gauw, RD [1 ]
Stoffolano, PJ [1 ]
Kuhlenbeck, DL [1 ]
Patel, VS [1 ]
Garver, SM [1 ]
Baker, TR [1 ]
Wehmeyer, KR [1 ]
机构
[1] Procter & Gamble Pharmaceut, Mason, OH 45040 USA
来源
JOURNAL OF CHROMATOGRAPHY B | 2000年 / 744卷 / 02期
关键词
fluprostenol;
D O I
10.1016/S0378-4347(00)00254-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Semi-automated 96-well plate solid-phase extraction (SPE) was used for sample preparation of fluprostenol, a prostaglandin analog, in rat plasma prior to detection by gas chromatography-negative chemical ionization tandem mass spectrometry (GC-NCI-MS-MS). A liquid handling system was utilized for all aspects of sample handling prior to SPE including transferring of samples into a 96-well format, preparation of standards as well as addition of internal standard to standards, quality control samples and study samples. SPE was performed in a 96-well plate format using octadecylsilane packing and the effluent from the SPE was dried in a custom-made 96-well apparatus. The sample residue was derivatized sequentially with pentafluorobenzylbromide followed by N-methyl-N-trimethylsilyltrifluoroacetamide. The derivatized sample was then analyzed using GC-NCI-MS-MS. The dynamic range for the method was from 7 to 5800 pg/ml with a 0.1-ml plasma sample. The methodology was evaluated over a 4-day period and demonstrated an accuracy of 90-106% with a precision of 2.4-12.9%. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:283 / 291
页数:9
相关论文
共 50 条