Challenges in hemodynamics assessment in complex neurovascular geometries using computational fluid dynamics and benchtop flow simulation in 3D printed patient specific phantoms

被引:0
作者
Paccione, Eric [1 ]
Ionita, Ciprian N. [1 ,2 ]
机构
[1] Univ Buffalo, Univ Dept Biomed Engn, Buffalo, NY 14203 USA
[2] Canon Stroke & Vasc Res Ctr, Buffalo, NY USA
来源
MEDICAL IMAGING 2021: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING | 2021年 / 11600卷
关键词
carotid stenosis; stroke; CFD; 3D-printing; hemodynamics;
D O I
10.1117/12.2582169
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Purpose: Complex hemodynamics assessments, as those related to carotid stenosis, are not always easily straightforward due to multifaceted challenges presented by the collateral flow in the Circle of Willis (CoW) and brain flow autoregulation. Advanced computational and benchtop methods to investigate hemodynamics aspects related to such complex flows are often used, however both have limitations and could lead to results which may diverge. In this study we investigated these aspects by performing correlated computational fluid dynamics (CFD) simulations and benchtop experiments in patient specific 3D printed phantoms. Materials and Methods: To investigate the flow in patients with carotid stenosis, we built two patient specific phantoms which contained the arterial lesion of interest, all main arteries leading to the brain, the CoW and main arteries branching from it. Each phantom was connected to a generic aortic arch. A programmable pump was connected and flow parameters were measured proximal and distal to the lesion and the contralateral arteries. The patient 3D geometry was used to perform a set of CFD simulations where inflow boundary conditions matched the experimental ones. Flow conditions were recorded at the same locations as the experimental setup. Further exploration into the translation from experimental to CFD was also performed by customizing vascular segmentation and physically manipulating arterial compliance properties. Results: We initially observed significant differences between the CFD recordings and the experimental setup. Most of the differences were due to changes in phantom geometry when subjected to physiological pressures and simplistic outflow boundary conditions in the CFD simulations which do not account for pulsatility and nonlinear phenomena. Further work confirms the need for dynamic mesh behavior within CFD simulations attempting to computationally mimic 3D-printed benchtop experiments. Additionally, CFD simulation may benefit from considering geometry specific to a 3D-printed vascular phantom.
引用
收藏
页数:8
相关论文
共 19 条
  • [1] Benjamin EJ, 2017, CIRCULATION, V135, pE146, DOI [10.1161/CIR.0000000000000485, 10.1161/CIR.0000000000000558, 10.1161/CIR.0000000000000530]
  • [2] Carotid Artery Stenosis as a Cause of Stroke
    Flaherty, Matthew L.
    Kissela, Brett
    Khoury, Jane C.
    Alwell, Kathleen
    Moomaw, Charles J.
    Woo, Daniel
    Khatri, Pooja
    Ferioli, Simona
    Adeoye, Opeolu
    Broderick, Joseph R.
    Kleindorfer, Dawn
    [J]. NEUROEPIDEMIOLOGY, 2013, 40 (01) : 36 - 41
  • [3] Challenges and limitations of patient-specific vascular phantom fabrication using 3D Polyjet printing
    Ionita, Ciprian N.
    Mokin, Maxim
    Varble, Nicole
    Bednarek, Daniel R.
    Xiang, Jianping
    Snyder, Kenneth V.
    Siddiqui, Adnan H.
    Levy, Elad I.
    Meng, Hui
    Rudin, Stephen
    [J]. MEDICAL IMAGING 2014: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2014, 9038
  • [4] Particle image velocimetry (PIV) evaluation of flow modification in aneurysm phantoms using asymmetric stents
    Ionita, CN
    Hoi, Y
    Meng, H
    Rudin, S
    [J]. MEDICAL IMAGING 2004: PHYSIOLOGY, FUNCTION, AND STRUCTURE FROM MEDICAL IMAGES, 2004, 5 (23): : 295 - 306
  • [5] 3D Printed Cardiac Phantom for Procedural Planning of a Transcatheter Native Mitral Valve Replacement
    Izzo, Richard L.
    O'Hara, Ryan P.
    Iyer, Vijay
    Hansen, Rose
    Meess, Karen M.
    Nagesh, S. V. Setlur
    Rudin, Stephen
    Siddiqui, Adnan H.
    Springer, Michael
    Ionita, Ciprian N.
    [J]. MEDICAL IMAGING 2016: PACS AND IMAGING INFORMATICS: NEXT GENERATION AND INNOVATIONS, 2016, 9789
  • [6] 3D Printed Abdominal Aortic Aneurysm Phantom for Image Guided Surgical Planning with a Patient Specific Fenestrated Endovascular Graft System
    Meess, Karen M.
    Izzo, Richard L.
    Dryjski, Maciej L.
    Curl, Richard E.
    Harris, Linda M.
    Springer, Michael
    Siddiqui, Adnan H.
    Rudin, Stephen
    Ionita, Ciprian N.
    [J]. MEDICAL IMAGING 2017: IMAGING INFORMATICS FOR HEALTHCARE, RESEARCH, AND APPLICATIONS, 2017, 10138
  • [7] Comparison of Modern Stroke Thrombectomy Approaches Using an In Vitro Cerebrovascular Occlusion Model
    Mokin, M.
    Nagesh, S. V. Setlur
    Ionita, C. N.
    Levy, E. I.
    Siddiqui, A. H.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (03) : 547 - 551
  • [8] Assessment of distal access catheter performance during neuroendovascular procedures: measuring force in three-dimensional patient specific phantoms
    Mokin, Maxim
    Waqas, Muhammad
    Nagesh, Swetadri Vasan Setlur
    Karkhanis, Nitant Vivek
    Levy, Elad I.
    Ionita, Ciprian N.
    Siddiqui, Adnan H.
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2019, 11 (06) : 619 - 623
  • [9] Stent retriever thrombectomy with the Cover accessory device versus proximal protection with a balloon guide catheter: in vitro stroke model comparison
    Mokin, Maxim
    Nagesh, Swetadri Vasan Setlur
    Ionita, Ciprian N.
    Mocco, J.
    Siddiqui, Adnan H.
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2016, 8 (04) : 413 - 417
  • [10] Primary stentriever versus combined stentriever plus aspiration thrombectomy approaches: in vitro stroke model comparison
    Mokin, Maxim
    Ionita, Ciprian N.
    Nagesh, Swetadri Vasan Setlur
    Rudin, Stephen
    Levy, Elad I.
    Siddiqui, Adnan H.
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2015, 7 (06) : 453 - 457