Investigation on drying conditions and assays of amidosulfuric acid and sodium carbonate by acidimetric coulometric titration and gravimetric titration

被引:18
作者
Asakai, Toshiaki [1 ]
Hioki, Akiharu [1 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, NMIJ, Tsukuba, Ibaraki 3058563, Japan
关键词
Coulometric titration; Gravimetric titration; Certified reference material; Acid-base titration; High purity amidosulfuric acid; High purity sodium carbonate; SULFAMIC ACID; PURITY;
D O I
10.1007/s00769-010-0646-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Amidosulfuric acid and sodium carbonate as standards for acid-base titrimetry were assayed by coulometric titration and gravimetric titration. Amidosulfuric acid was directly assayed by coulometric titration with electrogenerated hydroxide ions, and sodium carbonate was assayed by gravimetric back-titration. For sodium carbonate, excess amount of sulfuric acid, whose concentration was determined by coulometric titration, was added to sodium carbonate, and then gravimetrically back-titrated using a sodium hydroxide solution whose concentration was determined by gravimetric titration using the sulfuric acid. The accuracy of the coulometric titration for amidosulfuric acid and sulfuric acid was evaluated by examining the current efficiency of pulse electrolysis, the amount of the electrolysis current used, and the time spent for a titration. In addition, the drying conditions for high purity primary standards have a significant effect on the titration results due to changes in the acid-base assay. The suitable drying conditions for amidosulfuric acid and sodium carbonate were evaluated by mass-change measurements, coulometric titration and gravimetric titration. The measurement uncertainties were estimated from the uncertainties on the titration processes. Finally, the assays of amidosulfuric acid and sodium carbonate were 99.986% +/- A 0.010% (k = 2) after drying at 50 A degrees C for 2 h, and 99.970% +/- A 0.016% (k = 2) after drying at 280 A degrees C for 4 h, respectively. In addition, the international consistency was confirmed by measuring certified reference materials (CRMs) available from different National Metrology Institutes, and the compatibility of values among CRMs was experimentally ascertained.
引用
收藏
页码:391 / 399
页数:9
相关论文
共 23 条
[1]  
American Chemical Society, 2000, ACS SPEC REAG CHEM
[2]  
*AN METH COMM, 1969, PUR APPL CHEM, V19, P445
[3]  
Analytical Methods Committee, 1965, ANALYST, V90, P251
[4]  
[Anonymous], 1967, ANALYST, V92, P587
[5]   Investigation of the drying conditions for amidosulfuric acid [J].
Asakai, T ;
Minegishi, R ;
Inaba, N ;
Ishimura, Y ;
Murayama, M ;
Tanaka, T .
ANALYTICAL SCIENCES, 2006, 22 (03) :461-463
[6]   Scheme and studies of reference materials for volumetric analysis in Japan [J].
Asakai, Toshiaki ;
Murayama, Mariko .
ACCREDITATION AND QUALITY ASSURANCE, 2008, 13 (07) :351-360
[7]   Investigation of drying conditions for high-purity sodium carbonate [J].
Asakai, Toshiaki ;
Ogura, Tatsunori ;
Murayama, Mariko ;
Tanaka, Tatsuhiko .
BUNSEKI KAGAKU, 2008, 57 (01) :49-53
[8]   Determination of the purity of acidimetric standards by constant-current coulometry, and the intercomparison between CRMs [J].
Asakai, Toshiaki ;
Murayama, Mariko ;
Tanaka, Tatsuhiko .
ACCREDITATION AND QUALITY ASSURANCE, 2007, 12 (3-4) :151-155
[9]   ACCURACY OF THE MEASUREMENT OF PURITY OF SULFAMIC ACID BY COULOMETRIC TITRATION [J].
HIOKI, A ;
KOKUBUN, A ;
KUBOTA, M .
ANALYST, 1994, 119 (08) :1879-1882
[10]   PRECISE COULOMETRIC TITRATION OF ANTIMONY(III) IN A HIGHLY ACIDIC SOLUTION [J].
HIOKI, A ;
KUBOTA, M ;
KAWASE, A .
ANALYST, 1992, 117 (06) :997-1001