Design of facemasks for delivery of aerosol-based medication via pressurized metered dose inhaler with valved holding chamber: Key issues that affect performance

被引:34
作者
Morton, R. W. [1 ]
Mitchell, J. P. [1 ]
机构
[1] Trudell Med Int, London, ON N5V 5G4, Canada
来源
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG | 2007年 / 20卷
关键词
inhaler; holding chamber; facemask design;
D O I
10.1089/jam.2007.0571
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Valved holding chambers (VHCs) are widely prescribed for use with pressurized metered dose inhalers (pMDIs) for the treatment of respiratory disease by aerosol therapy. The facemask is the preferred patient interface for use by infants and small children, as well as by geriatric patients, due primarily to poor coordination skills. However, care is required in the design of the facemask-VHC system to optimize the delivery of medication. In particular, it is essential to achieve an effective mask-to-face seal and to minimize the volume of dead space. It is also important to ensure that the fit of the facemask is comfortable to the patient when applied with sufficient force to create a seal. We review each of these design principles and their application in the evolution of a range of VHCs from the same family of devices during the past fifteen years. We also examine the various methods available for evaluating VHC-facemasks as a system, recommending where future work might be directed.
引用
收藏
页码:S29 / S42
页数:14
相关论文
共 60 条
[1]   Ability to learn inhaler technique in relation to cognitive score and tests of praxis in old age [J].
Allen, SC ;
Ragab, S .
POSTGRADUATE MEDICAL JOURNAL, 2002, 78 (915) :37-39
[2]  
Amirav I, 2006, RESPIR CARE, V51, P123
[3]   Redesigned face mask improves "Real life" aerosol delivery for nebuchamber [J].
Amirav, I ;
Mansour, Y ;
Mandelberg, A ;
Bar-Ilan, I ;
Newhouse, MT .
PEDIATRIC PULMONOLOGY, 2004, 37 (02) :172-177
[4]   Aerosol therapy with valved holding chambers in young children: Importance of the facemask seal [J].
Amirav, I ;
Newhouse, MT .
PEDIATRICS, 2001, 108 (02) :389-394
[5]  
Amirav I., 2004, ITAL J PEDIATR, V30, P147
[6]  
[Anonymous], 2004, STAND DIAGN MAN PAT
[7]  
Becker A, 2005, CMAJ S, V173, pS1
[8]  
BISGAARD H, 1995, EUR RESPIR J, V8, P856
[9]  
Bisgaard H, 2002, LUNG BIOL HEALTH DIS, V162, P389
[10]   Automatic actuation of a dry powder inhaler into a nonelectrostatic spacer [J].
Bisgaard, H .