Clinical use and material wear of polymeric tracheostomy tubes

被引:20
作者
Bjoerling, Gunilla [1 ]
Axelsson, Sara
Johansson, Unn-Britt
Lysdahl, Michael
Markstroem, Agneta
Schedin, Ulla
Aune, Ragnhild E.
Frostell, Claes
Karlsson, Sigbritt
机构
[1] Danderyd Hosp, Natl Resp Ctr, Karolinska Inst, Div Anaesthesia & Intens Care,Dept Clin Sci, SE-18288 Stockholm, Sweden
[2] Sophiahemmet Univ Coll, Stockholm, Sweden
[3] Royal Inst Technol KTH, Fiber & Polymer Technol, Sch Chem Sci & Engn, Stockholm, Sweden
[4] Royal Inst Technol KTH, Sch Ind Technol & Management, Stockholm, Sweden
[5] Karolinska Univ Hosp, Karolinska Inst, Dept Anaesthesia & Intens Care Med, Stockholm, Sweden
关键词
tracheostomy; respiratory failure; material; material degradation; artificial airway; silicone; polyvinyl chloride; polyurethane; tracheostomy management; tracheostomy tube;
D O I
10.1097/MLG.0b013e31806911e3
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objectives: The objectives were to compare the duration of use of polymeric tracheostomy tubes, i.e., silicone (Si), polyvinyl chloride (PVC), and polyurethane (PU), and to determine whether surface changes in the materials could be observed after 30 days of patient use. Methods: Data were collected from patient and technical records for all tracheostomized patients attending the National Respiratory Center in Sweden. In the surface study, 19 patients with long-term tracheostomy were included: six with Bivona TTS Si tubes, eight with Shiley PVC tubes, and five with Trachoe Twist PU tubes. All tubes were exposed in the trachea for 30 days before being analyzed by scanning electron microscopy (SEM) and attenuated total reflectance Fourier transform infrared spectroscopy (ATR-FTIR). New tubes and tubes exposed in phosphate-buffered saline were used as reference. Results: Si tubes are used for longer periods of time than those made of PVC (P <.0001) and PU (P =.021). In general, all polymeric tubes were used longer than the recommended 30-day period. Eighteen of the 19 tubes exposed in patients demonstrated, in one or more areas of the tube, evident surface changes. The morphologic changes identified by SEM correlate well with the results obtained by ATR-FTIR. Conclusions: Si tracheostomy tubes are in general used longer than those made of PVC and PU. Most of the tubes exposed in the trachea for 30 days suffered evident surface changes, with degradation of the polymeric chains as a result.
引用
收藏
页码:1552 / 1559
页数:8
相关论文
共 13 条
[1]   FRACTURE OF A SYNTHETIC FENESTRATED TRACHEOSTOMY TUBE - CASE-REPORT AND REVIEW OF THE LITERATURE [J].
ALVI, A ;
ZAHTZ, GD .
AMERICAN JOURNAL OF OTOLARYNGOLOGY, 1994, 15 (01) :63-67
[2]  
Bauer T T, 2002, Monaldi Arch Chest Dis, V57, P84
[3]   A retrospective survey of outpatients with long-term tracheostomy [J].
Björling, G ;
Johansson, UB ;
Andersson, G ;
Schedin, U ;
Markström, A ;
Frostell, C .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2006, 50 (04) :399-406
[4]   Relevance of biofilms for the biodeterioration of surfaces of polymeric materials [J].
Flemming, HC .
POLYMER DEGRADATION AND STABILITY, 1998, 59 (1-3) :309-315
[5]   The concomitant development of poly(vinyl chloride)-related biofilm and antimicrobial resistance in relation to ventilator-associated pneumonia [J].
Gorman, SP ;
McGovern, JG ;
Woolfson, AD ;
Adair, CG ;
Jones, DS .
BIOMATERIALS, 2001, 22 (20) :2741-2747
[6]   DEGRADATION OF POLYURETHANES IN BIOMEDICAL APPLICATIONS - A REVIEW [J].
GRIESSER, HJ .
POLYMER DEGRADATION AND STABILITY, 1991, 33 (03) :329-354
[7]  
Jarrett William A, 2002, Ear Nose Throat J, V81, P659
[8]  
Lewarski Joseph S, 2005, Respir Care, V50, P534
[9]   In-vivo degradation of polyurethanes: transmission-FTIR microscopic characterization of polyurethanes sectioned by cryomicrotomy [J].
McCarthy, SJ ;
Meijs, GF ;
Mitchell, N ;
Gunatillake, PA ;
Heath, G ;
Brandwood, A ;
Schindhelm, K .
BIOMATERIALS, 1997, 18 (21) :1387-1409
[10]  
Sherman JM, 2000, AM J RESP CRIT CARE, V161, P297