The thermal neutron response of miniature silicon carbide semiconductor detectors

被引:58
作者
Dulloo, AR
Ruddy, FH
Seidel, JG
Adams, JM
Nico, JS
Gilliam, DM
机构
[1] Westinghouse Sci & Technol Dept, Pittsburgh, PA 15235 USA
[2] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
关键词
silicon carbide; neutron detection; radiation sensor;
D O I
10.1016/S0168-9002(02)01987-3
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Neutron response calibrations have been performed for miniature SiC semiconductor detectors based on a Schottky diode design in neutron fields maintained at the National Institute of Standards and Technology (NIST). The SiC diodes detect neutrons via neutron-induced charged particles (tritons) produced by the Li-6(n,alpha)H-3 reaction. The neutron response was calibrated at fluence rates from 1.76 x 10(4) to 3.59 x 10(10) cm(-2) s(-1) in NIST neutron fields. The maximum deviation from linearity of the fit of the neutron response of the SiC detectors to the NIST-measured fluence rates is much less than 5%, which is the estimated uncertainty in the measured fluence rates. A direct comparison of the SiC count rates to count rates obtained with a NIST double fission chamber over limited ranges where the detector positions and configurations remain unchanged shows a relative precision of +/-0.6%. In a separate set of calibrations, a SiC neutron detector that had been previously irradiated with a fast (E > 1 MeV) neutron fluence of 1.3 x 10(16) cm(-2) was found to have an absolute neutron fluence rate response that was indistinguishable from that of a previously unirradiated detector. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:415 / 423
页数:9
相关论文
共 12 条
[1]  
BABCOCK RV, 1963, IAEA TECHN P, P613
[2]   The neutron response of miniature silicon carbide semiconductor detectors [J].
Dulloo, AR ;
Ruddy, FH ;
Seidel, JG ;
Adams, JM ;
Nico, JS ;
Gilliam, DM .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3) :47-48
[3]   Simultaneous measurement of neutron and gamma-ray radiation levels from a TRIGA reactor core using silicon carbide semiconductor detectors [J].
Dulloo, AR ;
Ruddy, FH ;
Seidel, JG ;
Davison, C ;
Flinchbaugh, T ;
Daubenspeck, T .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (03) :275-279
[4]   MEASUREMENT OF ABSOLUTE FISSION RATES [J].
GRUNDL, JA ;
GILLIAM, DM ;
DUDEY, ND ;
POPEK, RJ .
NUCLEAR TECHNOLOGY, 1975, 25 (02) :237-257
[5]  
*ISO INT ORG STAND, 1999, GUID EXPR UNC MEAS
[6]  
LAMAZE GP, 1988, NBS SPECIAL PUBLICAT
[7]  
LOTT RG, 1998, Patent No. 5726453
[8]  
MCELROY WN, 1971, HEDLASA304
[9]   Development of a silicon carbide radiation detector [J].
Ruddy, FH ;
Dulloo, AR ;
Seidel, JG ;
Seshadri, S ;
Rowland, LB .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :536-541
[10]  
RUDDY FH, 2002, 11 INT S REACT DOS B