共 27 条
Preliminary kinetic evaluation of an immobilized polysaccharide sub-2 μm column using a low dispersion supercritical fluid chromatograph
被引:20
作者:
Berger, Terry A.
[1
]
机构:
[1] SFC Solut Inc, 9435 Downing St, Englewood, FL 34224 USA
关键词:
Sub-2 mu m particles;
Immobilized polysaccharide chiral column;
ChiralPak IA-U;
Ultra-high performance supercritical fluid chromatography (SFC);
Kinetic performance;
Sub-1 min chromatograms;
CHIRAL STATIONARY PHASES;
TOTALLY POROUS SILICA;
LIQUID-CHROMATOGRAPHY;
DRUG DISCOVERY;
M PARTICLES;
PERFORMANCE;
ENANTIOSEPARATIONS;
PURIFICATION;
EFFICIENCY;
SHELL;
D O I:
10.1016/j.chroma.2017.06.021
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
The performance of a 3 x 50 mm, 1.6 mu m d(p) column with an immobilized polysaccharide stationary phase (ChiralPak IA-U) was evaluated for efficiency, and pressure drop, with respect to flow rate and modifier concentration using supercritical fluid chromatography (SFC). This appears to be the first such report using such a column in SFC. A unique low dispersion (ultra-high performance) SFC was used for the evaluation. The minimum reduced plate height of 2.78, indicates that the maximum efficiency was similar to or better than coated polysaccharide columns. Selectivity was different from ChiralPak AD, with the same chiral selector, as reported by many others. At high flows and high methanol concentrations, pump pressures sometimes approached 600 bar. With 5% methanol, pressure vs. flow rate was non-linear suggesting turbulent flow in the connector tubing. The optimum flow rate (F-opt) at 40% methanol was approximate to 0.8 mL/min, where the column efficiency was highest. At 5% methanol, F-opt increased to approximate to 1.6 mL/min, but efficiency degraded noticeably. The differences in F-opt suggests that the solute diffusion coefficients are a strong function of modifier concentration. Several sub-1 min separations, including a 7.5 s separation, are presented. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:82 / 88
页数:7
相关论文