Calcium phosphate cement and locked plate augmentation of distal femoral defects: A biomechanical analysis

被引:5
作者
DeBaun, Malcolm R. [4 ]
Williams, Joel C. [2 ]
Bennett, Chase G. [4 ]
Pridgen, Eric M. [1 ]
Tamurian, Robert M. [3 ]
Amanatullah, Derek F. [4 ]
机构
[1] Univ Penn, Dept Orthopaed Surg, Perelman Sch Med, 3737 Market St,Suite 6121, Philadelphia, PA 19104 USA
[2] Rush Univ, Med Ctr, Dept Orthopaed Surg, 1611 W Harrison St,Suite 30,0, Chicago, IL 60612 USA
[3] Multicare Hlth Syst, Orthoped & Sports Med, 3400 South 19th St,Suite 340, Tacoma, WA 98405 USA
[4] Stanford Univ, Dept Orthoped Surg, 450 Broadway St, Redwood City, CA 94063 USA
关键词
Distal femur; Metaphyseal; Defect; Tumor; Axial; Torsion; Stiffness; GIANT-CELL TUMOR; BONE-GRAFT; STRENGTH REDUCTION; INTERNAL-FIXATION; CURETTAGE; DISEASE; FEMUR;
D O I
10.1016/j.knee.2019.07.006
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: Bone tumors are common in the distal femur and often treated with intralesional curettage. The optimal method of stabilization of large distal femoral defects after curettage remains unclear. The goal of this study is to compare stabilization techniques for large distal femoral defects. Methods: Large defects (60 cm(3)) were milled in the distal lateral metaphysis of 45 adult composite sawbone femurs. The defect was either (1) left untreated or reconstructed with (2) locked plate fixation, (3) calcium phosphate cement packing, or (4) locked plate fixation with calcium phosphate cement packing, or (5) polymethylmethacrylate packing. Each specimen then underwent axial and torsional stiffness testing followed by torsional loading to failure. The data were analyzed using ANOVA with Tukey-Kramer post-hoc analysis. Results: The calcium phosphate cement filled defect with a locked plate was the stiffest construct in axial and torsional loading as well as the strongest in torque to failure. However, this difference only reached significance with respect to all other groups in torque to failure testing. The calcium phosphate cement filled defect with a locked plate was significantly stiffer than three of the four other groups in both axial and torsional stiffness testing. Conclusions: These results indicate that calcium phosphate cement, with or without the addition of locked plate fixation, may provide improved construct stability under time zero testing conditions. This result warrants further testing under cyclic loading condition and consideration for fixation of large femoral metaphyseal defects in future clincal trails. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页码:1020 / 1025
页数:6
相关论文
共 27 条
[1]   Torsional Properties of Distal Femoral Cortical Defects [J].
Amanatullah, Derek F. ;
Williams, Joel C. ;
Fyhrie, David P. ;
Tamurian, Robert M. .
ORTHOPEDICS, 2014, 37 (03) :158-162
[2]  
BINI SA, 1995, CLIN ORTHOP RELAT R, P245
[3]   Properties of calcium phosphate ceramics in relation to their in vivo behavior [J].
Blokhuis, TJ ;
Termaat, MF ;
den Boer, FC ;
Patka, P ;
Bakker, FC ;
Haarman, HJTM .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2000, 48 (01) :179-186
[4]   REMODELING OF LARGE BONE DEFECTS IN THE TREATMENT OF SPACE-OCCUPYING LESIONS - CURETTAGE WITHOUT BONE-GRAFT FOR TREATING BENIGN BONE-TUMORS [J].
CHIGIRA, M ;
WATANABE, H ;
ARITA, S ;
NODA, K ;
SHIMIZU, T ;
SHINOZAKI, T ;
NAGASE, M .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1992, 111 (02) :61-65
[5]  
Dennis D A, 2001, Instr Course Lect, V50, P379
[6]   TORSIONAL STRENGTH REDUCTION DUE TO CORTICAL DEFECTS IN BONE [J].
EDGERTON, BC ;
AN, KN ;
MORREY, BF .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1990, 8 (06) :851-855
[7]   Fracture site augmentation with calcium phosphate cement reduces screw penetration after open reduction-internal fixation of proximal humeral fractures [J].
Egol, Kenneth A. ;
Sugi, Michelle T. ;
Ong, Crispin C. ;
Montero, Nicole ;
Davidovitch, Roy ;
Zuckerman, Joseph D. .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2012, 21 (06) :741-748
[8]   Bone-grafting and bone-graft substitutes [J].
Finkemeier, CG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (03) :454-464
[9]   Volume effect on biological properties of a calcium phosphate hydraulic cement:: Experimental study in sheep [J].
Flautre, B ;
Delecourt, C ;
Blary, MC ;
Van Landuyt, P ;
Lemaître, J ;
Hardouin, P .
BONE, 1999, 25 (02) :35S-39S
[10]   Biomechanical and histological evaluation of a calcium phosphate cement [J].
Frankenburg, EP ;
Goldstein, SA ;
Bauer, TW ;
Harris, SA ;
Poser, RD .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (08) :1112-1124