A recurrent neural network for rapid detection of delivery errors during real-time portal dosimetry

被引:9
作者
Bedford, James L. [1 ,2 ]
Hanson, Ian M. [3 ]
机构
[1] Inst Canc Res, Joint Dept Phys, London SM2 5PT, England
[2] Royal Marsden NHS Fdn Trust, London SM2 5PT, England
[3] Auckland Radiat Oncol, Auckland, New Zealand
关键词
In vivo dosimetry; Electronic portal imaging device; Artificial neural network; Volumetric modulated arc therapy; IN-VIVO DOSIMETRY; DOSE VERIFICATION; EPID IMAGES; RADIOTHERAPY; PROSTATE; OPTIMIZATION; EXPERIENCE; THERAPY; IMRT;
D O I
10.1016/j.phro.2022.03.004
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Real-time portal dosimetry compares measured images with predicted images to detect delivery errors as the radiotherapy treatment proceeds. This work aimed to investigate the performance of a recurrent neural network for processing image metrics so as to detect delivery errors as early as possible in the treatment.Materials and methods: Volumetric modulated arc therapy (VMAT) plans of six prostate patients were used to generate sequences of predicted portal images. Errors were introduced into the treatment plans and the modified plans were delivered to a water-equivalent phantom. Four different metrics were used to detect errors. These metrics were applied to a threshold-based method to detect the errors as soon as possible during the delivery, and also to a recurrent neural network consisting of four layers. A leave-two-out approach was used to set thresholds and train the neural network then test the resulting systems.Results: When using a combination of metrics in conjunction with optimal thresholds, the median segment index at which the errors were detected was 107 out of 180. When using the neural network, the median segment index for error detection was 66 out of 180, with no false positives. The neural network reduced the rate of false negative results from 0.36 to 0.24.Conclusions: The recurrent neural network allowed the detection of errors around 30% earlier than when using conventional threshold techniques. By appropriate training of the network, false positive alerts could be prevented, thereby avoiding unnecessary disruption to the patient workflow.
引用
收藏
页码:36 / 43
页数:8
相关论文
共 52 条
[1]   Optimisation of a composite difference metric for prompt error detection in real-time portal dosimetry of simulated volumetric modulated arc therapy [J].
Bedford, James L. ;
Hanson, Ian M. .
BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1120)
[2]   A method to verify sections of arc during intrafraction portal dosimetry for prostate VMAT [J].
Bedford, James L. ;
Hanson, Ian M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (20)
[3]   Comparison of forward- and back-projection in vivo EPID dosimetry for VMAT treatment of the prostate [J].
Bedford, James L. ;
Hanson, Ian M. ;
Hansen, Vibeke N. .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (02)
[4]   Portal dosimetry for VMAT using integrated images obtained during treatment [J].
Bedford, James L. ;
Hanson, Ian M. ;
Hansen, Vibeke Nordmark .
MEDICAL PHYSICS, 2014, 41 (02)
[5]   Sinogram analysis of aperture optimization by iterative least-squares in volumetric modulated arc therapy [J].
Bedford, James L. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (05) :1235-1250
[6]   Safety and Efficacy of a Five-Fraction Stereotactic Body Radiotherapy Schedule for Centrally Located Non-Small-Cell Lung Cancer: NRG Oncology/RTOG 0813 Trial [J].
Bezjak, Andrea ;
Paulus, Rebecca ;
Gaspar, Laurie E. ;
Timmerman, Robert D. ;
Straube, William L. ;
Ryan, William F. ;
Garces, Yolanda I. ;
Pu, Anthony T. ;
Singh, Anurag K. ;
Videtic, Gregory M. ;
McGarry, Ronald C. ;
Iyengar, Puneeth ;
Pantarotto, Jason R. ;
Urbanic, James J. ;
Sun, Alexander Y. ;
Daly, Megan E. ;
Grills, Inga S. ;
Sperduto, Paul ;
Normolle, Daniel P. ;
Bradley, Jeffrey D. ;
Choy, Hak .
JOURNAL OF CLINICAL ONCOLOGY, 2019, 37 (15) :1316-+
[7]   Quantifying the performance of in vivo portal dosimetry in detecting four types of treatment parameter variations [J].
Bojechko, C. ;
Ford, E. C. .
MEDICAL PHYSICS, 2015, 42 (12) :6912-6918
[8]   Hypofractionated breast radiotherapy for 1 week versus 3 weeks (FAST-Forward): 5-year efficacy and late normal tissue effects results from a multicentre, non-inferiority, randomised, phase 3 trial [J].
Brunt, Adrian Murray ;
Haviland, Joanne S. ;
Wheatley, Duncan A. ;
Sydenham, Mark A. ;
Alhasso, Abdulla ;
Bloomfield, David J. ;
Chan, Charlie ;
Churn, Mark ;
Cleator, Susan ;
Coles, Charlotte E. ;
Goodman, Andrew ;
Harnett, Adrian ;
Hopwood, Penelope ;
Kirby, Anna M. ;
Kirwan, Cliona C. ;
Morris, Carolyn ;
Nabi, Zohal ;
Sawyer, Elinor ;
Somaiah, Navita ;
Stones, Liba ;
Syndikus, Isabel ;
Bliss, Judith M. ;
Yarnold, John R. .
LANCET, 2020, 395 (10237) :1613-1626
[9]   AI applications to medical images: From machine learning to deep learning [J].
Castiglioni, Isabella ;
Rundo, Leonardo ;
Codari, Marina ;
Leo, Giovanni Di ;
Salvatore, Christian ;
Interlenghi, Matteo ;
Gallivanone, Francesca ;
Cozzi, Andrea ;
D'Amico, Natascha Claudia ;
Sardanelli, Francesco .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 83 :9-24
[10]  
Cho K., 2014, COMPUT SCI