Determination of selenium in Chinese herbs by hydride generation atomic fluorescence spectrometry

被引:0
作者
Yang, LL [1 ]
Zhang, DQ
Gao, Y
Yuan, CG
Zhang, YX
Sun, HW
机构
[1] Hebei Univ, Coll Chem & Environm Sci, Baoding 071002, Peoples R China
[2] Hebei Acad Sci, Shijiazhuang 050081, Peoples R China
关键词
hydride generation; atomic fluorescence spectrometry; Chinese herbs; selenium;
D O I
暂无
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A method for the determination of selenium by hydride generation atomic fluorescence spectrometry (HG-AFS) was proposed in this paper. The experimental conditions including the temperature of atomizer, height of atomizer, lamp current, flow rate of carrier gas, concentration of KBH4, and acidic medium that affect the generation of hydride and the sensitivity of the method were investigated and optimized. The influence of coexisting foreign ions on the determination of selenium was also investigated. The detection limit of the proposed method was found to be 0.21 ng.mL(-1). The proposed method was successfully applied to the determination of selenium in ten kinds of Chinese herbs, and the recoveries were in the range of 95%-105%.