Evaluation of dose profiles using Gafchromic EBT3 films in Leksell Gamma Knife 4C around inhomogeneities in the treatment of pituitary adenoma in anthropomorphic heterogeneous head phantom

被引:4
作者
Najafi, M. [1 ]
Shirazi, A. [1 ]
Geraily, Gh. [1 ]
Esfahani, M. [3 ]
Teimouri, J. [2 ]
机构
[1] Univ Tehran Med Sci, Med Phys & Biomed Engn Dept, Tehran, Iran
[2] Iran Gamma Knife Ctr, Tehran, Iran
[3] Univ Tehran Med Sci, Imam Khomeini Hosp, Tehran, Iran
关键词
Gamma knife; Heterogeneous phantom; EBT3; film; Pituitary adenoma; POLYMER GEL DOSIMETER; STEREOTACTIC RADIOSURGERY; TISSUE INHOMOGENEITIES; MC SIMULATION; UNIT;
D O I
10.1016/j.radphyschem.2018.04.003
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Purpose: The Leksell Gamma Knife is a complete system for radiosurgery. Treatment dose planning of Gamma Knife assumes the human head as a homogenous media; thus received dose by target tissue and organs at risk may be lower or higher than predicted dose. Our main interest is to compare measured dose profiles in an anthropomorphic heterogeneous head phantom by EBT3 films with calculated dose by Leksell Gamma Plan. Methods: An adult head phantom with size, shape and inhomogeneity the same as an adult human head was constructed. Then dose profiles across from isocenter and organs at risk were measured by EBT3 films and compared with those predicted by Leksell Gamma Plan. Results: Our findings indicate that when isocenter placed at the target tissue near the frontal sinus cavity, overestimation and underestimation of the dose value affect the dose profiles. In some points of organs at risk, measured dose is more than calculated dose. Comparison of isodose curves near air-tissue interface shows that in higher isodose levels, dose perturbation is more seen. Distance to Agreement parameter for some isodose lines exceeds the total positioning error for Gamma Knife ( +/- 2 mm). Also dose differences are more than numerical accuracy in dose delivery for stereotactic radiosurgery ( +/- 5%). Conclusions: Inhomogeneities in head impress dose distributions. Ignorance of inhomogeneities can cause the parts of target tissue receive under dosage and also in some points, organs at risk receive more dose. Results of this study are closer to reality because of anthropomorphic heterogeneous head phantom.
引用
收藏
页码:104 / 109
页数:6
相关论文
共 25 条
[1]   Effects of bone- and air-tissue inhomogeneities on the dose distributions of the Leksell Gamma Knife® calculated with PENELOPE [J].
Al-Dweri, FMO ;
Rojas, EL ;
Lallena, AM .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (23) :5665-5678
[2]   A new formalism for reference dosimetry of small and nonstandard fields [J].
Alfonso, R. ;
Andreo, P. ;
Capote, R. ;
Huq, M. Saiful ;
Kilby, W. ;
Kjaell, P. ;
Mackie, T. R. ;
Palmans, H. ;
Rosser, K. ;
Seuntjens, J. ;
Ullrich, W. ;
Vatnitsky, S. .
MEDICAL PHYSICS, 2008, 35 (11) :5179-5186
[3]   Radiosurgery for pituitary adenomas: evaluation of its efficacy and safety [J].
Castro, Douglas G. ;
Cecilio, Soraya A. J. ;
Canteras, Miguel M. .
RADIATION ONCOLOGY, 2010, 5
[4]   Analysis of the EBT3 Gafchromic film irradiated with 6 MV photons and 6 MeV electrons using reflective mode scanners [J].
Farah, Nicolas ;
Francis, Ziad ;
Abboud, Marie .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (06) :708-712
[5]  
Gray H., 2006, ANATOMY HUMAN BODY S
[6]  
Guide, 1998, GUID MATLAB US 5, P333
[7]   Gamma Knife 3-D dose distribution near the area of tissue inhomogeneities by normoxic gel dosimetry [J].
Isbakan, Fatih ;
Ulgen, Yekta ;
Bilge, Hatice ;
Ozen, Zeynep ;
Agus, Onur ;
Buyuksarac, Bora .
MEDICAL PHYSICS, 2007, 34 (05) :1623-1630
[8]   Hypofractionated stereotactic radiotherapy with CyberKnife for nonfunctioning pituitary adenoma: high local control with low toxicity [J].
Iwata, Hiromitsu ;
Sato, Kengo ;
Tatewaki, Koshi ;
Yokota, Naoki ;
Inoue, Mitsuhiro ;
Baba, Yoshimi ;
Shibamoto, Yuta .
NEURO-ONCOLOGY, 2011, 13 (08) :916-922
[9]  
Kim Won, 2012, Surg Neurol Int, V3, pS79, DOI 10.4103/2152-7806.95419
[10]   Dose-response tolerance of the visual pathways and cranial nerves of the cavernous sinus to stereotactic radiosurgery [J].
Leber, KA ;
Berglöff, J ;
Pendl, G .
JOURNAL OF NEUROSURGERY, 1998, 88 (01) :43-50