Evaluation of differences between fan-beam and pencil-beam densitometers

被引:44
作者
Mazess, RB [1 ]
Barden, HS [1 ]
机构
[1] Lunar Corp, Madison, WI 53717 USA
关键词
densitometry; X-ray; bones; osteoporosis; body composition;
D O I
10.1007/s002230001147
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We compared the bone and body composition results in vivo on two bone densitometers using fan-beam geometry (EXPERT and PRODIGY) with those using pencil-beam geometry (DPX). Measurements were made on large groups of adults ranging in weight from about 50 to 120 kg. Both spine and femur neck BMD on the fan-beam densitometers averaged within 1% of the pencil-beam results, and there was no magnitude dependence of the results by Bland-Altman analysis. Total body BMC and BMD on the PRODIGY and DPX were congruent, but on the EXPERT, BMC was about 2% lower and BMD 2% higher than corresponding values on the DPX. Soft-tissue composition was closely congruent for the PRODIGY and DPX; the comparable EXPERT-DPX differences showed greater scatter but no significant magnitude dependence. The smaller fan-angle of the PRODIGY (4 degrees) probably contributed to its better congruence to pencil-beam results compared with the EXPERT (12 degrees).
引用
收藏
页码:291 / 296
页数:6
相关论文
共 15 条
  • [1] CROSS CALIBRATION OF QDR-2000 AND QDR-1000 DUAL-ENERGY X-RAY DENSITOMETERS FOR BONE-MINERAL AND SOFT-TISSUE MEASUREMENTS
    ABRAHAMSEN, B
    GRAM, J
    HANSEN, TB
    BECKNIELSEN, H
    [J]. BONE, 1995, 16 (03) : 385 - 390
  • [2] Comparison of two Hologic DXA systems (QDR 1000 and QDR 4500/A)
    Barthe, N
    Braillon, P
    Ducassou, D
    BasseCathalinat, B
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (835) : 728 - 739
  • [3] DUAL X-RAY ABSORPTIOMETRY - A COMPARISON BETWEEN FAN-BEAM AND PENCIL BEAM SCANS
    BLAKE, GM
    PARKER, JC
    BUXTON, FMA
    FOGELMAN, I
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (790) : 902 - 906
  • [4] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [5] Cross-calibration of a fan-beam X-ray densitometer with a pencil-beam system
    Bouyoucef, SE
    Cullum, ID
    Jell, P
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (822) : 522 - 531
  • [6] DIESSEL E, 1998, OSTEOPOROSIS INT, V8, P511
  • [7] Bone mineral and body composition measurements: Cross-calibration of pencil-beam and fan-beam dual-energy X-ray absorptiometers
    Ellis, KJ
    Shypailo, RJ
    [J]. JOURNAL OF BONE AND MINERAL RESEARCH, 1998, 13 (10) : 1613 - 1618
  • [8] Bone mineral density of opposing hips using dual energy X-ray absorptiometry in single-beam and fan-beam design
    Franck, H
    Munz, M
    Scherrer, M
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1997, 61 (06) : 445 - 447
  • [9] Fuerst T, 1996, OSTEOPOROS INT S, V6, P202
  • [10] ENHANCED PRECISION WITH DUAL-ENERGY X-RAY ABSORPTIOMETRY
    MAZESS, R
    CHESNUT, CH
    MCCLUNG, M
    GENANT, H
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1992, 51 (01) : 14 - 17