Thalassemia Screening Using Different Automated Blood Cell Counters: Consideration of Appropriate Cutoff Values

被引:13
作者
Chaitraiphop, Chanintorn [1 ]
Sanchaisuriya, Kanokwan [2 ]
Inthavong, Sakoun [3 ]
Fucharoen, Goonnapa [2 ]
Sanchaisuriya, Pattara [4 ]
Changtrakun, Yossombut [5 ]
Fucharoen, Supan [2 ]
机构
[1] Khon Kaen Univ, Grad Sch, Med Sci Program, Khon Kaen 40002, Thailand
[2] Khon Kaen Univ, Fac Associated Med Sci, Ctr Res & Dev Med Diagnost Labs CMDL, Khon Kaen 40002, Thailand
[3] Minist Hlth, Bur Nutr, Viangchan, Laos
[4] Khon Kaen Univ, Fac Publ Hlth, Dept Nutr, Khon Kaen 40002, Thailand
[5] Khon Kaen Univ, Srinagarind Hosp, Diagnost Microscopy Unit, Fac Med, Khon Kaen 40002, Thailand
关键词
thalassemia screening; automated blood cell counter; mean corpuscular volume; mean corpuscular hemoglobin; cutoff values; NORTHEAST THAILAND; HB-H; HEMOGLOBINOPATHIES; TRAIT; ALPHA(0)-THALASSEMIA; DISORDERS; PREGNANCY; INDEXES;
D O I
10.7754/Clin.Lab.2015.150720
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: To evaluate the use of MCV and MCH cutoffs for thalassemia screening in areas with a high frequency of thalassemia and hemoglobinopathies using different hematology analyzers. Methods: Blood samples of known alpha(0)-thalassemia and beta-thalassemia carriers were analyzed to establish the appropriate cutoffs for each analyzer. These selected cutoffs were validated prospectively for screening of alpha(0)-thalassemia, beta-thalassemia in combination with the dichlorophenolindophenol test for hemoglobin E on another 288 Thai and 325 Laotian students. Genotypes were defined by standard hemoglobin and DNA analyses. Results: The appropriate cutoffs for the Sysmex XS-800i and Sysmex XN-1000, Coulter LH 780, and Pentra ES-60 were found to be 78 IL for MCV and 25 pg for MCH. These were 82 fL and 25 pg for the Cell-Dyn Ruby. Further validation on Thai and Laotian students revealed 100% sensitivity and specificity of higher than 80% for all analyzers. Conclusions: While using MCV in screening requires establishment of appropriate cutoffs, MCH appears to be applicable to all analyzers. Each analyzer should be evaluated and appropriate cutoffs should be established before application in the field.
引用
收藏
页码:545 / 552
页数:8
相关论文
共 19 条
[1]   Carrier detection for beta-thalassemia trait in general Pakistani population: a way forward [J].
Ali, Natasha ;
Moiz, Bushra ;
Bin Azhar, Waleed ;
Zaidi, Nazneen ;
Memon, Rabnawaz .
HEMATOLOGY, 2012, 17 (04) :237-240
[2]  
BAIN B, 1994, J CLIN PATHOL, V47, P289
[3]   The laboratory diagnosis of haemoglobinopathies [J].
Bain, BJ ;
Amos, RJ ;
Bareford, D ;
Chapman, C ;
Davies, SC ;
Old, JM ;
Wild, BJ .
BRITISH JOURNAL OF HAEMATOLOGY, 1998, 101 (04) :783-792
[4]   The diverse molecular basis and hematological features of Hb H and AEBart's diseases in northeast Thailand [J].
Boonsa, S ;
Sanchaisuriya, K ;
Fucharoen, G ;
Wiangnon, S ;
Jetsrisuparb, A ;
Fucharoen, S .
ACTA HAEMATOLOGICA, 2004, 111 (03) :149-154
[5]  
Chaibunruang A, 2010, CLIN LAB, V56, P111
[6]   Should we screen for globin gene mutations in blood samples with mean corpuscular volume (MCV) greater than 80 fL in areas with a high prevalence of thalassaemia? [J].
Chan, LC ;
Ma, SK ;
Chan, AYY ;
Ha, SY ;
Waye, JS ;
Lau, YL ;
Chui, DHK .
JOURNAL OF CLINICAL PATHOLOGY, 2001, 54 (04) :317-320
[7]  
Fucharoen G, 2004, B WORLD HEALTH ORGAN, V82, P364
[8]  
Fucharoen Suthat, 2011, Indian J Med Res, V134, P498
[9]   Evaluation of the URIT-2900 Automated Hematology Analyzer for screening of thalassemia and hemoglobinopathies in Southeast Asian populations [J].
Karnpean, Rossarin ;
Pansuwan, Anupong ;
Fucharoen, Goonnapa ;
Fucharoen, Supan .
CLINICAL BIOCHEMISTRY, 2011, 44 (10-11) :889-893
[10]   α0-THALASSEMIA TRAIT WITH NORMAL RED CELL INDICES: A REPORT OF TWO CASES [J].
Liao, Can ;
Li, Dong-Zhi .
HEMOGLOBIN, 2012, 36 (06) :589-591