3 Tesla MRI of Patients With a Vagus Nerve Stimulator: Initial Experience Using a T/R Head Coil Under Controlled Conditions

被引:17
作者
Gorny, Krzysztof R. [1 ]
Bernstein, Matt A. [1 ]
Watson, Robert E., Jr. [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
关键词
MRI; safety; vagus nerve stimulation; VNS; 3T; 3.0T; heating; IN-VITRO EVALUATION; DEEP BRAIN-STIMULATION; THERAPY SYSTEM; VNS; EPILEPSY; MECHANISMS; IMPLANT; ADULTS;
D O I
10.1002/jmri.22037
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess safety of clinical MRI of the head in patients with implanted model 100, 102, and 103 vagus nerve stimulation (VNS) Therapy Systems (Cyberonics, Inc., Houston. TX) in 3.0 Tesla MRI (GE Healthcare. Milwaukee, WI). Materials and Methods: The distributions of the radiofrequency B-1(+)-field produced by the clinically used transmit/receive (T/R) head coil (Advanced Imaging Research Incorporated, Cleveland, OH) and body coil were measured in a head and shoulders phantom. These measurements were supplemented by temperature measurements on the lead tips and the implantable pulse generator (IPG) of the VNS devices in a head and torso phantom with the same two coils. Clinical 3T MRI head scans were then acquired under highly controlled conditions in a series of 17 patients implanted with VNS. Results: Phantom studies showed only weak B-1(+) fields at the location of the VNS IPG and leads for MRI scans using the T/R head coil. The MRI-related heating on a VNS scanned in vitro at 3T was also found to be minimal (0.4-0.8 degrees C at the leads, negligible at the IPG). The patient MRI examinations were completed successfully without any adverse incidents. No patient reported any heating, discomfort, or any other unusual sensation. Conclusion: Safe clinical MRI head scanning of patients with implanted VNS is shown to be feasible on a GE Signa Excite 3T MRI system using one specific T/R head coil. These results apply to this particular MRI system configuration. Extrapolation or generalization of these results to more general or less controlled imaging situations without supporting data of safety is highly discouraged.
引用
收藏
页码:475 / 481
页数:7
相关论文
共 37 条
[1]  
ASTM standard F2182-02a, 2002, ASTM STANDARD F2182, DOI [10.1520/F2182-02A, DOI 10.1520/F2182-02A]
[2]   Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating [J].
Baker, KB ;
Tkach, JA ;
Nyenhuis, JA ;
Phillips, M ;
Shellock, FG ;
Gonzalez-Martinez, J ;
Rezai, AR .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (02) :315-320
[3]   Variability in RF-induced heating of a deep brain stimulation implant across MR systems [J].
Baker, Kenneth B. ;
Tkach, Jean A. ;
Phillips, Micheal D. ;
Rezai, Ali R. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 24 (06) :1236-1242
[4]  
BENDABIS SR, 2001, SEIZURE, V10, P512
[5]   Vagus nerve stimulation acutely alters food craving in adults with depression [J].
Bodenlos, Jamie S. ;
Kose, Samet ;
Borckardt, Jeffrey J. ;
Nahas, Ziad ;
Shaw, Darlene ;
O'Neil, Patrick M. ;
George, Mark S. .
APPETITE, 2007, 48 (02) :145-153
[6]  
Borckardt Jeffrey J, 2005, Pain Res Manag, V10, P9
[7]   Acute and long-term VNS effects on pain perception in a case of treatment-resistant depression [J].
Borckardt, Jeffrey J. ;
Anderson, Berry ;
Kozel, F. Andrew ;
Nahas, Ziad ;
Smith, A. Richard ;
Thomas, K. Jackson ;
Kose, Samet ;
George, Mark S. .
NEUROCASE, 2006, 12 (04) :216-220
[8]  
Briellmann RS, 2003, EPILEPTIC DISORD, V5, P3
[9]   Weight loss associated with vagus nerve stimulation [J].
Burneo, JG ;
Faught, E ;
Knowlton, R ;
Morawetz, R ;
Kuzniecky, R .
NEUROLOGY, 2002, 59 (03) :463-464
[10]   Long-term outcome of vagus nerve stimulation for refractory partial epilepsy [J].
Chavel, SM ;
Westerveld, M ;
Spencer, S .
EPILEPSY & BEHAVIOR, 2003, 4 (03) :302-309